A whole-heart motion-correction algorithm: Effects on CT image quality and diagnostic accuracy of mechanical valve prosthesis abnormalities.

Young Joo Suh1, Young Jin Kim2, Jin Young Kim1, Suyon Chang1, Dong Jin Im1, Yoo Jin Hong1, Byoung Wook Choi1.   

Abstract

BACKGROUND: We aimed to determine the effect of a whole-heart motion-correction algorithm (new-generation snapshot freeze, NG SSF) on the image quality of cardiac computed tomography (CT) images in patients with mechanical valve prostheses compared to standard images without motion correction and to compare the diagnostic accuracy of NG SSF and standard CT image sets for the detection of prosthetic valve abnormalities.
METHODS: A total of 20 patients with 32 mechanical valves who underwent wide-coverage detector cardiac CT with single-heartbeat acquisition were included. The CT image quality for subvalvular (below the prosthesis) and valvular regions (valve leaflets) of mechanical valves was assessed by two observers on a four-point scale (1 = poor, 2 = fair, 3 = good, and 4 = excellent). Paired t-tests or Wilcoxon signed rank tests were used to compare image quality scores and the number of diagnostic phases (image quality score≥3) between the standard image sets and NG SSF image sets. Diagnostic performance for detection of prosthetic valve abnormalities was compared between two image sets with the final diagnosis set by re-operation or clinical findings as the standard reference.
RESULTS: NG SSF image sets had better image quality scores than standard image sets for both valvular and subvalvular regions (P < 0.05 for both). The number of phases that were of diagnostic image quality per patient was significantly greater in the NG SSF image set than standard image set for both valvular and subvalvular regions (P < 0.0001). Diagnostic performance of NG SSF image sets for the detection of prosthetic abnormalities (20 pannus and two paravalvular leaks) was greater than that of standard image sets (P < 0.05).
CONCLUSION: Application of NG SSF can improve CT image quality and diagnostic accuracy in patients with mechanical valves compared to standard images.
Copyright © 2017 Society of Cardiovascular Computed Tomography. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Artifacts; Computed tomography; Diagnostic accuracy; Heart valve prostheses; Image quality; Motion correction

Mesh:

Year:  2017        PMID: 28966103     DOI: 10.1016/j.jcct.2017.09.011

Source DB:  PubMed          Journal:  J Cardiovasc Comput Tomogr        ISSN: 1876-861X


  2 in total

1.  Second-generation motion correction algorithm improves diagnostic accuracy of single-beat coronary CT angiography in patients with increased heart rate.

Authors:  Junfu Liang; Ying Sun; Ziqing Ye; Yanchun Sun; Lei Xu; Zhen Zhou; Brian Thomsen; Jianying Li; Zhonghua Sun; Zhanming Fan
Journal:  Eur Radiol       Date:  2019-01-07       Impact factor: 5.315

2.  Intraindividual evaluation of effects of image filter function on image quality in coronary computed tomography angiography.

Authors:  Liang Jin; Pan Gao; Kun Wang; Jianying Li; Ming Li
Journal:  Front Cardiovasc Med       Date:  2022-09-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.