| Literature DB >> 7025757 |
L L Lembke, R N Kniseley, R C van Nostrand, M D Hale.
Abstract
A method was devised to determine the precision of the all-glass impinger and the Andersen six-stage microbial impactor over a wide range of aerosol concentrations like those found in facilities which process solid waste. Simultaneous samples were collected inside a municipal solid-waste recovery system, and the data were treated statistically to estimate the precision of each air-sampling device. All-glass impingers yielded colony counts which indicated a linear relationship between samplers over an observed aerosol concentration of 1.1 X 10(3) to 2.8 X 10(7) colony-forming units per m3 of air. Impactors also yielded colony counts which indicated a linear relationship over an observed aerosol concentration range of 3.9 X 10(3) to 1.9 X 10(5) colony-forming units per m3 of air. The coefficients of variation for the all-glass impinger and the six-stage impactor in an environment with a high and variable dust level were determined to be 0.38 and 0.23, respectively.Mesh:
Substances:
Year: 1981 PMID: 7025757 PMCID: PMC243994 DOI: 10.1128/aem.42.2.222-225.1981
Source DB: PubMed Journal: Appl Environ Microbiol ISSN: 0099-2240 Impact factor: 4.792