Literature DB >> 6738452

Investigation of basic imaging properties in digital radiography. I. Modulation transfer function.

M L Giger, K Doi.   

Abstract

The effect of various digital parameters, such as the sampling aperture, sampling distance, and display aperture, on the modulation transfer function (MTF) of digital radiographic imaging systems was investigated by means of theoretical simulation studies. The MTFs were also determined experimentally to confirm the relationship used in the simulation studies. The results indicate that the overall MTF of a digital system cannot specify the resolution properties in the same way as can the MTFs of analog systems. The MTF of a digital system may include a "false" response due to aliasing, which could lead to an incorrect interpretation of the resolution properties. The magnitude of aliasing that will occur in a digitized signal depends on the sampling parameters chosen and on the frequency content of the radiologic object being imaged. Thus, the type of object to be detected as well as various digital parameters must be considered in the design and evaluation of digital imaging systems.

Mesh:

Year:  1984        PMID: 6738452     DOI: 10.1118/1.595629

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Enhancement of chest images by automatic adaptive spatial filtering.

Authors:  M Souto; J Correa; P G Tahoces; D Tucker; K S Malagari; J J Vidal; R G Fraser
Journal:  J Digit Imaging       Date:  1992-11       Impact factor: 4.056

2.  Image-quality assessment method for digital phase-contrast imaging based on two-dimensional power spectral analysis.

Authors:  Satoru Matsuo; Junji Morishita; Tetsuro Katafuchi; Chika Honda; Hiroshi Fujita
Journal:  Radiol Phys Technol       Date:  2011-11-10

3.  Two K versus 4 K storage phosphor chest radiography: detection performance and image quality.

Authors:  Claus Koelblinger; Mathias Prokop; Michael Weber; Johannes Sailer; Fabiola Cartes-Zumelzu; Cornelia Schaefer-Prokop
Journal:  Eur Radiol       Date:  2007-05-10       Impact factor: 5.315

Review 4.  Computer-aided diagnosis in medical imaging: historical review, current status and future potential.

Authors:  Kunio Doi
Journal:  Comput Med Imaging Graph       Date:  2007-03-08       Impact factor: 4.790

Review 5.  Anniversary paper: History and status of CAD and quantitative image analysis: the role of Medical Physics and AAPM.

Authors:  Maryellen L Giger; Heang-Ping Chan; John Boone
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

6.  Singular value description of a digital radiographic detector: theory and measurements.

Authors:  Iacovos S Kyprianou; Aldo Badano; Brandon D Gallas; Kyle J Myers
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

7.  Effect of readout direction in the edge profile on the modulation transfer function of computed radiographic systems by use of the edge method.

Authors:  Nobukazu Tanaka; Junji Morishita; Norisato Tsuda; Masafumi Ohki
Journal:  Radiol Phys Technol       Date:  2013-06-04

8.  Application of a variable filter for presampled modulation transfer function analysis with the edge method.

Authors:  Ryo Higashide; Katsuhiro Ichikawa; Hiroshi Kunitomo; Kazuya Ohashi
Journal:  Radiol Phys Technol       Date:  2015-06-19

9.  Medical display application for degraded image sharpness restoration based on the modulation transfer function: initial assessment for a five-megapixel mammography display monitor.

Authors:  Shogo Tokurei; Yoichiro Ikushima; Kazuki Takegami; Munemasa Okada; Junji Morishita
Journal:  Phys Eng Sci Med       Date:  2021-05-17

10.  The effects of x-ray beam hardening on detective quantum efficiency and radiation dose.

Authors:  Molly Donovan Wong; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2011       Impact factor: 1.535

View more