V Fiacco, Y J Bryson, D A Bruckner.
Abstract
A fluorescein isothiocyanate-labeled monoclonal cytomegalovirus antibody reagent incorporating Evans blue (Syva/Genetic Systems, Palo Alto, Calif.) was compared to a fluorescein isothiocyanate-labeled polyclonal antibody (Dynatech Diagnostics, Inc., South Windham, Maine) for identification of cytomegalovirus in tissue culture. A total of 191 cultures, 86 positive and 105 negative for cytomegalovirus by cytopathic effect in WI38 cells, were stained in parallel and examined by fluorescent microscopy. No false-negative results were noted with either reagent; however, the polyclonal antibody-stained slides required the use of a separate counterstain to decrease nonspecific fluorescence. The polyclonal reagent gave a false-positive reaction with one varicella-zoster virus isolate and four simian cytomegalovirus isolates. No false-positive results were noted with the monoclonal reagent.Entities:
Mesh:
Substances:
Year: 1984 PMID: 6088576 PMCID: PMC271217 DOI: 10.1128/jcm.19.6.928-930.1984