Yoshitsugu Matsumoto1, Etsuo Kunieda1, Natsumi Futakami2, Takeshi Akiba2, Ryuta Nagao1, Tsuyoshi Fukuzawa1, Tomomi Katsumata1, Toshihisa Kuroki2, Tatsuya Mikami1, Yasuhiro Okumura3, Kenji Souda3, Eride Mutu1, Akitomo Sugawara1.
Abstract
Background: A high-definition multi-leaf collimator (HD-MLC) with 5- and 10-mm fine MLCs is useful for radiotherapy. However, it is difficult to irradiate the mammary gland and supraclavicular region using a HD-MLC because of the narrow field of volumetric modulated arc radiotherapy (VMAT). Therefore, we aimed to evaluate the dose distribution of the VMAT dose using a HD-MLC in 15 patients with left breast cancer undergoing postoperative irradiation of breast and regional lymph nodes, including the internal mammary node. Materials and methods: The following four plans were generated: three-arc VMAT using HD-MLC (HD-VMAT), two tangential arcs and one-arc VMAT using HD-MLC (tHD-VMAT), three-dimensional conformal radiotherapy (3DCRT) using HD-MLC, and two-arc VMAT using the Millennium 120-leaf MLC (M-VMAT). We assessed the doses to the target volume and organs at risk.Entities:
Keywords: Millennium 120-leaf MLC; breast cancer; high-definition multi-leaf collimator; three-dimensional conformal radiotherapy; volumetric modulated arc therapy
Year: 2022 PMID: 36196412 PMCID: PMC9521697 DOI: 10.5603/RPOR.a2022.0081
Source DB: PubMed Journal: Rep Pract Oncol Radiother ISSN: 1507-1367
Figure 1Comparisons of dose indices among HD-MLC, tHD-VMAT, 3DCRT, and M-VMAT. A. Target doses; B. Homogeneity index (HI) and conformity index (CI); C. Organs at risk. **p < 0.01, *p < 0.05. HD-VMAT — three-arc volumetric modulated arc radiotherapy using HD-MLC; tHD-VMAT — tangential volumetric modulated arc radiotherapy using HD-MLC; 3DCRT — three-dimensional conformal radiotherapy; M-VMAT — volumetric modulated arc radiotherapy using Millennium 120-leaf MLC; CTV — clinical target volume; PTV — planning target volume; D — dose; IM — internal mammary
Dosing parameters according to radiotherapy plan
| Dose parameters | HD-VMAT | tHD-VMAT | 3DCRT | M-VMAT |
|---|---|---|---|---|
| Mean ± SD | Mean ± SD | Mean ± SD | Mean ± SD | |
|
| ||||
| D95% [Gy] | 47.90 ± 0.59 | 47.56 ± 0.9 | 47.13 ± 0.81 | 48.07 ± 0.46 |
| D90% [Gy] | 48.65 ± 0.32 | 48.31 ± 2.54 | 48.19 ± 0.66 | 48.69 ± 0.33 |
|
| ||||
| D95% [Gy] | 39.93 ± 3.04 | 40.13 ± 2.96 | 37.60 ± 4.14 | 38.52 ± 2.75 |
| D90% [Gy] | 45.07 ± 1.12 | 45.20 ± 1.20 | 43.76 ± 2.26 | 44.11 ± 1.17 |
|
| ||||
| D95% [Gy] | 43.53 ± 2.47 | 43.98 ± 1.58 | 39.92 ± 4.58 | 43.96 ± 0.94 |
| D90% [Gy] | 45.07 ± 1.12 | 45.42 ± 1.43 | 42.49 ± 3.03 | 45.28 ± 0.89 |
|
| ||||
| V30 Gy (%) | 9.43 ± 5.46 | 11.25 ± 4.87 | 14.59 ± 2.65 | 12.29 ± 2.93 |
| V20 Gy (%) | 15.80 ± 6.48 | 17.40 ± 6.02 | 16.20 ± 2.73 | 19.07 ± 3.83 |
| V10 Gy (%) | 30.49 ± 5.38 | 31.00 ± 4.68 | 18.74 ± 2.91 | 32.67 ± 4.99 |
| V5 Gy (%) | 54.80 ± 8.41 | 48.71 ± 7.81 | 22.49 ± 3.01 | 57.34 ± 10.78 |
| Dmean [Gy] | 10.75 ± 2.87 | 10.38 ± 1.94 | 8.17 ± 1.28 | 11.51 ± 1.59 |
|
| ||||
| V30 Gy (%) | 2.33 ± 2.01 | 6.10 ± 4.41 | 10.52 ± 5.76 | 3.54 ± 2.71 |
| V20 Gy (%) | 5.86 ± 3.55 | 10.53 ± 6.09 | 12.79 ± 6.24 | 8.05 ± 5.37 |
| V10 Gy (%) | 16.17 ± 6.37 | 25.35 ± 13.91 | 15.83 ± 6.75 | 20.69 ± 15.67 |
| Dmean [Gy] | 6.73 ± 1.54 | 8.95 ± 3.13 | 7.22 ± 2.77 | 8.00 ± 2.73 |
|
| ||||
| V20 Gy (%) | 3.78 ± 7.73 | 2.78 ± 4.68 | 2.09 ± 2.8 | 4.71 ± 5.90 |
| V10 Gy (%) | 25.03 ± 13.46 | 20.93 ± 13.15 | 2.98 ± 3.53 | 22.59 ± 13.02 |
| Dmean [Gy] | 7.84 ± 2.29 | 6.38 ± 2.15 | 1.91 ± 1.32 | 7.44 ± 2.20 |
| Homogeneity index (%) | 42.85 ± 7.76 | 43.84 ± 8.49 | 49.23 ± 11.55 | 47.69 ± 10.26 |
| Conformity index | 0.36 ± 0.05 | 0.36 ± 0.05 | 0.32 ± 0.06 | 0.39 ± 0.05 |
CTV — clinical target volume; PTV — planning target volume; PTV_IM — PTV internal mammary node; VMAT — volumetric modulated arc radiotherapy; HD-VMAT — three-arc VMAT using HD-MLC; tHD-VMAT — two tangential arcs and one arc VMAT using HD-MLC; 3DCRT — three-dimensional conformal radiotherapy; M-VMAT — two-arc VMAT using Millennium 120-leaf MLC; Breast_Rt — contralateral breast
Difference in each dosing parameter of three-arc VMAT using HD-MLC (HD-VMAT) relative to other radiotherapy plans
| Dose parameters | tHD-VMAT | 3DCRT | M-VMAT | |||
|---|---|---|---|---|---|---|
| Difference | p | Difference | p-value | Difference | p-value | |
|
| ||||||
| D95% [Gy] | −0.34 | 0.211 | −0.77 | 0.008 | 0.17 | 0.865 |
| D90% [Gy] | −0.34 | 0.001 | −0.45 | 0.027 | 0.04 | 0.733 |
|
| ||||||
| D95% [Gy] | 0.20 | 0.308 | −2.33 | 0.001 | −1.40 | 0.001 |
| D90% [Gy] | 0.13 | 0.916 | −1.31 | 0.020 | −0.96 | 0.001 |
|
| ||||||
| D95% [Gy] | 0.45 | 0.570 | −3.61 | 0.009 | 0.43 | 0.427 |
| D90% [Gy] | 0.35 | 0.394 | −2.58 | 0.006 | 0.21 | 0.394 |
|
| ||||||
| V30 Gy (%) | 1.82 | 0.023 | 5.16 | 0.009 | 2.93 | 0.061 |
| V20 Gy (%) | 1.60 | 0.019 | 0.40 | 0.427 | 3.32 | 0.061 |
| V10 Gy (%) | 0.51 | 0.279 | −11.75 | 0.001 | 2.25 | 0.191 |
| V5 Gy (%) | −6.10 | 0.003 | −32.31 | 0.001 | 2.54 | 0.650 |
| Dmean [Gy] | −0.37 | 0.733 | −2.58 | 0.002 | 0.76 | 0.069 |
|
| ||||||
| V30 Gy (%) | 3.78 | 0.001 | 8.20 | 0.001 | 1.22 | 0.061 |
| V20 Gy (%) | 4.67 | 0.004 | 6.93 | 0.002 | 2.19 | 0.069 |
| V10 Gy (%) | 9.18 | 0.004 | −0.34 | 0.776 | 4.51 | 0.570 |
| Dmean [Gy] | 2.23 | 0.002 | 0.50 | 0.532 | 1.27 | 0.031 |
|
| ||||||
| V20 Gy (%) | −1.00 | 0.583 | −1.69 | 0.929 | 0.93 | 0.069 |
| V10 Gy (%) | −4.09 | 0.069 | −22.04 | 0.001 | −2.43 | 0.334 |
| Dmean [Gy] | −1.46 | 0.002 | −5.93 | 0.001 | −0.40 | 0.334 |
| Homogeneity index | 0.99 | 0.069 | 6.38 | 0.001 | 4.84 | 0.001 |
| Conformity index | 0.00 | 0.875 | −0.04 | 0.023 | 0.03 | 0.112 |
CTV — clinical target volume; PTV — planning target volume; PTV_IM — PTV internal mammary node; VMAT — volumetric modulated arc radiotherapy; tHD_VMAT — two tangential arcs and one arc VMAT using HD-MLC; 3DCRT — three-dimensional conformal radiotherapy; M-VMAT, two-arc VMAT using Millennium 120-leaf MLC; Breast_Rt — contralateral breast; Difference, the respective parameter of the indicated plan minus the HD-VMAT parameter.
p < 0.01,
p < 0.05
Figure 2Dose distributions of volumetric modulated arc radiotherapy using high-definition multi-leaf collimator (HD-MLC) (first column), tangential volumetric modulated arc radiotherapy using HD-MLC (second column), and three-dimensional conformal radiotherapy (third column). Dose distributions of volumetric modulated arc radiotherapy using Millennium 120-leaf MLC (fourth column). The dose in the isodose curve follows the left label. The contours shown are the clinical target volume (navy) and planning target volume (red)