Literature DB >> 3963334

The Nellcor N-101 pulse oximeter. A clinical evaluation in anaesthesia and intensive care.

J A Tytler, H F Seeley.   

Abstract

The accuracy of the Nellcor N-101 pulse oximeter has been evaluated in adult patients receiving general anaesthesia or intensive care. Readings obtained noninvasively with this instrument were compared with measurements made on arterial blood using a Radiometer OSM2 oximeter. The pulse oximeter was easy to use and within the range tested (70-100 percent saturation of haemoglobin with oxygen) the readings were within I digit of the values obtained by in vitro measurement.

Entities:  

Mesh:

Year:  1986        PMID: 3963334     DOI: 10.1111/j.1365-2044.1986.tb12793.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955