Literature DB >> 7364839

Irradiation of the surgeon during pinning of femoral fractures.

A A Giachino, M Cheng.   

Abstract

The intensity of scattered radiation in both a forward and a backward direction from a portable fluoroscope during pinning of the neck of the femur was measured by experiments on a cadaver. The intensity decreased rapidly with increasing distance from the flare of the greater trochanter. Hazard to the surgeon's hands from scattered radiation can be greatly reduced by positioning the C-arm of the fluoroscope so that the x-ray beam is directed laterally.

Entities:  

Mesh:

Year:  1980        PMID: 7364839

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  10 in total

1.  Radiation exposure during 3D fluoroscopy of the knee: an experimental study.

Authors:  T Tharmviboonsri; K Riansuwan; A Nitising; B Mahaisavariya
Journal:  Eur J Trauma Emerg Surg       Date:  2011-11-19       Impact factor: 3.693

2.  Occupational radiation exposure from C arm fluoroscopy during common orthopaedic surgical procedures and its prevention.

Authors:  Anupam Mahajan; Sumant Samuel; Atul K Saran; M K Mahajan; M K Mam
Journal:  J Clin Diagn Res       Date:  2015-03-01

3.  [Radiation burden to the hands of surgeons in intramedullary nailing].

Authors:  L P Müller; J Suffner; K Wenda; W Mohr; L Rudig
Journal:  Unfallchirurgie       Date:  1996-12

4.  Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?

Authors:  Talia Chapman; Dennis P Martin; Christopher Williamson; Brian Tinsley; Mark L Wang; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2017-06-23

Review 5.  C-arm fluoroscopy in orthopaedic surgical practice.

Authors:  Ishaq Ojodu; Ayodele Ogunsemoyin; Sascha Hopp; Tim Pohlemann; Oluwole Ige; Oluwaseun Akinola
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-05-23

6.  Are Indian orthopaedic surgeons aware of the health hazards of radiation exposure? A survey and review on awareness and ways to mitigate them.

Authors:  Binoti Sheth; Arzan Jesia; Rohan Parihar; Rohit Somani; Pankaj Pawar
Journal:  J Clin Orthop Trauma       Date:  2022-08-07

Review 7.  Radiation exposure of eyes, thyroid gland and hands in orthopaedic staff: a systematic review.

Authors:  Chandrasekharan Nair Kesavachandran; Frank Haamann; Albert Nienhaus
Journal:  Eur J Med Res       Date:  2012-10-30       Impact factor: 2.175

8.  Exposure of the surgical team to ionizing radiation during orthopedic surgical procedures.

Authors:  Evandro Pereira Palácio; André Araújo Ribeiro; Bruno Moreira Gavassi; Gabriel Guimarães Di Stasi; José Antônio Galbiatti; Alcides Durigam Junior; Roberto Ryuiti Mizobuchi
Journal:  Rev Bras Ortop       Date:  2014-04-25

9.  FRACTURES OF THE DISTAL EXTREMITY OF THE TIBIA TREATED WITH INTRAMEDULLARY NAIL OR BRIDGE PLATE: COMPARISON OF RADIATION EXPOSURE DURATION BETWEEN THE TWO METHODS.

Authors:  Pedro José Labronici; Fábio Soares Lyra; Ildeu Leite Moreira; Rolix Hoffmann; José Sergio Franco; Paulo Roberto Barbosa de Toledo Lourenço; Gustavo José Labronici
Journal:  Rev Bras Ortop       Date:  2015-11-16

10.  Pulsation and Collimation During Fluoroscopy to Decrease Radiation: A Cadaver Study.

Authors:  Kazuta Yamashita; Kosaku Higashino; Hiroaki Hayashi; Fumio Hayashi; Yoshihiro Fukui; Koichi Sairyo
Journal:  JB JS Open Access       Date:  2017-12-19
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.