V S Shaiju1, Rajesh Kumar2, Debjani Phani1, K V Rajasekhar3, George Zacharia1, Saju Bhasi1, Raghuram K Nair1.
Abstract
PURPOSE: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans.Entities:
Keywords: Gamma analysis; head phantom; point dose
Year: 2020 PMID: 32831488 PMCID: PMC7416869 DOI: 10.4103/jmp.JMP_21_20
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Head phantom concept to reality: (a) designed part (b) machining the outer contour of the phantom using a Computer numerical control machine (c) fabricated head phantom, and (d) phantom mounted on linear accelerator couch for dose measurements
Figure 2(a) Individually designed slice three dimensional model and (b) the corresponding machined part
Figure 3(a) 40 mm× 40 mm × 40 mm cuboid three dimensional model with detector insert and (b) the corresponding machined part
Figure 4Cavity insert drawing with external dimension 40 mm × 40 mm × 40 mm for (a) Gel dosimetry (b) radio-chromic film and (c) fabricated optically stimulated luminescence dosimeter insert cuboid
Figure 5Sagittal computed tomography view of the head phantom with detector CC13
Figure 6(a) 40 mm × 40 mm Gafchromic EBT3 film placed in the phantom for irradiation and (b) the exposed film with fiducial marks on the films
Validation results of the head phantom using a CC13 ion chamber
| Point dose measurements using a CC13 ion chamber for VMAT plans | |||
|---|---|---|---|
| Number of patients | TPS dose (cGy) | Measured dose (cGy) | Percentage of variation |
| 1 | 180.90 | 183.86 | 1.61 |
| 2 | 190.70 | 194.43 | 1.92 |
| 3 | 186.50 | 187.14 | 0.34 |
| 4 | 178.40 | 178.87 | 0.26 |
| 5 | 181.40 | 182.32 | 0.50 |
| 6 | 188.40 | 192.03 | 1.89 |
| 7 | 178.50 | 179.94 | 0.80 |
| 8 | 190.00 | 191.11 | 0.58 |
| 9 | 172.40 | 173.87 | 0.85 |
| 10 | 196.20 | 197.60 | 0.71 |
| 11 | 187.30 | 188.74 | 0.76 |
| 12 | 184.10 | 184.90 | 0.43 |
| 13 | 177.40 | 179.81 | 1.34 |
| 14 | 177.70 | 178.33 | 0.35 |
| 15 | 186.40 | 187.67 | 0.68 |
| Mean percentage of deviation | 0.87 | ||
VMAT: Volumetric Modulated Arc Therapy, TPS: Treatment Planning System
Figure 7(a) Transverse plane film image, (b) Transverse unaltered planar dose, (c) combined isodose lines and (d) the corresponding gamma analysis
Gamma analysis statistics of stereotactic radiotherapy patients in the transverse plane
| Serial number of patients | Gamma Analysis (DTA, DD) | |
|---|---|---|
| 3 mm/3% (SD) | 2 mm/2% (SD) | |
| 1 | 99.88±0.15 | 98.89±0.23 |
| 2 | 99.17±0.25 | 93.16±0.38 |
| 3 | 99.73±0.18 | 96.50±0.27 |
| 4 | 99.81±0.19 | 95.38±0.28 |
| 5 | 99.85±0.16 | 96.23±0.24 |
SD: Standard deviation, DD: Dose difference, DTA: Distance to agreement