Literature DB >> 557503

Processing urinary endoscopes in a low-temperature steam and formaldehyde autoclave.

G L Gibson.   

Abstract

Methods of disinfection and sterilisation of urinary endoscopes are considered. A small mobile low-temperature steam and formaldehyde autoclave (Miniclave 80) is evaluated and shown to be satisfactory for this purpose as judged by a variety of relevant microbiological test pieces.

Entities:  

Mesh:

Substances:

Year:  1977        PMID: 557503      PMCID: PMC476371          DOI: 10.1136/jcp.30.3.269

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  6 in total

1.  Disinfection of woollen blankets in steam at subatmospheric pressure.

Authors:  V G ALDER; W A GILLESPIE
Journal:  J Clin Pathol       Date:  1961-09       Impact factor: 3.411

2.  Testing a steam-formaldehyde sterilizer for gas penetration efficiency.

Authors:  S J Line; J K Pickerill
Journal:  J Clin Pathol       Date:  1973-09       Impact factor: 3.411

3.  Disinfection of cystoscopes by subatmospheric steam and steam and formaldehyde at 80 degrees C.

Authors:  V G Alder; J C Gingell; J P Mitchell
Journal:  Br Med J       Date:  1971-09-18

4.  The disinfection of urological endoscopes.

Authors:  J P Mitchell; V G Alder
Journal:  Br J Urol       Date:  1975-10

5.  Disinfection of heat-sensitive material by low-temperature steam and formaldehyde.

Authors:  V G Alder; A M Brown; W A Gillespie
Journal:  J Clin Pathol       Date:  1966-01       Impact factor: 3.411

6.  Residual formaldehyde after low-temperature steam and formaldehyde sterilization.

Authors:  G L Gibson; H P Johnston; V E Turkington
Journal:  J Clin Pathol       Date:  1968-11       Impact factor: 3.411

  6 in total
  1 in total

1.  Measurement of formaldehyde concentrations in a subatmospheric steam-formaldehyde autoclave.

Authors:  D Marcos; D Wiseman
Journal:  J Clin Pathol       Date:  1979-06       Impact factor: 3.411