Literature DB >> 7294989

Mechanical failure modes in total knee replacement.

H U Cameron, G M McNeice.   

Abstract

In an attempt to define failure modes, a study of 39 knees which failed mechanically has been carried out. In 7 cases failure occurred on the femoral side and in 32 on the tibial side. There was fracture of one femoral component. All femoral loosenings and/or sinkage occurred in malaligned components. Two main tibial failure modes were noted, lateral subluxation of the tibia on the femur leading to loosening and sinkage of the tibial component. Release of popliteus prevents lateral subluxation. Correct component alignment and the use of a large area tibial component resting on the cortices should decrease the incidence of tibial sinkage.

Entities:  

Mesh:

Year:  1981        PMID: 7294989     DOI: 10.1007/bf00460801

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0344-8444


  6 in total

1.  Mechanical causes of loosening in knee joint replacement.

Authors:  A Kagan
Journal:  J Biomech       Date:  1977       Impact factor: 2.712

2.  Polycentric total knee arthroplasty.

Authors:  R S Bryan; L F Peterson
Journal:  Orthop Clin North Am       Date:  1973-04       Impact factor: 2.472

3.  Loosening of the femoral component in surface replacement of the knee.

Authors:  J Colley; H U Cameron; M A Freeman; S A Swanson
Journal:  Arch Orthop Trauma Surg       Date:  1978-08-11

4.  Review of a failed knee replacement and some observations on the design of a knee resurfacing prosthesis.

Authors:  H U Cameron; D M Fedorkow
Journal:  Arch Orthop Trauma Surg       Date:  1980

5.  Mechanical tests on the tibial components of non-hinged knee prostheses.

Authors:  J H Bargren; W H Day; M A Freeman; S A Swanson
Journal:  J Bone Joint Surg Br       Date:  1978-05

6.  Total knee dislocation due to rotatory malalignment of tibial component: a case report.

Authors:  J H Bargren
Journal:  Clin Orthop Relat Res       Date:  1980 Mar-Apr       Impact factor: 4.176

  6 in total
  4 in total

1.  Bone strength measurements at the proximal tibia. Penetration tests and epiphyseal compressive strength.

Authors:  I Hvid; J Jensen; S Nielsen
Journal:  Int Orthop       Date:  1986       Impact factor: 3.075

2.  Dual-photon absorptiometry of the proximal tibia.

Authors:  I Hvid; C Hasling; S L Hansen; H H Hansen
Journal:  Arch Orthop Trauma Surg       Date:  1987

3.  Torsional characteristics of the Minns meniscal knee prosthesis.

Authors:  R J Minns
Journal:  Arch Orthop Trauma Surg       Date:  1989       Impact factor: 3.067

4.  Total condylar knee arthroplasty. A report of 2-year follow-up on 247 cases.

Authors:  S Nielsen; I Hvid; O Sneppen
Journal:  Arch Orthop Trauma Surg       Date:  1985
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.