In vitro hydrodynamic comparison of mitral valve prostheses at high flow rates.

S Gabbay, D M McQueen, E L Yellin, R M Becker, R W Frater.   

Abstract

A pulse duplicator system for evaluating the hemodynamic performance of mitral prostheses is described. Under conditions stimulating normal resting physiology, all valves tested had measurable but acceptably small pressure drops. Under conditions simulating exercise, all were moderately to severely stenotic. Valves with nearly equal mounting diameters were compared. The Hancock, Beall, and Starr-Edwards valves (Group A) were found to be significantly more stenotic than the Björk-Shiley, Cutter-Cooley, Ionescu-Shiley, and Lillehei-Kaster valves (Group B). In the 29 to 30 mm. mounting diameter size at cardiac outputs of 5 and 9 L. per minute, Group A had average pressure drops of 3.2 and 10.5 mm. Hg and Group B, pressure drops of 1.6 and 5.3 mm. Hg, respectively. In the 24 to 26 mm. mounting diameter size, at cardiac outputs of 9 L. per minute, all the valves had critically large pressure drops (9 to 17.6 mm. Hg). The standard Gorlin formula is inappropriate for computing the orifice area of prosthetic valves. The discharge coefficient for a valve (a measure of how well the valve uses its primary flow area) and a performance index (a measure of how well the valve uses its mounting area) have been computed from a knowledge of the orifice size, without the necessity of assuming a value for the discharge coefficient required by the Gorlin formula. The biological valves (Hancock and Ionescu-Shiley) provide an efficient orifice for fluid flow at the free leaflet margins and have large discharge coefficients. On the basis of the fluid dynamic equation of motion, steady flow, root mean square (RMS) flow, and peak flow, combined with the appropriate transvalvular gradients, were all shown to yield equally accurate characterizations of valvular hydrodynamic performance. Mean flow, unfortunately the only value obtainable clinically, yielded effective orifice areas 10 percent smaller than either of the other three flow values.

Mesh:

Year:  1978        PMID: 713584

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  9 in total

1.  Limitations of Doppler ultrasound in the assessment of the function of prosthetic mitral valves.

Authors:  J Chambers; G Jackson; D Jewitt
Journal:  Br Heart J       Date:  1990-03

2.  Development and use of a remote-controlled mitral valve.

Authors:  J S Meisner; S Nikolić; T Tamura; K Tamura; R W Frater; E L Yellin
Journal:  Ann Biomed Eng       Date:  1986       Impact factor: 3.934

3.  Changing trends in cardiac disease and cardiac surgery.

Authors:  D C McGoon
Journal:  Can Anaesth Soc J       Date:  1982-07

4.  Multiple valve replacement with pericardial xenograft. Clinical and haemodynamic study.

Authors:  A P Tandon; W Whitaker; M I Ionescu
Journal:  Br Heart J       Date:  1980-11

5.  Computer-controlled in vitro model of the human left heart.

Authors:  P Verdonck; A Kleven; R Verhoeven; B Angelsen; J Vandenbogaerde
Journal:  Med Biol Eng Comput       Date:  1992-11       Impact factor: 2.602

6.  Ionescu-Shiley pericardial xenograft valve: Hemodynamic evaluation and early clinical follow-up of 326 patients.

Authors:  David A. Ott; Aldemire T. Coelho; Denton A. Cooley; George J. Reul
Journal:  Cardiovasc Dis       Date:  1980-06

7.  SEQUENTIAL HEMODYNAMIC STUDIES OF THE IONESCU-SHILEY PERICARDIAL XENOGRAFT VALVE UP TO SIX YEARS AFTER IMPLANTATION.

Authors:  Anand P. Tandon; Donald R. Smith; Marian I. Ionescu
Journal:  Cardiovasc Dis       Date:  1979-09

8.  Development and characterization of acellular porcine pulmonary valve scaffolds for tissue engineering.

Authors:  Ji Luo; Sotirios A Korossis; Stacy-Paul Wilshaw; Louise M Jennings; John Fisher; Eileen Ingham
Journal:  Tissue Eng Part A       Date:  2014-06-12       Impact factor: 3.845

9.  The effect of decellularisation on the real time mechanical fatigue of porcine aortic heart valve roots.

Authors:  Amisha Desai; Eileen Ingham; Helen E Berry; John Fisher; Louise M Jennings
Journal:  PLoS One       Date:  2022-04-01       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.