Effect of temperature on the cyclic fatigue resistance of thermally treated reciprocating instruments.

Michel Espinosa Klymus1, Murilo Priori Alcalde2, Rodrigo Ricci Vivan1, Marcus Vinicius Reis Só3, Bruno Carvalho de Vasconselos4, Marco Antonio Hungaro Duarte1.   

Abstract

OBJECTIVES: The aim of this study was to evaluate the impact of body temperature on the cyclic fatigue resistance of different NiTi alloys used for the manufacturing of Reciproc Blue R25 (RB 25.08; VDW, Munich, Germany), X1 Blue File 25 (X1 25.06; MK Life Medical and Dental Products, Porto Alegre, Brazil) and WaveOne Gold Primary (WOG 25.07; Dentsply Maillefer, Ballaigues, Switzerland).
MATERIALS AND METHODS: Sixty instruments of the RB 25.08, X1 25.06 and WOG 25.07 systems were used (n = 20). Cyclic fatigue tests were performed at room temperature (20° ± 1 °C) and at body temperature (37° ± 1 °C). The instruments were reciprocated until fracture occurred in an artificial stainless steel canal with a 60° angle and a 5-mm radius of curvature. The time to fracture (TTF) was recorded. Also, the number of cycles to fracture (NCF) was calculated. Data were analysed using one-way ANOVA and Tukey's tests for inter-group comparison at both temperatures and for the reduction of cyclic fatigue at body temperature. For intra-group comparison at the different temperatures, the unpaired t test was used.
RESULTS: The cyclic fatigue test at 20 °C showed that RB 25.08 and X1 25.06 presented significantly higher TTF and NCF than WOG 25.07 (P < 0.05). At 37 °C, all groups presented significant reduction of TTF and NCF (P < 0.05). RB 25.08 presented significant higher TTF than WOG 25.07 (P < 0.05). Regarding the NCF, there was no significant difference among the groups (P > 0.05). The WOG 25.07 presented the lowest percentage reduction of cyclic fatigue (P < 0.05).
CONCLUSION: The body temperature treatment caused a marked reduction of the cyclic fatigue resistance for all reciprocating instruments tested. The RB 25.08 and X1 25.06 systems presented similar results at both temperatures tested. However, WOG 25.07 presented the lowest percentage reduction in fatigue resistance at body temperature. CLINICAL RELEVANCE: The cyclic fatigue resistance of NiTi reciprocating instruments has been evaluated at room temperature. However, the fatigue resistance significantly decreases upon exposure to body temperature, which could affect the mechanical behaviour of the NiTi instruments during root canal preparation.

Entities:  

Keywords:  Body temperature; Cyclic fatigue; Nickel titanium; Reciprocating systems

Mesh:

Substances:

Year:  2018        PMID: 30397733     DOI: 10.1007/s00784-018-2718-1

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  7 in total

1.  Comparison of the cyclic fatigue resistance of VDW.ROTATE, TruNatomy, 2Shape, and HyFlex CM nickel-titanium rotary files at body temperature.

Authors:  Mustafa Gündoğar; Gülşah Uslu; Taha Özyürek; Gianluca Plotino
Journal:  Restor Dent Endod       Date:  2020-06-22

2.  Differences in cyclic fatigue lifespan between two different heat treated NiTi endodontic rotary instruments: WaveOne Gold vs EdgeOne Fire.

Authors:  Gianluca Gambarini; Massimo Galli; Dario Di Nardo; Marco Seracchiani; Orlando Donfrancesco; Luca Testarelli
Journal:  J Clin Exp Dent       Date:  2019-07-01

3.  Mechanical Properties of ProTaper Gold, EdgeTaper Platinum, Flex Gold and Pro-T Rotary Systems.

Authors:  Murilo Alcalde; Marco Antonio Hungaro Duarte; Pablo Andrés Amoroso Silva; Pedro Henrique Souza Calefi; Emmanuel Silva; Jussaro Duque; Rodrigo Vivan
Journal:  Eur Endod J       Date:  2020-12

4.  Efficacy of reciprocating instruments and final irrigant activation protocols on retreatment of mesiobuccal roots of maxillary molars: a micro-CT analysis.

Authors:  Lilian Tietz; Renan Diego Furlan; Ricardo Abreu da Rosa; Marco Antonio Hungaro Duarte; Murilo Priori Alcalde; Rodrigo Ricci Vivan; Theodoro Weissheimer; Marcus Vinicius Reis Só
Journal:  Restor Dent Endod       Date:  2022-02-15

5.  Cyclic Fatigue Resistance of Nickel-Titanium Rotary Instruments after Simulated Clinical Use.

Authors:  Hana R Ubaed; Diyar Kh Bakr
Journal:  Appl Bionics Biomech       Date:  2022-10-11       Impact factor: 1.664

6.  Influence of NiTi Wire Diameter on Cyclic and Torsional Fatigue Resistance of Different Heat-Treated Endodontic Instruments.

Authors:  Eugenio Pedullà; Francesco Saverio Canova; Giusy Rita Maria La Rosa; Alfred Naaman; Franck Diemer; Luigi Generali; Walid Nehme
Journal:  Materials (Basel)       Date:  2022-09-22       Impact factor: 3.748

Review 7.  Effect of body temperature on the cyclic fatigue resistance of the nickel-titanium endodontic instruments: A systematic review and meta-analysis of in vitro studies.

Authors:  Selventhra Savitha; Sidhartha Sharma; Vijay Kumar; Amrita Chawla; Perumal Vanamail; Ajay Logani
Journal:  J Conserv Dent       Date:  2022-08-02
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.