Contrast-detail assessment of computed tomography scanners.

K Faulkner, B M Moores.   

Abstract

The design and construction of contrast-detail detectability phantoms for the assessment of the imaging performance of CT scanners are described. These phantoms have been employed to undertake a series of measurements on a number of different machines. The results are expressed in terms of object contrast which is consistent with the conventional definition of radiological contrast. For one particular scanner, the imaging performance of a number of image reconstruction and post-processing filters has been investigated. The results indicate that there was no observed energy dependence of contrast-detail detectability once differences in technique factors and attenuation were taken into account. A similar conclusion was drawn from the results of a set of measurements made to examine the effect on detectability of additional filtration.

Mesh:

Year:  1986        PMID: 3774879     DOI: 10.1088/0031-9155/31/9/005

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

Review 1.  Radiologic and near-infrared/optical spectroscopic imaging: where is the synergy?

Authors:  Brian W Pogue; Frederic Leblond; Venkataramanan Krishnaswamy; Keith D Paulsen
Journal:  AJR Am J Roentgenol       Date:  2010-08       Impact factor: 3.959

2.  Observer detection limits for a dedicated SPECT breast imaging system.

Authors:  S J Cutler; K L Perez; H X Barnhart; M P Tornai
Journal:  Phys Med Biol       Date:  2010-03-12       Impact factor: 3.609

3.  Characterization of atherosclerotic plaque: a contrast-detail study using multidetector and cone-beam computed tomography.

Authors:  Nima Kasraie; Peter Mah; Carl R Keener; Geoffrey D Clarke
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.