Literature DB >> 7058746

Analysis of pacemaker malfunction and complications of temporary pacing in the coronary care unit.

J L Austin, L K Preis, R S Crampton, G A Beller, R P Martin.   

Abstract

The medical records of 100 patients who received 113 temporary transvenous pacemakers were reviewed to determine the incidence of complications and malfunction. Malfunction, defined as failure to capture or sense, or both, occurred in 42 (37 percent) of 113 temporary pacemakers. The initial malfunction occurred within 24 hours in 21 (50 percent) and within 48 hours in 36 (86 percent) of the 42 pacemakers. Although the incidence of malfunction was not significantly different for brachial and femoral venous pacing catheters, 7 (37 percent) of 19 brachial venous pacemakers required repositioning or replacement compared with 8 (9 percent) of 91 femoral venous catheters (p = 0.005). Thirty-seven complications occurred in 23 (20 percent) of 113 episodes of pacing; ventricular tachycardia during catheter insertion, fever and phlebitis were the most common complications. No complication resulted in death. The incidence of complications and perforation was greater for brachial than for femoral venous pacemakers (p less than 0.05). Sepsis, local infection and pulmonary embolus occurred only with femoral venous pacemakers. Sepsis, phlebitis and pulmonary embolus were more common with temporary pacemakers in place for 7 hours or longer (p = 0.04). Recognition to the problems peculiar to each pacing catheter site and shortening the duration of pacing should help minimize problems with temporary pacing.

Entities:  

Mesh:

Year:  1982        PMID: 7058746     DOI: 10.1016/0002-9149(82)90505-7

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  17 in total

Review 1.  Temporary cardiac pacing.

Authors:  M D Gammage
Journal:  Heart       Date:  2000-06       Impact factor: 5.994

2.  Accidental mediastinal entry via left internal jugular vein cannulation.

Authors:  T E Albertson; C J Fisher; Z Vera
Journal:  Intensive Care Med       Date:  1985       Impact factor: 17.440

3.  A transient, closed-loop network of wireless, body-integrated devices for autonomous electrotherapy.

Authors:  Yeon Sik Choi; Hyoyoung Jeong; Rose T Yin; Raudel Avila; Anna Pfenniger; Jaeyoung Yoo; Jong Yoon Lee; Andreas Tzavelis; Young Joong Lee; Sheena W Chen; Helen S Knight; Seungyeob Kim; Hak-Young Ahn; Grace Wickerson; Abraham Vázquez-Guardado; Elizabeth Higbee-Dempsey; Bender A Russo; Michael A Napolitano; Timothy J Holleran; Leen Abdul Razzak; Alana N Miniovich; Geumbee Lee; Beth Geist; Brandon Kim; Shuling Han; Jaclyn A Brennan; Kedar Aras; Sung Soo Kwak; Joohee Kim; Emily Alexandria Waters; Xiangxing Yang; Amy Burrell; Keum San Chun; Claire Liu; Changsheng Wu; Alina Y Rwei; Alisha N Spann; Anthony Banks; David Johnson; Zheng Jenny Zhang; Chad R Haney; Sung Hun Jin; Alan Varteres Sahakian; Yonggang Huang; Gregory D Trachiotis; Bradley P Knight; Rishi K Arora; Igor R Efimov; John A Rogers
Journal:  Science       Date:  2022-05-26       Impact factor: 63.714

4.  Morphological changes in human myocardium during permanent pacing.

Authors:  P Châtelain; R Adamec; J N Cox
Journal:  Virchows Arch A Pathol Anat Histopathol       Date:  1985

Review 5.  Cardiac pacing in the accident and emergency department: a review.

Authors:  J Heyworth
Journal:  J R Soc Med       Date:  1988-07       Impact factor: 18.000

6.  A review of temporary cardiac pacing wires.

Authors:  Peter McCann
Journal:  Indian Pacing Electrophysiol J       Date:  2007-01-01

7.  Temporary Endocavitary Pacemakers and their Use and Misuse: the Least is Better.

Authors:  Antoine Kossaify
Journal:  Clin Med Insights Cardiol       Date:  2014-01-08

8.  Pediatric primary cardiac tumors: diagnosis and treatment.

Authors:  Avihu Z Gazit; Sanjiv K Gandhi
Journal:  Curr Treat Options Cardiovasc Med       Date:  2007-10

9.  Fully implantable and bioresorbable cardiac pacemakers without leads or batteries.

Authors:  Yeon Sik Choi; Rose T Yin; Anna Pfenniger; Jahyun Koo; Raudel Avila; K Benjamin Lee; Sheena W Chen; Geumbee Lee; Gang Li; Yun Qiao; Alejandro Murillo-Berlioz; Alexi Kiss; Shuling Han; Seung Min Lee; Chenhang Li; Zhaoqian Xie; Yu-Yu Chen; Amy Burrell; Beth Geist; Hyoyoung Jeong; Joohee Kim; Hong-Joon Yoon; Anthony Banks; Seung-Kyun Kang; Zheng Jenny Zhang; Chad R Haney; Alan Varteres Sahakian; David Johnson; Tatiana Efimova; Yonggang Huang; Gregory D Trachiotis; Bradley P Knight; Rishi K Arora; Igor R Efimov; John A Rogers
Journal:  Nat Biotechnol       Date:  2021-06-28       Impact factor: 68.164

10.  Myocardial Injury After Temporary Transvenous Cardiac Pacing.

Authors:  Meng Liu; Pingsheng Wu
Journal:  Ther Clin Risk Manag       Date:  2021-05-18       Impact factor: 2.423

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.