Abstract
The performance of a fiberoptic probe colorimeter (model PC800; Brinkmann Instruments, Inc., Westbury, N.Y.) for quantitating enzymatic or colorimetric assays in 96-well microtiter plates was compared with the performances of a spectrophotometer (model 240; Gilford Instrument Laboratories, Inc., Oberlin, Ohio) and a commercially available enzyme immunoassay reader (model MR590; Dynatech Laboratories, Inc., Alexandria, Va.). Alkaline phosphatase-p-nitrophenyl phosphate in 3 M NaOH was used as the chromophore source. Six types of plates were evaluated for use with the probe colorimeter; they generated reproducibility values (100% coefficient of variation) ranging from 91 to 98% when one individual made 24 independent measurements on the same dilution of chromophore on each plate. Eleven individuals each performed 24 measurements with the colorimeter on either a visually light (absorbance of 0.10 at 420 nm) or a dark (absorbance of 0.80 at 420 nm) dilution of chromophore; reproducibilities averaged 87% for the light dilution and 97% for the dark dilution. When one individual measured the same chromophore sample at least 20 times in the colorimeter, in the spectrophotometer or in the enzyme immunoassay reader, reproducibility for each instrument was greater than 99%. Measurements of a dilution series of chromophore in a fixed volume indicated that the optical responses of each instrument were linear in a range of 0.05 to 1.10 absorbance units.Entities:
Mesh:
Year: 1983 PMID: 6341399 PMCID: PMC272656 DOI: 10.1128/jcm.17.3.410-413.1983