| Literature DB >> 36193239 |
Vicki Trier Taasti1, Colien Hazelaar1, Femke Vaassen1, Ana Vaniqui1, Karolien Verhoeven1, Frank Hoebers1, Wouter van Elmpt1, Richard Canters1, Mirko Unipan1.
Abstract
Background and purpose: Treatment quality of proton therapy can be monitored by repeat-computed tomography scans (reCTs). However, manual re-delineation of target contours can be time-consuming. To improve the workflow, we implemented an automated reCT evaluation, and assessed if automatic target contour propagation would lead to the same clinical decision for plan adaptation as the manual workflow. Materials and methods: This study included 79 consecutive patients with a total of 250 reCTs which had been manually evaluated. To assess the feasibility of automated reCT evaluation, we propagated the clinical target volumes (CTVs) deformably from the planning-CT to the reCTs in a commercial treatment planning system. The dose-volume-histogram parameters were extracted for manually re-delineated (CTVmanual) and deformably mapped target contours (CTVauto). It was compared if CTVmanual and CTVauto both satisfied/failed the clinical constraints. Duration of the reCT workflows was also recorded.Entities:
Keywords: Automation; Deformable image registration; Efficiency gain; Proton therapy; Repeat-CT evaluation; Scripting
Year: 2022 PMID: 36193239 PMCID: PMC9525894 DOI: 10.1016/j.phro.2022.09.009
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Schematic workflow for the reCT evaluation leading to the decision whether the plan was still acceptable (i.e., satisfying the target coverage constraints) on the reCT. The top row shows the previous manual workflow which typically took a couple of days to complete. In the third step, the RTT created the image registration, here the * indicates that both a rigid and a deformable image registration (DIR) was created (since a rigid registration was needed to create a DIR), and then the OARs were propagated using the DIR. In the bottom row, the new workflow is shown. In this workflow the largest work-burdens were replaced by scripting. A single script was created to perform all the scripted steps.
Fig. 2(Left) Volume difference between the manually delineated and the deformably propagated target contours. (Right) DICE value for the overlap between the manually delineated target contours on the reCTs and the deformed contours propagated from the pCT to the following reCTs.
Number of reCTs presented with each result; the numbers in the parentheses are the rates relative to the total number of reCTs.
| Total | Brain | Head-and-neck | Breast | Lung | Lymphoma | |
|---|---|---|---|---|---|---|
| # Patients | 12 | 14 | 35 | 14 | 4 | |
| # reCTs evaluated | 59 | 63 | 69 | 50 | 9 | |
| # Contours evaluated | 58 | 153 | 303 | 110 | 19 | |
| # True positives (rate) | 35 (59%) | 34 (54%) | 22 (32%) | 19 (38%) | 8 (89%) | |
| # True negatives (rate) | 24 (41%) | 19 (30%) | 39 (57%) | 28 (56%) | 1 (11%) | |
| # False positives (rate) | 0 (0%) | 5 (8%) | 5 (7%) | 2 (4%) | 0 (0%) | |
| # False negatives (rate) | 0 (0%) | 5 (8%) | 3 (4%) | 1 (2%) | 0 (0%) |
Details on false negatives. Second column specifies the reCT number (no.), (x2) means that two target dose constraints were violated for CTVmanual but not for CTVauto on the given reCT. Fourth column states how much the target dose-volume-histogram (DVH) constraint was violated for CTVmanual, computed as , where is the DVH parameter extracted for CTVmanual. The DVH parameter difference listed in column five is given as . If and are both close to 0%, the constraint was just barely violated for CTVmanual and just barely fulfilled for CTVauto, meaning that the two DVH parameters were almost the same. The second reCT for the third HN patient is marked in boldface, since a plan adaptation was made on this reCT.
| Site (no.) | reCT no. | Contour volumes | CTVmanual failure ( | Goal value difference ( | |||
|---|---|---|---|---|---|---|---|
| HN (1) | reCT 5 | 23.1 vs 26.0 | −0.6 | 1.8 | |||
| HN (2) | reCT 4 (x2) | 135.7 vs 126.8 | −0.04 | −0.05 | −0.1 | −0.1 | |
| HN (3) | reCT 1 (x2) | 32.6 vs 23.6 | −0.1 | −0.2 | −0.5 | −0.6 | |
| 33.0 vs 26.3 | −2.2 | 3.1 | |||||
| HN (4) | reCT 2 | 232.9 vs 234.1 | −0.5 | 1.1 | |||
| Breast (5) | reCT 1 | 17.3 vs 16.7 | −0.9 | 2.6 | |||
| Breast (6) | reCT 1 (x2) | 960.7 vs 941.8 | 886.1 vs 866.5 | −0.4 | −0.6 | 0.7 | 0.7 |
| Breast (7) | reCT 2 | 510.9 vs 499.1 | −0.2 | 0.6 | |||
| Lung (8) | reCT 2 | 100.3 vs 67.9 | −0.1 | 4.4 | |||
Fig. 3Example of false negative repeat-CT (reCT), for HN patient 3 who had a plan adaption on this reCT (reCT 2; see Table 2). (Left) Planning-CT (pCT), here an air cavity, representing the right piriform sinus, is seen (dashed orange circle and arrow) which is not present in the same location on the reCT (middle), likely due to radiation-induced edema. The blue contour is the delineated CTV on the pCT (it has been rigidly copied to the reCT for visualization purposes), the yellow contour is the delineated GTV, the purple contour is the delineated CTV on the reCT, and the red contour is the deformed CTV on the reCT propagated from the pCT. (Right) Deformation vector field between the pCT and the reCT; the colors of the vectors indicate their length. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Average duration of the reCT evaluation workflow before and after implementing the script-based workflow. The number in parenthesis after the tumor site indicates how many reCTs were included in each subgroup (before and after).
| Tumor site | Timing of reCT evaluation workflow (h) | Time reduction (h/%) | |
|---|---|---|---|
| Manual workflow | Applying script | ||
| Brain (100) | 68.4 | 15.0 | 53.4 (78%) |
| HN (35) | 66.4 | 19.2 | 47.2 (71%) |
| Lung (150) | 63.2 | 16.1 | 47.1 (75%) |
| Total (285) | 65.4 | 16.1 | 49.3 (75%) |