Arterial pressure measurement in preterm infants.

G Briassoulis.   

Abstract

We compared automatic oscillometric measurements of systolic, diastolic, and mean arterial pressures with simultaneous pressures measured from arterial catheters in six preterm infants. Oscillometric arterial pressure measurements were performed with the Omega automatic instrument, using the recommended cuff and the size-larger cuff. Despite highly significant overall correlation coefficients for oscillometric and catheter measurements, 29% to 42% of oscillometric measurements with both cuff sizes were outside the 5-mm Hg error limits, and 4.0% to 9.3% were more than 10 mm Hg different from catheter measurements. Changing the cuff size did not reduce inherent errors in the oscillometric method.

Entities:  

Mesh:

Year:  1986        PMID: 3720331     DOI: 10.1097/00003246-198608000-00014

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  4 in total

1.  Accuracy of oscillometric blood pressure measurement in critically ill neonates with reference to the arterial pressure wave shape.

Authors:  M Gevers; H R van Genderingen; H N Lafeber; W W Hack
Journal:  Intensive Care Med       Date:  1996-03       Impact factor: 17.440

2.  Systolic blood pressure in babies of less than 32 weeks gestation in the first year of life. Northern Neonatal Nursing Initiative.

Authors: 
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-01       Impact factor: 5.747

Review 3.  Hemodynamic dysfunction in neonatal sepsis.

Authors:  Ashraf Kharrat; Amish Jain
Journal:  Pediatr Res       Date:  2021-11-24       Impact factor: 3.756

4.  Use of pulse oximetry for blood pressure measurement after cardiac surgery.

Authors:  A J Movius; S L Bratton; G K Sorensen
Journal:  Arch Dis Child       Date:  1998-05       Impact factor: 3.791

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.