Strut fracture in the new Bjørk-Shiley mitral valve prosthesis.

O Brubakk, S Simonsen, L Källman, A Fredriksen.   

Abstract

The case of a patient with the new type Bjørk-Shiley aortic and mitral valve prosthesis is described. Three months after implant she suffered acute heart failure and died. Post-mortem examination revealed a fractured outlet strut in the mitral valve prosthesis with dislocation of the disc. The fracture was regarded as due to excessive brittleness caused by demonstrated deposition of chromium-tungsten-carbide.

Entities:  

Mesh:

Year:  1981        PMID: 6166068     DOI: 10.1055/s-2007-1023454

Source DB:  PubMed          Journal:  Thorac Cardiovasc Surg        ISSN: 0171-6425            Impact factor:   1.827


  5 in total

1.  Outlet-strut fracture and disc embolization in 27-mm and smaller Björk-Shiley convexo-concave valves: historical overview and 3 case reports.

Authors:  C G Montero; R Burgos; J E Rodriguez; D Figuera
Journal:  Tex Heart Inst J       Date:  1989

2.  Discrimination in vitro between the acoustic emissions from Bjork-Shiley convexo-concave valves with and without a broken minor strut.

Authors:  D K Walker; L N Scotten
Journal:  Med Biol Eng Comput       Date:  1991-09       Impact factor: 2.602

3.  A case of strut fracture of the new Björk-Shiley mitral valve prosthesis.

Authors:  J Ninomiya; N Yamate; T Mizutani; R Soga; T Shoji
Journal:  Jpn J Surg       Date:  1983-09

4.  Late strut fracture in a Björk-Shiley valve prosthesis (current series).

Authors:  S H Sacks; A D Northeast; J Watkins
Journal:  Br Heart J       Date:  1984-05

5.  Twenty-five years of heart valve substitutes: a review.

Authors:  M M Black; P J Drury; W B Tindale
Journal:  J R Soc Med       Date:  1983-08       Impact factor: 18.000

  5 in total