iCycle: Integrated, multicriterial beam angle, and profile optimization for generation of coplanar and noncoplanar IMRT plans.

Sebastiaan Breedveld1, Pascal R M Storchi, Peter W J Voet, Ben J M Heijmen.   

Abstract

PURPOSE: To introduce iCycle, a novel algorithm for integrated, multicriterial optimization of beam angles, and intensity modulated radiotherapy (IMRT) profiles.
METHODS: A multicriterial plan optimization with iCycle is based on a prescription called wish-list, containing hard constraints and objectives with ascribed priorities. Priorities are ordinal parameters used for relative importance ranking of the objectives. The higher an objective priority is, the higher the probability that the corresponding objective will be met. Beam directions are selected from an input set of candidate directions. Input sets can be restricted, e.g., to allow only generation of coplanar plans, or to avoid collisions between patient/couch and the gantry in a noncoplanar setup. Obtaining clinically feasible calculation times was an important design criterium for development of iCycle. This could be realized by sequentially adding beams to the treatment plan in an iterative procedure. Each iteration loop starts with selection of the optimal direction to be added. Then, a Pareto-optimal IMRT plan is generated for the (fixed) beam setup that includes all so far selected directions, using a previously published algorithm for multicriterial optimization of fluence profiles for a fixed beam arrangement Breedveld et al. [Phys. Med. Biol. 54, 7199-7209 (2009)]. To select the next direction, each not yet selected candidate direction is temporarily added to the plan and an optimization problem, derived from the Lagrangian obtained from the just performed optimization for establishing the Pareto-optimal plan, is solved. For each patient, a single one-beam, two-beam, three-beam, etc. Pareto-optimal plan is generated until addition of beams does no longer result in significant plan quality improvement. Plan generation with iCycle is fully automated.
RESULTS: Performance and characteristics of iCycle are demonstrated by generating plans for a maxillary sinus case, a cervical cancer patient, and a liver patient treated with SBRT. Plans generated with beam angle optimization did better meet the clinical goals than equiangular or manually selected configurations. For the maxillary sinus and liver cases, significant improvements for noncoplanar setups were seen. The cervix case showed that also in IMRT with coplanar setups, beam angle optimization with iCycle may improve plan quality. Computation times for coplanar plans were around 1-2 h and for noncoplanar plans 4-7 h, depending on the number of beams and the complexity of the site.
CONCLUSIONS: Integrated beam angle and profile optimization with iCycle may result in significant improvements in treatment plan quality. Due to automation, the plan generation workload is minimal. Clinical application has started.

Entities:  

Mesh:

Year:  2012        PMID: 22320804     DOI: 10.1118/1.3676689

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  69 in total

1.  Automating proton treatment planning with beam angle selection using Bayesian optimization.

Authors:  Vicki T Taasti; Linda Hong; Jin Sup Andy Shim; Joseph O Deasy; Masoud Zarepisheh
Journal:  Med Phys       Date:  2020-05-27       Impact factor: 4.071

2.  Optimization approaches to volumetric modulated arc therapy planning.

Authors:  Jan Unkelbach; Thomas Bortfeld; David Craft; Markus Alber; Mark Bangert; Rasmus Bokrantz; Danny Chen; Ruijiang Li; Lei Xing; Chunhua Men; Simeon Nill; Dávid Papp; Edwin Romeijn; Ehsan Salari
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

Review 3.  Automation in intensity modulated radiotherapy treatment planning-a review of recent innovations.

Authors:  Mohammad Hussein; Ben J M Heijmen; Dirk Verellen; Andrew Nisbet
Journal:  Br J Radiol       Date:  2018-09-04       Impact factor: 3.039

4.  Automatic replanning of VMAT plans for different treatment machines: A template-based approach using constrained optimization.

Authors:  Luise A Künzel; Oliver S Dohm; Markus Alber; Daniel Zips; Daniela Thorwarth
Journal:  Strahlenther Onkol       Date:  2018-05-30       Impact factor: 3.621

5.  Many-isocenter optimization for robotic radiotherapy.

Authors:  Qihui Lyu; Ryan Neph; Victoria Y Yu; Dan Ruan; Salime Boucher; Ke Sheng
Journal:  Phys Med Biol       Date:  2020-02-10       Impact factor: 3.609

6.  Automatic configuration of the reference point method for fully automated multi-objective treatment planning applied to oropharyngeal cancer.

Authors:  Rens van Haveren; Ben J M Heijmen; Sebastiaan Breedveld
Journal:  Med Phys       Date:  2020-03-05       Impact factor: 4.071

7.  Software-based evaluation of a class solution for prostate IMRT planning.

Authors:  Sarah Clarke; Josie Goodworth; Justin Westhuyzen; Brendan Chick; Matthew Hoffmann; Jacqueline Pacey; Stuart Greenham
Journal:  Rep Pract Oncol Radiother       Date:  2017-08-30

8.  Using deep learning to predict beam-tunable Pareto optimal dose distribution for intensity-modulated radiation therapy.

Authors:  Gyanendra Bohara; Azar Sadeghnejad Barkousaraie; Steve Jiang; Dan Nguyen
Journal:  Med Phys       Date:  2020-08-02       Impact factor: 4.071

9.  Automated intensity modulated treatment planning: The expedited constrained hierarchical optimization (ECHO) system.

Authors:  Masoud Zarepisheh; Linda Hong; Ying Zhou; Jung Hun Oh; James G Mechalakos; Margie A Hunt; Gig S Mageras; Joseph O Deasy
Journal:  Med Phys       Date:  2019-05-29       Impact factor: 4.071

10.  A surrogate-based metaheuristic global search method for beam angle selection in radiation treatment planning.

Authors:  H H Zhang; S Gao; W Chen; L Shi; W D D'Souza; R R Meyer
Journal:  Phys Med Biol       Date:  2013-03-21       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.