Long-term follow-up of the Ionescu-Shiley mitral pericardial xenograft.

S Gabbay, U Bortolotti, F Wasserman, N Tindel, S M Factor, R W Frater.   

Abstract

In a group of 40 patients discharged from the hospital after mitral valve replacement with an Ionescu-Shiley pericardial xenograft from January, 1977, to December, 1980, seven instances of valve failure occurred. This unprecedented incidence of bioprosthetic dysfunction after mitral valve replacement with the Ionescu-Shiley pericardial xenograft led us to update the follow-up of our patients receiving this valve in the mitral position. The results of our survey showed, at 6 years postoperatively, an actuarial survival rate of 72%, an actuarial probability of being free from emboli of 62%, and an actuarial freedom from prosthetic failure of 60%. The explanted valves showed complete absence of the neoendothelial lining of the Dacron-covered frame and lesions resembling those observed in Ionescu-Shiley pericardial xenografts removed from our fatigue test system after a mean of 29 +/- 17 X 10(6) cycles. This similarity prompted a classification of tears occurring in the mitral Ionescu-Shiley pericardial xenograft, which is of extreme importance, since the clinical presentation and outcome of patients with a failing valve differs according to the type of rupture. The lesions of the cusp observed in clinical specimens were possibly related to the continuous trauma of the tissue against the bare Dacron cloth during closure of the valve. It is concluded that (1) actual durability and thrombogenicity of the Ionescu-Shiley pericardial xenograft in the mitral position needs to be carefully reassessed, (2) close follow-up of such patients by clinical and two-dimensional echocardiographic evaluation is advisable after the third postoperative year, and (3) failure of the mitral Ionescu-Shiley pericardial xenograft may occur suddenly, and awareness of this complication is the clue to prompt recognition and treatment of such patients.

Entities:  

Mesh:

Year:  1984        PMID: 6492841

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  11 in total

Review 1.  A look at recent improvements in the durability of tissue valves.

Authors:  Takahiro Nishida; Ryuji Tominaga
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-01-24

2.  Comparison of transcatheter aortic valve and surgical bioprosthetic valve durability: A fatigue simulation study.

Authors:  Caitlin Martin; Wei Sun
Journal:  J Biomech       Date:  2015-08-07       Impact factor: 2.712

3.  Ionescu-Shiley bovine pericardial bioprostheses. Histologic and ultrastructural studies.

Authors:  S L Hilbert; V J Ferrans; H A McAllister; D A Cooley
Journal:  Am J Pathol       Date:  1992-05       Impact factor: 4.307

4.  Mitral replacement: clinical experience with a ball-valve prosthesis. Twenty-five years later.

Authors:  A Cobanoglu; G L Grunkemeier; G M Aru; C L McKinley; A Starr
Journal:  Ann Surg       Date:  1985-09       Impact factor: 12.969

5.  Simulation of long-term fatigue damage in bioprosthetic heart valves: effects of leaflet and stent elastic properties.

Authors:  Caitlin Martin; Wei Sun
Journal:  Biomech Model Mechanobiol       Date:  2013-10-04

6.  An analysis of autopsy findings in 108 patients who died after valve replacement.

Authors:  H Ishibashi-Ueda; M Imakita; M Katsuragi; H Fujita; H Hao; C Yutani
Journal:  Virchows Arch A Pathol Anat Histopathol       Date:  1993

Review 7.  Biomechanical Behavior of Bioprosthetic Heart Valve Heterograft Tissues: Characterization, Simulation, and Performance.

Authors:  Joao S Soares; Kristen R Feaver; Will Zhang; David Kamensky; Ankush Aggarwal; Michael S Sacks
Journal:  Cardiovasc Eng Technol       Date:  2016-08-09       Impact factor: 2.495

8.  Early-phase events with the mitroflow pericardial valve.

Authors:  L B McGrath; L Gonzalez-Lavin
Journal:  Tex Heart Inst J       Date:  1988

9.  Total ellipse of the heart valve: the impact of eccentric stent distortion on the regional dynamic deformation of pericardial tissue leaflets of a transcatheter aortic valve replacement.

Authors:  Paul S Gunning; Neelakantan Saikrishnan; Ajit P Yoganathan; Laoise M McNamara
Journal:  J R Soc Interface       Date:  2015-12-06       Impact factor: 4.118

Review 10.  Fatigue damage of collagenous tissues: experiment, modeling and simulation studies.

Authors:  Caitlin Martin; Wei Sun
Journal:  J Long Term Eff Med Implants       Date:  2015
View more