Literature DB >> 3994864

Mechanism of late failure of the Alvarez disc valve prosthesis.

J L Gibbs, G A Davies, A Schofield, G A Wharton, D A Watson, L M Gerlis.   

Abstract

An Alvarez disc valve prosthesis had been implanted in the mitral and tricuspid positions in 15 patients between 1964 and 1967 in this hospital. Of these 15 patients, only three survived longer than 10 years, all three dying unexpectedly between 16 and 18 years postoperatively. Two deaths were a direct result of disruption of the Alvarez prosthesis. The third patient was not examined at necropsy, but in retrospect the history was highly suggestive of valve malfunction. Cross sectional and pulsed Doppler echocardiography made valuable contributions to the diagnosis in both cases of valve disruption and, when available, should be part of the primary investigation of suspected prosthetic heart valve failure. The Alvarez valve prosthesis is liable to erosive wear which may cause late detachment of the disc from the valve ring with abrupt and catastrophic haemodynamic consequences.

Entities:  

Mesh:

Year:  1985        PMID: 3994864      PMCID: PMC481801          DOI: 10.1136/hrt.53.5.510

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  23 in total

1.  Failure of Beal mitral valve prosthesis. Clinical and radiographic features.

Authors:  R G Jost; R C McKnight; C L Roper
Journal:  J Thorac Cardiovasc Surg       Date:  1975-07       Impact factor: 5.209

2.  Aortic valve replacement with the De Bakey valve.

Authors:  B C Paton; M B Pine
Journal:  J Thorac Cardiovasc Surg       Date:  1976-10       Impact factor: 5.209

3.  Acute massive mitral regurgitation from prosthetic valve dysfunction.

Authors:  D K Cooper; M F Sturridge
Journal:  Br Heart J       Date:  1976-07

4.  Survival following fracture of strut from mitral prosthesis with disc translocation.

Authors:  M T McEnany; E O Wheeler; W G Austen
Journal:  J Thorac Cardiovasc Surg       Date:  1979-07       Impact factor: 5.209

5.  Double opening sound related to dysfunction in a caged disc mitral valve prosthesis.

Authors:  E C Harrison; E J Roschke; M Matsuno; J W Allen
Journal:  Am J Cardiol       Date:  1974-02       Impact factor: 2.778

6.  Fracture and embolization of a Björk-Shiley disc. Fatal failure of a prosthetic mitral valve.

Authors:  D D Norenberg; R W Evans; A E Gundersen; R M Abellera
Journal:  J Thorac Cardiovasc Surg       Date:  1977-12       Impact factor: 5.209

7.  An ultrasound Doppler technique for the noninvasive determination of the pressure gradient in the Björk-Shiley mitral valve.

Authors:  J Holen; S Simonsen; T Frøysaker
Journal:  Circulation       Date:  1979-03       Impact factor: 29.690

8.  Durability of prosthetic heart valves.

Authors:  R E Clark; W M Swanson; J L Kardos; R W Hagen; R A Beauchamp
Journal:  Ann Thorac Surg       Date:  1978-10       Impact factor: 4.330

9.  Wear in Björk-Shiley heart valve prostheses recovered at necropsy or operation.

Authors:  M D Silver
Journal:  J Thorac Cardiovasc Surg       Date:  1980-05       Impact factor: 5.209

10.  Ball variance and fracture of mitral valve prosthesis causing recurrent thromboemboli.

Authors:  L D Joyce; R W Emery; D M Nicoloff
Journal:  J Thorac Cardiovasc Surg       Date:  1978-02       Impact factor: 5.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.