Assessment of neonatal ventilator performances.

P Jouvet1, P Hubert, P H Jarreau, F Lofaso, M Cloup, A Harf.   

Abstract

OBJECTIVE: To analyze efficiency and reliability of 4 modern neonatal ventilators under difficult test conditions. The ventilators tested were: Babylog 8000 (Dräger Medical), BP 2001 (Bear Medical Systems), Sechrist IV 100 B (Sechrist Industries), Infant Star (Infrasonics INC). MEASUREMENTS AND
RESULTS: Gas flow generation was tested by comparison of preset flow values with no resistance in the circuit to flow values obtained during interposition of a resistance in the inspiratory circuit. A decrease in gas flow was observed when interposition of a resistance in the inspiratory circuit increased peak inspiratory pressure to 60 cmH2O (gas flow decreased by 8% to 24% depending on the ventilator tested). The pressure limiting valve and the positive end-expiratory pressure valve were also evaluated in order to test their behaviour under different flow conditions. Flow-dependence of the pressure was noted for all ventilators except Babylog 8000. Assessment of the reliability of pressure monitoring revealed either 'under' or 'over' estimation of peak inspiratory pressure and positive end-expiratory pressure depending on the ventilator tested.
CONCLUSION: For the best clinical use of mechanical ventilators, neonatologists should be aware of these limitations. Therefore a regular assessment of ventilator performance and monitoring reliability is recommended.

Entities:  

Mesh:

Year:  1995        PMID: 8847431     DOI: 10.1007/bf01704743

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  11 in total

1.  Test of 20 similar intensive care ventilators in daily use conditions--evaluation of accuracy and performances.

Authors:  L Beydon; N Liu; J Hassapopoulos; F Rua; F Bonnet; A Harf; P Duvaldestin
Journal:  Intensive Care Med       Date:  1992       Impact factor: 17.440

2.  Bench testing of the CPU-1 ventilator.

Authors:  J F Nunn; D J Lyle
Journal:  Br J Anaesth       Date:  1986-06       Impact factor: 9.166

3.  Factors influencing mechanical performance of neonatal high-frequency ventilators.

Authors:  J J Fredberg; G M Glass; B R Boynton; I D Frantz
Journal:  J Appl Physiol (1985)       Date:  1987-06

4.  The physical properties of ventilators in the inspiratory phase.

Authors:  R L Peslin
Journal:  Anesthesiology       Date:  1969-03       Impact factor: 7.892

5.  Flow resistance of exhalation valves and positive end-expiratory pressure devices used in mechanical ventilation.

Authors:  J J Marini; B H Culver; W Kirk
Journal:  Am Rev Respir Dis       Date:  1985-06

Review 6.  Retinopathy of prematurity.

Authors:  J T Flynn
Journal:  Pediatr Clin North Am       Date:  1987-12       Impact factor: 3.278

7.  Some infant ventilators do not limit peak inspiratory pressure reliably during active expiration.

Authors:  H Kirpalani; R Santos-Lyn; R Roberts
Journal:  Crit Care Med       Date:  1988-09       Impact factor: 7.598

8.  Inspiratory work and response times of a modified pediatric volume ventilator during synchronized intermittent mandatory ventilation and pressure support ventilation.

Authors:  L D Martin; J F Rafferty; R C Wetzel; F R Gioia
Journal:  Anesthesiology       Date:  1989-12       Impact factor: 7.892

Review 9.  Principles of neonatal assisted ventilation.

Authors:  W A Carlo; R J Martin
Journal:  Pediatr Clin North Am       Date:  1986-02       Impact factor: 3.278

10.  Methods of assessment and findings regarding pulmonary function in infants less than 1000 grams.

Authors:  M D Cunningham; N S Desai
Journal:  Clin Perinatol       Date:  1986-06       Impact factor: 3.430

View more
  2 in total

1.  Comparison of two different CPAP systems by tidal breathing parameters.

Authors:  Thomas Hückstädt; Bertram Foitzik; Roland R Wauer; Gerd Schmalisch
Journal:  Intensive Care Med       Date:  2003-05-27       Impact factor: 17.440

2.  Simulations for mechanical ventilation in children: review and future prospects.

Authors:  Olivier Flechelles; Annie Ho; Patrice Hernert; Guillaume Emeriaud; Nesrine Zaglam; Farida Cheriet; Philippe A Jouvet
Journal:  Crit Care Res Pract       Date:  2013-03-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.