| Literature DB >> 36195916 |
Bipin Adhikari1,2, Rupam Tripura3,4, Lek Dysoley5, James J Callery3,4, Thomas J Peto3,4, Chhoeun Heng3, Thy Vanda3, Ou Simvieng3, Sarah Cassidy-Seyoum6, Benedikt Ley6, Kamala Thriemer6, Arjen M Dondorp3,4, Lorenz von Seidlein3,4.
Abstract
BACKGROUND: Quantitative measurement of Glucose-6-Phosphate Dehydrogenase (G6PD) enzyme activity is critical to decide on appropriate treatment and provision of radical cure regimens for vivax malaria. Biosensors are point-of-care semi-quantitative analysers that measure G6PD enzyme activity. The main objective of this study was to evaluate the operational aspects of biosensor deployment in the hands of village malaria workers (VMWs) in Cambodia over a year.Entities:
Keywords: Cambodia; Community; G6PD; Quantitative; Radical cure; Village malaria workers; Vivax malaria
Mesh:
Substances:
Year: 2022 PMID: 36195916 PMCID: PMC9531219 DOI: 10.1186/s12936-022-04300-9
Source DB: PubMed Journal: Malar J ISSN: 1475-2875 Impact factor: 3.469
Fig. 1Schematic diagram of study activities of RORC
Fig. 2Stepwise illustration of G6PD test using biosensor
(reproduced from the PATH website at www.path.org dated 28 June 2022)
Socio-demographics of participants of Semi-Structured Interviews (SSIs)
| SSI | Age (years) | Sex | Qualification (years of education) |
|---|---|---|---|
| SSI-1 | 65 | M | 7 |
| SSI-2 | 55 | F | 7 |
| SSI-3 | 30 | F | 12 |
| SSI-4 | 42 | F | 6 |
| SSI-5 | 51 | F | 6 |
| SSI-6 | 40 | F | 9 |
| SSI-7 | 38 | F | 12 |
| SSI-8 | 54 | F | 12 |
| SSI-9 | 41 | F | 9 |
| SSI-10 | 59 | F | 9 |
| SSI-11 | 27 | M | 12 |
| SSI-12 | 23 | M | 12 |
| SSI-13 | 41 | M | 9 |
| SSI-14 | 43 | M | 7 |
| SSI-15 | 41 | F | 7 |
| SSI-16 | 50 | F | 6 |
| SSI-17 | 51 | M | 8 |
| SSI-18 | 54 | M | 9 |
| SSI-19 | 49 | F | 9 |
| SSI-20 | 36 | F | 5 |
| SSI-21 | 30 | F | 6 |
| SSI-22 | 25 | F | 7 |
| SSI-23 | 36 | F | 7 |
| SSI-24 | 42 | M | 9 |
| SSI-25 | 52 | M | 12 |
| SSI-26 | 54 | M | 12 |
| SSI-27 | 26 | M | University |
| SSI-28 | 50 | M | 12 |
| SSI-29 | 31 | F | 6 |
| SSI-30 | 43 | M | 9 |
| SSI-31 | 48 | M | 4 |
| SSI-32 | 48 | M | 11 |
| SSI-33 | 52 | M | 2 |
Information on their workplace is deliberately avoided in the table to prevent potential identification of the participants
Fig. 3Current scenario of vivax management referral system in Cambodia
Fig. 4Considerations for integrating biosensor into VMW/MMWs’ routine work at the community
Fig. 5A VMW using a biosensor machine with a patient at the community
The standard G6PD test assessment among the participants at T0 and T30 (n = 33)
| Items | T0 | T30 |
|---|---|---|
| Number (%) | Number (%) | |
| The STANDARD G6PD test identifies G6PD deficiency | 32 (97) | 33 (100) |
| Standard G6PD measures enzyme reaction | 24 (73) | 32 (97) |
| Optimal temperature for biosensor operation 15–40 °C | 30 (91) | 26 (79) |
| Type of samples for G6PD test: capillary and venous blood | 22 (67) | 23 (70) |
| Must matching of chip and strip pouch | 26 (79) | 25 (76) |
| Insert the test strip after the code chip is entered into the analyzer | 21 (64) | 19 (58) |
| Mix blood and buffer in EziTube 8–10 times | 26 (79) | 30 (91) |
| Use 10 µl or the black line on the EziTube | 29 (88) | 28 (85) |
| Apply the mixture immediately to the test strip | 11 (33) | 0 |
| Number of Ezi tubes to run the samples = 2 | 26 (79) | 26 (79) |
| Test strip cannot be reused | 30 (91) | 33 (100) |
| Correct precautions to avoid the injury | 14 (42) | 17 (52) |
| G6PD 1.2 is deficient | 27 (82) | 27 (82) |
| G6Pd 13.1 is read correct | 33 (100) | 33 (100) |
| Hb 15.7 is read correct | 31 (94) | 32 (97) |
| G6PD 13.1 is normal | 29 (88) | 31 (94) |
| G6PD 0.7 is read correct | 33 (100) | 33 (100) |
| Hb 11.1 is read correct | 32 (97) | 33 (100) |
| G6PD 0.7 is deficient | 33 (100) | 32 (97) |
| G6PD 4.5 is read correct | 33 (100) | 33 (100) |
| Hb 13.4 is read correct | 32 (97) | 32 (97) |
| G6PD 4.5 is intermediate | 29 (88) | 25 (76) |
| G6PD n-A is read correct | 32 (97) | 33 (100) |
| Hb Lo is read correct | 32 (97) | 33 (100) |
| Test did not work (error or NA) | 33 (100) | 33 (100) |
| G6PD 2.0 is read correct | 33 (100) | 33 (100) |
| Hb 13.2 is read correct | 33 (100) | 33 (100) |
| G6PD 2.0 is deficient | 33 (100) | 33 (100) |
| G6PD E-2 is read correct | 32 (97) | 31 (94) |
| Hb (none) | 33 (100) | 33 (100) |
| Test did not work (error or NA) | 30 (91) | 33 (100) |
| G6PD 9.2 is read correct | 32 (97) | 33 (100) |
| Hb 5.8 is read correct | 31 (94) | 33 (100) |
| G6PD 9.2 is normal | 27 (82) | 19 (58) |
Fig. 6Evaluation of biosensor use by health workers based on the observation by trainer at M8
Concerns and recommendations around the use of biosensor machine
| Items | Problems | Data |
|---|---|---|
| Pipette | Black line in pipette is hard to appreciate. Especially older VMWs had problem appreciating the black line Pipettes are too small. When the pipette takes bubble, it can affect the result Need to use two pipettes | |
| Sample Processing | The sample processing steps require a prompt completion, letting the measurement chamber flap open for little longer, machine will display an error | |
| Test device | The hole within the test device is too small and sometimes mixture of buffer and blood were not able to be placed in it | |
| Chip code | Insertion and removal of chip code is difficult. Removing chip code was difficult, some used knives to remove it and were scared that it could break | |
| Buffer tube | VMWs did not feel comfortable with buffer tube because there is nothing to support it, it can fall down any time by wind, or air from fan. It should have something to support; support for buffer tube so that it would be easy while processing | |
| Device package box | The device package box is made from paper and thus can absorb water and is vulnerable to easy damage | |
| Battery | There is no notification about battery and VMWs could not identify the status of the battery. Electricity-chargeable battery would have been great |