Literature DB >> 6721952

Late strut fracture in a Björk-Shiley valve prosthesis (current series).

S H Sacks, A D Northeast, J Watkins.   

Abstract

Fracture of the short strut of the latest series of Björk-Shiley mitral valve prostheses occurred 12 months after implantation. The patient died despite emergency surgery to excise and replace the prosthesis. Björk-Shiley 29-33 mm valves with a 60 degree opening angle manufactured between February 1981 and March 1982 have a predicted strut fracture rate of 0.3% (three times greater than that of valves produced outside these dates). Unimplanted valves from this batch have now been recalled by the manufacturers.

Entities:  

Mesh:

Year:  1984        PMID: 6721952      PMCID: PMC481552          DOI: 10.1136/hrt.51.5.578

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  4 in total

1.  Early disc dislodgment. An unusual complication after insertion of a Björk-Shiley mitral valve prosthesis.

Authors:  B J Messmer; M Rothlin; A Senning
Journal:  J Thorac Cardiovasc Surg       Date:  1973-03       Impact factor: 5.209

2.  Strut fracture in the new Bjørk-Shiley mitral valve prosthesis.

Authors:  O Brubakk; S Simonsen; L Källman; A Fredriksen
Journal:  Thorac Cardiovasc Surg       Date:  1981-04       Impact factor: 1.827

3.  Fracture and embolization of a Björk-Shiley disc. Fatal failure of a prosthetic mitral valve.

Authors:  D D Norenberg; R W Evans; A E Gundersen; R M Abellera
Journal:  J Thorac Cardiovasc Surg       Date:  1977-12       Impact factor: 5.209

4.  Strut fracture and disc embolization of a Björk-Shiley mitral valve prosthesis: localization of embolized disc by computerized axial tomography.

Authors:  A J Larrieu; E Puglia; P Allen
Journal:  Ann Thorac Surg       Date:  1982-08       Impact factor: 4.330

  4 in total
  3 in total

1.  Metallurgical analysis of failed Björk-Shiley cardiac valve prostheses.

Authors:  S H Sacks; M Harrison; P J Bischler; J W Martin; J Watkins; A Gunning
Journal:  Thorax       Date:  1986-02       Impact factor: 9.139

2.  Early clinical evaluation of a new Björk-Shiley valve prosthesis with integral monostrut.

Authors:  S Nakano; H Hirose; H Matsuda; K Taniguchi; T Kawamoto; Y Kawashima
Journal:  Heart Vessels       Date:  1985-11       Impact factor: 2.037

3.  Mechanism of late failure of the Alvarez disc valve prosthesis.

Authors:  J L Gibbs; G A Davies; A Schofield; G A Wharton; D A Watson; L M Gerlis
Journal:  Br Heart J       Date:  1985-05
  3 in total