Tissue hypoxia distal to a Peñaz finger blood pressure cuff.

J S Gravenstein, D A Paulus, J Feldman, G McLaughlin.   

Abstract

The Peñaz finger method to measure blood pressure uses a finger cuff in which the pressure level fluctuates in the vicinity of the mean arterial pressure level and thereby interferes with the circulation of blood to and from the fingertip. We measured capillary blood gases and saturation of hemoglobin in the finger during Peñaz finger blood pressure (PFBP) monitoring to assess the degree to which it impairs circulation in the fingertip. Within 2.5 minutes after initiating PFBP monitoring, capillary oxygen tension (PO2) had decreased significantly, from about 71 mm Hg to between 49 and 58 mm Hg for up to 50 minutes. These changes were quite different from those occurring when an occlusive tourniquet was applied around the finger. Within 10 minutes of tourniquet application, acidosis (pH 7.25), hypercapnia (carbon dioxide tension, 59.0 mm Hg), and hypoxemia (PO2, 29 mm Hg) resulted. Within 30 seconds of releasing the PFBP cuff, capillary blood gas values were back to normal. Interspersing 30-second rest periods every 5 minutes during 35 minutes of PFBP monitoring actually decreased capillary oxygen values compared with monitoring without such rest periods. A finger pulse oximeter distal to the PFBP cuff showed desaturation from an average of 97% to 93.7%, with much variability. However, desaturation was statistically significant within 1 minute of application of the PFBP cuff. Within 1 minute the finger volume increased an average of 0.05 ml. After 1 minute the volumes varied widely and, on the average, returned to normal despite continued PFBP monitoring.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 3831251     DOI: 10.1007/bf02832199

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  3 in total

1.  Carbon dioxide tensions in nonarterialised capillary and arterial blood during anaesthesia.

Authors:  E J KNUDSEN; P HANSEN
Journal:  Acta Anaesthesiol Scand       Date:  1962       Impact factor: 2.105

2.  The accuracy of capillary sampling for acid-base estimations.

Authors:  K Docrat; S Kenny
Journal:  Br J Anaesth       Date:  1965-11       Impact factor: 9.166

3.  Evaluation of two prototype devices producing noninvasive, pulsatile, calibrated blood pressure measurement from a finger.

Authors:  N T Smith; K H Wesseling; B de Wit
Journal:  J Clin Monit       Date:  1985-01
  3 in total
  3 in total

1.  On the reliability of the Penaz cuff during systemic and local fingertip vasodilatation at rest and in exercise.

Authors:  W Hildebrandt; H Schütze; J Stegemann
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1991

2.  Continuous, real-time, noninvasive monitor of blood pressure: Penaz methodology applied to the finger.

Authors:  R D Boehmer
Journal:  J Clin Monit       Date:  1987-10

3.  Method specificity of non-invasive blood pressure measurement: oscillometry and finger pulse pressure vs acoustic methods.

Authors:  C De Mey; V Schroeter; R Butzer; S Roll; G G Belz
Journal:  Br J Clin Pharmacol       Date:  1995-10       Impact factor: 4.335

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.