| Literature DB >> 36186969 |
Liang Jin1, Pan Gao1, Kun Wang1, Jianying Li2, Ming Li1,3.
Abstract
Objectives: To evaluate whether applying image filters (smooth 3D+ and edge-2) improves image quality in coronary CT angiography (CCTA).Entities:
Keywords: X-ray computed tomography; analysis; coronary artery disease; image quality; sharpness
Year: 2022 PMID: 36186969 PMCID: PMC9521173 DOI: 10.3389/fcvm.2022.840735
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Figure 1The flow chart of the patients' enrollment.
Figure 2Procedure for generating the SE group. Step 1: Click the “No Filter” option in red frame 1 and select smooth 3D+; step 2: Click the button in red frame 2 to generate the smoothed 3D+ group. Then, quit the software program and start again, repeat step 1 but choose the edge-2 filter, and repeat step 2 and generate the SE group. Finally, the original series 402 and the SE group 404 were showed in two red frame 3.
Diagnostic criteria for coronary stenosis by CCTA and DSA.
|
|
| |
|---|---|---|
| CAD-RADS 0 | 0% (No plaque or stenosis) | Documented absence of CAD |
| CAD-RADS 1 | 1–24% - Minimal stenosis or plaque | Minimal non-obstructive CAD |
| CAD-RADS 2 | 25–49% - Mild stenosis | Mild non-obstructive CAD |
| CAD-RADS 3 | 50–69% stenosis | Moderate stenosis |
| CAD-RADS 4 | 70–99% stenosis | Severe stenosis |
| CAD-RADS 5 | 100% (total occlusion) | Total coronary occlusion |
| CAD-RADS 6 | Non-diagnostic study | Obstructive CAD cannot be excluded |
CAD-RADS, Coronary Artery Disease - Reporting and Data System; CCTA, Coronary computed tomography angiography; DSA, digital subtraction angiography.
Figure 3The flow chart shows how to calculate the image sharpness. (A1,A2) are the same Digital Imaging and Communications in Medicine (DICOM) slices of the beginning of the coronary sinus of one patient [(A1) from the routine group and (A2) from the SE group], which are exported to be a PNG format (Portable Network Graphic). Red frames in (B1,B2) are the setting of ROI (region of interest) (C1,C2). Then, the software calculates the image sharpness through the ROI based on the MTF50, MTF50P, and MTF 30 [red underlines in (D1,D2)].
Patients' information and radiation dose.
| Age | 67.1 ± 11.7 |
| Sex (F/M) | 23/67 |
| BMI (kg/m2) | 27.4 ± 4.9 |
| Smoker | 56 (62%) |
| Hypertension | 70 (78%) |
| Diabetes | 61 (68) |
| Hyperlipemia | 39 (43%) |
| DLP, (mGy) | 344.69 ± 103.07 |
| ED, (mSv) | 4.8 ± 1.4 |
F/M, Female/Male; BMI, Body Mass Index; DLP, Dose-length Product; ED, Effective radiation dose.
Comparison of sharpness of image between the routine and SE groups.
|
|
|
|
|
|---|---|---|---|
| AO (HU) | 409.55 ± 77.48 | 407.42 ± 77.86 | <0.001 |
| Image noise | 25.73 ± 7.12 | 16.34 ± 6.12 | <0.001 |
|
| |||
| AO | 16.99 ± 5.16 | 28.34 ± 12.25 | <0.001 |
| LAD-P | 26.43 ± 15.06 | 46.51 ± 37.70 | <0.001 |
| LAD-M | 9.38 ± 5.74 | 12.48 ± 12.36 | 0.005 |
| LAD-D | 6.49 ± 5.20 | 7.80 ± 6.19 | 0.009 |
| LCX-P | 26.43 ± 15.06 | 46.51 ± 37.70 | <0.001 |
| LCX-M | 13.19 ± 7.43 | 18.04 ± 12.48 | <0.001 |
| LCX-D | 7.52 ± 4.95 | 9.74 ± 6.55 | <0.001 |
| RCA-P | 26.42 ± 15.30 | 32.29 ± 30.99 | <0.001 |
| RCA-M | 17.68 ± 11.92 | 25.75 ± 23.63 | <0.001 |
| RCA-D | 13.43 ± 9.43 | 18.28 ± 12.75 | <0.001 |
|
| |||
| AO | 40.48 ± 32.74 | 80.65 ± 51.39 | <0.001 |
| LAD-P | 42.01 ± 31.30 | 76.30 ± 48.96 | <0.001 |
| LAD-M | 29.57 ± 23.41 | 52.60 ± 39.45 | <0.001 |
| LAD-D | 23.96 ± 18.87 | 48.83 ± 33.68 | <0.001 |
| LCX-P | 42.01 ± 31.30 | 76.30 ± 48.96 | <0.001 |
| LCX-M | 32.59 ± 27.49 | 64.91 ± 43.19 | <0.001 |
| LCX-D | 25.59 ± 22.75 | 51.92 ± 33.44 | <0.001 |
| RCA-P | 42.71 ± 29.89 | 76.80 ± 50.84 | <0.001 |
| RCA-M | 32.79 ± 27.09 | 68.40 ± 45.64 | <0.001 |
| RCA-D | 34.06 ± 22.97 | 60.73 ± 39.33 | <0.001 |
|
| |||
| MTF50 | 0.0091 (0.0046, 0.0137) | 0.0093 (0.0051, 0.0164) | 0.036 |
| MTF50P | 0.0088 (0.0046, 0.0128) | 0.0090 (0.0059, 0.0136) | 0.049 |
| MTF30 | 0.0134 (0.0072, 0.0217) | 0.0211 (0.0146, 0.0258) | 0.122 |
SE, smooth 3D+ combined with edge-2; CT, computed tomography; SNR, signal-to-noise ratio; CNR, contrast-to-noise ratio; HU, household; AO, lumen of the aortic root; LAD-P, M, D, left anterior descending artery, proximal, middle, and distal; LCX-P, M, D, left circumflex artery, proximal, middle, and distal; RCA-P, M, D, right coronary artery, proximal, middle, and distal; MTF, Modulation Transfer Function. Significance level, P < 0.05.
Comparison of image quality scores between the routine and SE groups.
|
|
|
|
| ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| ||||||
| Score (Median (25%,75%) | Routine group | 3 (2, 3) | 3 (2, 3) | 0.81 | 3 (3, 4) | 3 (3, 4) | 0.83 | 3 (2, 3) | 3 (2, 3) | 0.84 | 3(2, 4) | 3(2, 4) | 0.83 |
| SE group | 2 (1, 3) | 2 (1, 3) | 0.78 | 2 (2, 3) | 3 (2, 3) | 0.80 | 2 (1, 3) | 2 (1, 2) | 0.71 | 2(1, 3) | 2(1, 3) | 0.77 | |
|
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | |||
| Improved Vessels | 48(57) | 59 (60) | 51 (49) | 60 (56) | 40 (69) | 57 (69) | 139(70) | 176(69) | |||||
SE, smooth 3D+ combined with edge-2; LAD, left anterior descending artery; LCX, left circumflex artery; ka., Kappa; R, radiologist; RCA, right coronary artery.
Comparison of diagnostic accuracy among the routine group, SE group, and the DSA group for coronary artery stenosis on per-segment and total levels.
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| |
| FPR, % | 0.36 | 0.39 | 0.43 | 0.43 | 0.41 | 0.41 | 0.39 | 0.40 |
| FNR, % | 0.04 | 0.03 | 0.02 | 0.02 | 0.01 | 0.00 | 0.02 | 0.02 |
| PPV, % | 0.85 | 0.84 | 0.84 | 0.84 | 0.91 | 0.89 | 0.87 | 0.87 |
| NPV, % | 0.88 | 0.90 | 0.92 | 0.92 | 0.94 | 1.00 | 0.90 | 0.93 |
| Se, % | 0.96 | 0.97 | 0.98 | 0.98 | 0.99 | 1.00 | 0.98 | 0.98 |
| Sp, % | 0.64 | 0.61 | 0.57 | 0.57 | 0.59 | 0.59 | 0.61 | 0.60 |
| Accuracy, % | 0.86 | 0.86 | 0.86 | 0.86 | 0.91 | 0.92 | 0.88 | 0.88 |
| 0.839 | 1.000 | 0.418 | 0.776 | |||||
| Non-diagnostic segment | 1 | 1 | 1 | 1 | 6 | 4 | 8 | 6 |
SE, smooth 3D+ combined with edge-2; FPR, False Positive Rate; FNR, False Negative Rate; PPV, Positive Predictive Value; NPV, Negative Predictive Value; Se, Sensitivity; Sp, Specificity; LAD, left anterior descending artery; LCX, left circumflex artery; RCA, right coronary artery; DSA, digital subtraction angiography.
Figure 4Comparison among the routine group, SE group, and digital subtraction angiography (DSA) group. (A) stenosis of the DSA group (red arrows point to the left anterior descending artery); (B,B-1), stenosis of the routine group; and (C,C-1), stenosis of the SE group.
Figure 5Images without using a combination of image filters in the routine group and images after using a combination of image filters in the SE (smooth 3D+ and edge-2) and Edge-3 (smooth 3D+ and edge-3) groups.