C Constantinou, F H Attix, B R Paliwal.
Abstract
The formulation, manufacture and testing of an epoxy resin-based solid substitute for water is presented. This "solid water" has radiation characteristics very close volumetrically to those of water. When it is used as a dosimetry phantom for x- and gamma-ray beams in the radiotherapy range, phantom-to-water corrections and density corrections are eliminated. Relative transmission measurements have shown that the transmission through 10 cm of solid water is within 0.2% of that through an equal thickness of water for x and gamma rays. The use of this material for calibration phantoms can help achieve the goal of radiotherapy beam calibrations within +/- 1.0% of the true dose rate, easier to achieve.Entities:
Mesh:
Substances:
Year: 1982 PMID: 7110074 DOI: 10.1118/1.595063
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071