| Literature DB >> 36210937 |
Rina P Duke1, Shasha Bai2, Joshua A Bornhorst3, Nahed O ElHassan4, Jeffrey R Kaiser5.
Abstract
Background: The American Academy of Pediatrics and Pediatric Endocrine Society neonatal hypoglycemia guidelines based their glucose concentration treatment thresholds on studies that predominantly used Beckman and Yellow Springs Glucose Oxidase Analyzers. Currently, a majority (76%) of U.S. hospital laboratories utilizing glucose oxidase methodology use Vitros® Glucose Analyzers. However, a bias of ~+5% between glucose concentrations from Beckman vs. Vitros Glucose Analyzers has been reported; this could have a clinically significant effect when using published guideline treatment thresholds.Entities:
Keywords: glucose; glucose analyzer; guideline; neonatal hypoglycemia; newborn
Year: 2022 PMID: 36210937 PMCID: PMC9532546 DOI: 10.3389/fped.2022.933508
Source DB: PubMed Journal: Front Pediatr ISSN: 2296-2360 Impact factor: 3.569
Demographic and birth characteristics of term newborns who had plasma glucose concentrations measured by Beckman vs. Vitros analyzers.
|
|
|
|
|
|---|---|---|---|
| Year born | 1998 | 2008 | |
| Gestational age, median (IQR), weeks | 39.0 (38.0, 40.0) | 39.0 (38.0, 40.0) | 0.003† |
| Birth weight, median (IQR), g | 3,282 (3,068, 3,545) | 3,335 (3,066, 3,583) | 0.276† |
| Race, | < 0.001* | ||
| Black | 416 (46.0) | 283 (26.1) | |
| White | 422 (46.7) | 436 (40.4) | |
| Hispanic | 47 (5.2) | 312 (28.8) | |
| Asian | 19 (2.2) | 52 (4.8) | |
| Female sex, | 429 (47.5) | 510 (47.1) | 0.871* |
| Apgar 5 mins, | 0.039* | ||
| 7 | 40 (4.4) | 39 (3.6) | |
| Time to specimen collection, median (IQR), minutes | 93 (75, 115) | 101 (79, 126) | < 0.001† |
| Glucose concentration, median (IQR), mg/dL | 59 (50, 68) | 53 (46, 62) | < 0.001† |
*Chi-Square test. †Mann Whitney U. IQR, interquartile range.
Figure 1Pattern of fitted glucose concentrations during the first 3 h after birth. Restricted cubic spline curves were fit from initial glucose concentrations from term, appropriate size for gestational age newborns in the Beckman and Vitros Analyzers. (A) Fitted glucose concentrations and 95% confidence intervals from newborns using Beckman and Vitros Analyzers. (B) Fitted glucose concentrations and 95% confidence intervals from newborns using Beckman and Vitros Analyzers, after 5 mg/dL was added to the Vitros glucose concentration spline curve.