| Literature DB >> 36185802 |
Frank-André Siebert1, Markus Hirt1, Marc Delaperrière1, Jürgen Dunst1.
Abstract
Background and purpose: Risk management in radiotherapy is of high importance. There is not much data published on errors occurring in the treatment planning process of external beam techniques. The aim of this study was to investigate errors occurring during physics plan review in external beam radiotherapy. Materials and methods: Over a period of 14 months errors observed during the physical review process are reported. The errors were grouped and evaluated regarding treatment machine, technique, and treatment site. In addition, a correlation between frequency of errors and staff shortage was analyzed.Entities:
Keywords: Error detection; Quality assurance; Radiotherapy; Review process; Risk management
Year: 2022 PMID: 36185802 PMCID: PMC9519775 DOI: 10.1016/j.phro.2022.09.006
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Illustration of the treatment planning and review process in ARIA. The approval status of the treatment plans is also included as well as the person who is carrying out the approval. The dashed rectangle shows the portion of the plan review that is addressed in this study.
Short description of errors (g-errors and s-errors) observed in this study during the review process. In addition, the frequency (number of occurrences) of the errors are given for all errors with frequency > 1. In the last column it is marked by a “T” whether a technical barrier in the TPS for the error is already implemented. “(T)” indicates the errors that could be trapped by the TPS, but this had not been implemented at the time of this study.
| Description | Frequency | Barrier in TPS | |
|---|---|---|---|
| Wrong gantry angle at setup field | Two setup fields are defined at 0° and 90°/270°. If a gantry angle is not correct, the setup field will be modified prior acquisition. | 19 | (T) |
| Wrong dose/number of fractions | A wrong dose or number of fractions compared to prescription was entered in the TPS. This can result in over or under dose to the target volume and over dose to healthy tissues. | 5 | (T) |
| Wrong gantry angle | Wrong gantry angle means the gantry angle is erroneously deviating from a predefined planning technique like a box; e.g. 95° gantry angle instead of 90°. | 5 | (T) |
| MLC not conformal to planning target volume (PTV) | The MLC was not properly adjusted in regard to the PTV. | 2 | |
| Isocenter not defined or at false position | If the isocenter is not or falsely defined the TPS uses the Dicom center as isocenter. This means after patient setup using simulation markers, a wrong couch shift can be applied for the setup images. A wrong delivery of a treatment field will most likely not occur as the correct field position is checked by a setup field. | 2 | |
| Wrong field label | Treatment fields are labelled according to the gantry rotation. Even if this is incorrect the correct treatment will be applied. | 24 | |
| Wrong machine | The treatment plan is calculated for a wrong treatment machine. This can mean a swap between the two Artiste linacs (which are identical in regard of dosimetry) or a swap between Artiste and TrueBeamSTx. Incorrect treatment is not possible. This issue can result in a new treatment planning or affect the patient logistic. | 7 | |
| Setup field with MLC | In the considered clinic open setup fields without MLCs are created; in other clinics the MLC might be included in the setup field. To include an MLC in the setup field is a deviation from internal clinical rules and no g-error as the acquisition of the setup field with MLC would not be repeated with a setup field without MLC. | 7 | (T) |
| Isocenter position not rounded | Typically, in the considered workflow isocenter positions are rounded to half of a centimeter. To omit this is a violation of internal clinical rules but cannot result in wrong treatment. | 4 | |
| Primary reference point for IMRT/VMAT with location | In the considered treatment plans, normally a primary reference point with location and plan normalization (100 % dose) is used. For VMAT or IMRT and some 3D technique, the normalization is done differently (e.g. 100 % at target mean). In these cases, a location of the primary reference point might result in an incorrect dose reporting. | 4 | |
| No DRR for setup field created | DRRs (digital reconstructed radiographs) are created in the TPS and are linked to the setup field. If a DRR is not created this will be done at the console of the linac. | 3 | |
| Fields with zero dose rate | If a treatment field is defined without a dose-rate definition, a plan approval is not possible. | 3 | T |
| MLC exceeds physical constraints | Leaf positions and movements are restricted by physical constraints as maximum leaf overtravel and leaf span. In some cases, e.g. after manual adjustment of leaf positions, it can occur that these are not fully met after dose calculation. When such plans are approved an error message appears. A verification of MLC positions in the TPS is needed followed by new dose calculation. | 3 | T |
| Jaws not collimating the MLC field | For MLC fields the collimation by the jaws must match the MLC aperture to minimize dose leakage within the range of the leaf width. This is not the case in clinics where back-up jaws are set to the maximum aperture of all field segments. | 2 | |
| Old plan version printed | Treatment plans are still printed in the considered clinic for archiving and handout at the linac. If a wrong version was printed this cannot result in a wrong treatment, because the approved correct digital treatment plan is transferred to the linac and differences with the printed plan will attract attention. | 2 | |
| Bad printout quality | By bad printout quality of a treatment plan, like improper zoom settings in isodose representation, no wrong treatment can emerge as the correct plan parameters are used in ARIA. | 2 | |
| Not enough dose from field at primary reference point | The primary reference point receives a dose contribution from every field in the treatment plan. If the reference point is located far away from one or more treatment fields their dose contribution will be too low to be considered. A plan approval is not possible. In such cases either the location of the primary reference point has to be adjusted or the reference point is changed to a reference point without location. | 2 | T |
| Primary reference point undefined | Without a primary reference point plan approval is not possible. | 2 | T |
| Wrong energy for setup field | If a wrong energy of a setup field is set, the treatment plan can be approved, but the setup field cannot be acquired. | 2 | T |
Number of treatment plans and errors are listed with respect to treatment machines, treatment technique, and anatomical site. Errors are given in absolute numbers and in percent.
| Number of plans | Percentage of plans | Number of errors | Percentage of plans with errors | |
|---|---|---|---|---|
| 2x Artiste | 689 | 65 % | 67 | 10 % |
| 1x TrueBeamSTx | 367 | 35 % | 43 | 12 % |
| 3D | 767 | 73 % | 84 | 11 % |
| IMRT | 18 | 2 % | 1 | 6 % |
| VMAT | 220 | 21 % | 20 | 9 % |
| AMAT | 51 | 5 % | 5 | 10 % |
| Brain | 116 | 11 % | 8 | 7 % |
| H&N | 100 | 10 % | 12 | 12 % |
| Breast | 210 | 20 % | 29 | 14 % |
| Extremities | 43 | 4 % | 3 | 7 % |
| Chest | 274 | 26 % | 22 | 8 % |
| Pelvis (no prostate) | 195 | 19 % | 28 | 14 % |
| Prostate | 118 | 11 % | 8 | 7 % |
Fig. 2Percentage of errors in treatment plans per month. Bars of months in which more than 25% of the physics staff was not in the clinic (e.g. holiday period, sickness leave) are marked in dark grey.