| Literature DB >> 63463 |
M M Salpeter, H C Fertuck, E E Salpeter.
Abstract
Resolution for 125I-labeled specimens under electron microscope (EM) autoradiographic conditions was assessed experimentally. With this isotope the size of the silver halide crystal was the most important resolution-limiting factor. Heavy metal staining such as is routinely used in preparing animal tissues for EM autoradiography produced an improvement in resolution of approximately 15-20%. For a 500-1,000-A biological tissue section fixed with OsO4 and stained with uranyl acetate, we obtained resolution (half distance, HD) values of approximately 800 +/- 120 A using Ilford L4 emulsion and 500 +/- 70 A using a Kodak NTE-type emulsion. General aspects of resolution-limiting factors and comparison with 3H and 14C values are discussed.Entities:
Mesh:
Substances:
Year: 1977 PMID: 63463 PMCID: PMC2110981 DOI: 10.1083/jcb.72.1.161
Source DB: PubMed Journal: J Cell Biol ISSN: 0021-9525 Impact factor: 10.539