M A Gordon, R A Duncan, L C Kingsley.
Abstract
A soluble-antigen fluorescent-antibody technique for serodiagnosis of toxoplasmosis, employing a special fluorometer (FIAX; International Diagnostic Technology, Santa Clara, Calif.), was evaluated in comparison with a standard indirect fluorescent-antibody method which uses a fluorescence microscope. Titers obtained by the new method on 324 serum samples correlated very well with those of the conventional procedure. There was good intralaboratory and interlaboratory agreement in runs on selected samples. The FIAX method is simple, rapid, and reliable and has the advantage of objectivity.Entities:
Mesh:
Substances:
Year: 1981 PMID: 7009636 PMCID: PMC273778 DOI: 10.1128/jcm.13.2.283-285.1981
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948