Alef Vermudt1, Milton Carlos Kuga2, João Felipe Besegato2, Eliane Cristina Gulin de Oliveira2, Thaís Piragine Leandrin2, Marcus Vinicius Reis Só3, João Carlos Silos Moraes4, Jefferson Ricardo Pereira1.
Abstract
Resin-based composites (RBCs) have transformed restorative dentistry and its procedures. However, the characteristics of RBCs have been modified over the years to enhance the physical and chemical properties of the materials. This context raises the need for studies that evaluate whether the properties of the RBCs that are commercially available are clinically adequate with different curing modes. This study aimed to evaluate the mechanical behavior of commercial RBCs after undergoing different curing modes. Twenty-three RBCs of different classes were evaluated. For curing the specimens, a microwave (BMS45, Brastemp) (for 3 min at 450 W) and three LED units were used: an Emitter A Fit (Schuster (second generation)) (light-curing for 15 s with an irradiance of 1250 mW/cm2), VALO (Ultradent (third generation)) (light-curing for 15 s with an irradiance of 1100 mW/cm2), and Emitter Now Duo (Schuster (second generation)) (light-curing for 15 s with an irradiance of 1100 mW/cm2). A total of 670 RBC specimens of 8 mm in diameter and 1 mm in depth were obtained. Afterward, a biaxial flexure strength test was performed until the failure of the specimens, using a universal testing machine set at a speed of 0.5 mm/min. The same specimens were subjected to infrared spectroscopy for evaluating the degree of conversion. Tukey's test was used for multiple comparisons at a significance level of 5%. The light-curing mode did not affect the flexure strength of the RBCs (p > 0.05), but the type and shade of RBCs did so (p < 0.05). In conclusion, the type of RBC directly interferes with the mechanical behavior of the material. However, the curing modes within the same RBC did not change the mechanical properties.Entities:
Keywords: LED source; curing; degree of conversion; flexure strength; resin composites
Year: 2022 PMID: 36235967 PMCID: PMC9571923 DOI: 10.3390/polym14194020
Source DB: PubMed Journal: Polymers (Basel) ISSN: 2073-4360 Impact factor: 4.967
Resin-based composites (RBC) used in this study.
| RBC Trademark | Classification | Manufacturer |
|---|---|---|
| Charisma | Microhybrid | Heraeus Kulzer |
| Z100 | Microhybrid | 3M ESPE |
| Z250 | Microhybrid | 3M ESPE |
| Polofil Supra | Microhybrid | VOCO |
| Empress Direct | Nanohybrid | Ivoclar Vivadent |
| Harmonize | Nanohybrid | Kerr |
| Herculite Précis | Nanohybrid | Kerr |
| Amaris | Nanohybrid | VOCO |
| Admira Fusion | Nanohybrid | VOCO |
| Grandio | Nanohybrid | VOCO |
| GrandioSO | Nanohybrid | VOCO |
| Vittra | Nanofilled | FGM |
| Z350 | Nanofilled | 3M ESPE |
| Palfique LX5 | Supra-nanofilled | Tokuyama |
Sample size (n = 10, following other studies [4,7,13]) according to the shade of RBC and the curing mode.
| RBC Trademarks | Shade A2 | Additional Curing | Effects Shade | ||||
|---|---|---|---|---|---|---|---|
| VALO | Emitter | Emitter Now Duo | Emitter a Fit | VALO | Emitter | Emitter Now Duo | |
| Group 1 | Group 2 | Group 3 | Group 4 | Group 5 | Group 6 | Group 7 | |
| Charisma | 10 | 10 | - | - | - | - | - |
| Z100 | 10 | 10 | 10 | 10 | - | - | - |
| Z250 | 10 | 10 | - | - | - | - | - |
| Polofil Supra | 10 | 10 | - | - | - | - | - |
| Empress Direct | 10 | 10 | - | - | 10 | 10 | 10 |
| Harmonize | 10 | 10 | 10 | 10 | - | - | - |
| Herculite Précis | 10 | 10 | - | - | 10 | 10 | 10 |
| Amaris | 10 | 10 | - | - | 10 | 10 | 10 |
| Admira Fusion | 10 | 10 | - | - | 10 | 10 | 10 |
| Grandio | 10 | 10 | 10 | 10 | 20 | 20 | 20 |
| GrandioSO | 10 | 10 | - | - | 10 | 10 | 10 |
| Vittra | 10 | 10 | 10 | 10 | 10 | 10 | 10 |
| Z350 | 10 | 10 | 10 | 10 | 10 | 10 | - |
| Palfique LX5 | 10 | 10 | - | - | 10 | 10 | 10 |
| TOTAL | 140 | 140 | 50 | 50 | 100 | 100 | 90 |
“-” means that there were no specimens tested.
Figure 1Scheme of the Teflon matrix used to obtain the discs of resin-based composites.
Photopolymerization units, classification, and wavelength.
| Photopolymerization Units | Classification | Wavelength |
|---|---|---|
| VALO | Third generation | 395–480 nm |
| Emitter Now Duo | Third generation | 385–515 nm |
| Emitter Fit A | Second generation | 420–480 nm |
Figure 2Scheme of the biaxial flexure strength test.
Figure 3Absorption spectrum of resin−based composites before (SPol) and after (Pol) polymerization.
Mean and standard deviation of the biaxial flexure strength tests (in N) of each resin-based composite (shade A2) using LEDs from second- and third-generation curing units.
| RBC Trademark | Emitter a Fit (1250 mW/cm2) | VALO (1100 mW/cm2) |
|---|---|---|
| Charisma A2 | 265.14 A 1,2 (±51.13) | 298.20 A 2,3 (±75.33) |
| Palfique A2 | 175.07 A 1 (±24.28) | 168.63 A 1 (±31.73) |
| Vittra A2 | 329.69 A 2,4 (±59.75) | 328.86 A 3 (±85.45) |
| Empress A2 | 203.17 A 1,3 (±40.03) | 186.95 A 1,2 (±54.38) |
| Z100 A2 | 307.04 A 2,3,4 (±29.16) | 363.85 A 3 (±81.17) |
| Z250 A2 | 329.93 A 2,4 (±126.23) | 363.16 A 3 (±81.17) |
| Z350 A2 | 375.28 A 2 (±83.16) | 355.32 A 3 (±90.56) |
| Harmonize A2 | 368.53 A 2 (±46.30) | 348.88 A 3 (±51.50) |
| Herculite Precis A2 | 268.12 A 1,2 (±74.46) | 301.68 A 2,3 (±71.03) |
| Polofil Supra A2 | 357.81 A 2 (±72.88) | 287.21 A 1,3 (±49.82) |
| Amaris O2 | 212.73 A 1,4 (±39.55) | 185.97 A 1,2 (±30.97) |
| Admira Fusion A2 | 227.02 A 1,4 (±53.09) | 224.06 A 1,2,4 (±31.14) |
| GrandioSO A2 | 379.65 A 2 (±91.65) | 299.53 A 2,3 (±58.59) |
| Grandio A2 | 381.51 A 2 (±53.34) | 328.27 A 3,4 (±70.17) |
Different letters in the same row denote a statistically significant difference (p > 0.05). Different numbers in the same column denote a statistically significant difference (p > 0.05).
Mean and standard deviation of the biaxial flexure strength test (in N) of each resin-based composite (shade A2) cured with a second-generation LED unit and microwave.
| RBC Trademark | Emitter a Fit | Emitter a Fit + Microwave |
|---|---|---|
| Z350 A2 | 375.28 (±83.16) | 334.99 (±50.19) |
| Z100 A2 | 307.04 (±29.14) | 317.42 (±83.02) |
| Grandio A2 | 381.51 (±53.84) | 301.08 (±76.42) |
| Harmonize A2 | 368.53 (±46.30) | 346.81 (±49.38) |
| Vittra A2 | 329.69 (±59.75) | 251.58 (±59.62) |
No statistically significant differences were found within the same RBC and the same curing mode (p > 0.05).
Mean and standard deviation of the biaxial flexure strength test (in Newtons (N)) of each resin-based composite (effects shade) cured with LED units from the second and third generations.
| RBC Trademark | Emitter a Fit | VALO | Emitter Now Duo |
|---|---|---|---|
| Palfique CE | 187.69 A 1 (±32.11) | 163.00 A 1,5 (±26.76) | 183.50 A 1 (±46.70) |
| Z350 GT | 365.93 A 2,3 (±62.08) | 385.28 A 2 (±72.85) | ----- |
| Herculite precis LTI | 295.45A1,2,3 (±66.87) | 297.55 A 2,3 (±51.80) | 308.10 A 2,3 (±79.27) |
| Grandio Incisal | 358.81 A 2,3 (±52.58) | 309.01 A 2,4 (±33.96) | 321.41 A 2 (±111.78) |
| Empress BL-L | 223.61 A 1,3 (±50.46) | 175.06 A 1,3 (±66.43) | 136.35 A 1 (±42.99) |
| Amaris Translúcida | 211.48 A 1,3 (±50.81) | 197.81 A 1,3,4 (±34.14) | 190.60 A 1,3 (±53.80) |
| Admira fusion Incisal | 241.63 A 3 (±42.26) | 207.76 A 1,3,4 (±52.76) | 153.87 A 1 (±39.51) |
| Grandio BL | 325.16 A 3 (±63.08) | 310.57 A 4,5,6 (±60.64) | 316.52 A 2 (±109.39) |
| GrandioSO BL | 315.55 A 3 (±64.50) | 292.31 A 2,3,4 (±79.83) | 308.31 A 2,3 (±59.65) |
| Vittra E Bleach | 310.73 A 1,3 (±88.72) | 394.57 A 2,6 (±115.73) | 306.97 A 2,3 (±117.74) |
Different letters in the same row denote a statistically significant difference (p > 0.05). Different numbers in the same column denote a statistically significant difference (p > 0.05). Shade effects mean that RBCs do not have hue and chroma.
Mean and standard deviation of the biaxial flexure strength test (in N) of resin-based composites of different shades cured with second- and third-generation LED units.
| RBC Trademark | Emitter a Fit | VALO | Emitter Now Duo | |||
|---|---|---|---|---|---|---|
| Shade A2 | Effect Shade | Shade A2 | Effect Shade | Shade A2 | Effect Shade | |
| Palfique | 175.07 (±24.28) | 187.69 (±32.11) | 168.63 (±31.73) | 163.00 (±26.76) | ----- | 183.5 (±46.70) |
| Z350 | 375.28 (±83.16) | 365.93 (±62.08) | 298.2 (±75.33) | 385.28 (±72.85) | 297.72 (±51.80) | ----- |
| Herculite precis | 268.12 (±74.46) | 295.45 (±66.87) | 186.95 (±54.38) | 297.55 (±51.80) | ----- | 308.10 (±79.27) |
| Grandio | 381.51 (±53.34) | 358.81 (±52.58) | 363.85 (±81.17) | 309.01 (±33.96) | 332.05 (±63.81) | 321.41 (±111.78) |
| Empress | 203.17 (±40.03) | 223.61 (±50.46) | 363.16 (±81.17) | 175.06 (±66.43) | ----- | 136.35 (±42.99) |
| Amaris | 212.73 (±39.55) | 211.48 (±50.81) | 355.32 (±90.56) | 197.81 (±34.14) | ----- | 190.6 (±53.80) |
| Admisa fusion | 227.02 (±53.09) | 241.63 (±42.26) | 348.88 (±51.50) | 207.76 (±52.76) | ----- | 153.87 (±39.51) |
| GrandioSO | 379.65 (±91.65) | 315.55 (±64.80) | 301.68 (±71.03) | 292.31 (±79.83) | ----- | 308.31 (±59.65) |
| Vittra | 329.69 (±59.75) | 310.73 (±88.72) | 287.21 (±49.82) | 394.57 (±115.73) | 251.58 (±59.62) | 306.97 (±117.74) |
No statistically significant differences were found between the RBC of shade A2 and shade effects within the same curing mode (p > 0.05). Shade effects mean that the RBCs do not have hue and chroma.
Degree of conversion (%) of each RBC, according to the curing mode.
| RBC Trademark | Emitter a Fit | VALO | Emitter Now Duo |
|---|---|---|---|
| Admira Fusion Incisal | 53.1 | 54.2 | 51.6 |
| Admira Fusion A2 | 53.4 | 51.5 | 54.4 |
| Amaris Translúcido | 37.8 | 38.7 | 37.3 |
| Amaris O2 | 37.4 | 37.9 | 37.5 |
| Charisma A2 | 43.1 | 47.9 | 43.3 |
| Empress BL-L | 30.4 | 32.6 | 30.3 |
| Empress A2 | 38.1 | 34.3 | 30.9 |
| Grandio Incisal | 44.7 | 47.5 | 45.0 |
| Grandio BL | 43.3 | 43.3 | 41.8 |
| Grandio A2 | 42.2 | 44.0 | 40.4 |
| GransioSO BL | 45.9 | 46.8 | 46.8 |
| Grandioso A2 | 46.9 | 48.6 | 45.7 |
| Harmonize A2 | 43.7 | 47.9 | 44.1 |
| Herculite Précis A2 | 50.0 | 51.2 | 50.1 |
| Palfique LX5 CE | 34.8 | 35.3 | 37.1 |
| Palfique A2 | 40.1 | 42.8 | 39.6 |
| Polofil Supra A2 | 58.2 | 56.3 | 54.8 |
| Vittra E Bleach | 31.3 | 66.2 | 48.5 |
| Vittra A2 | 42.8 | 50.0 | 35.9 |
| Z100 A2 | 40.0 | 39.5 | 38.2 |
| Z250 A2 | 49.3 | 48.3 | 47.4 |
| Z350 GT | 51.1 | 50.0 | 51.0 |
| Z350 A2 | 44.6 | 50.6 | 46.1 |
Statistical analysis was not performed since only one specimen of each RBC was used.