P A Levine, S S Barold, R D Fletcher, P Talbot.
Abstract
Six cases are presented in which a transient or chronic rise in the stimulation threshold of a permanently implanted unipolar pacemaker resulted in the loss of effective pacing after therapeutic defibrillation or cardioversion. Although damage to the pulse generator may still occur, leading to a loss of function as demonstrated in a seventh patient, improvements in the internal protection circuits of the present generation of pacemakers makes this less likely while possibly predisposing to endocardial burns and increased fibrosis at the electrode-endocardial interface. The theoretical explanations for this phenomenon are discussed, along with recommendations for the prospective and retrospective management of the pacemaker patient who requires defibrillation or cardioversion.Entities:
Mesh:
Year: 1983 PMID: 6853897 DOI: 10.1016/s0735-1097(83)80044-8