Literature DB >> 36173238

The Insulin-Only Bionic Pancreas Pivotal Trial Extension Study: A Multi-Center Single-Arm Evaluation of the Insulin-Only Configuration of the Bionic Pancreas in Adults and Youth with Type 1 Diabetes.

Jane Lynch1, Lauren G Kanapka2, Steven J Russell3, Edward R Damiano4,5, Firas H El-Khatib4,5, Katrina J Ruedy2, Courtney Balliro3, Peter Calhoun2, Roy W Beck2.   

Abstract

Objective: To evaluate a transition from standard-of-care (SC) management of type 1 diabetes (any insulin delivery method including hybrid closed-loop systems plus real-time continuous glucose monitoring [CGM]) to use of the insulin-only configuration of the iLet® bionic pancreas (BP) in 90 adults and children (age 6-71 years). Research Design and
Methods: After the SC group completed the randomized controlled trial (RCT) portion of the Insulin-Only BP Pivotal Trial, 90 of the 107 participants participated in a 13-week study using the BP. The key outcomes were change from baseline in HbA1c and CGM metrics after 13 weeks on the BP.
Results: Using the BP, mean HbA1c decreased from 7.7% ± 1.0% (61 ± 10.9 mmol/mol) at baseline to 7.1% ± 0.6% (54 ± 6.6 mmol/mol) at 13 weeks (mean change -0.55% ± 0.72% [-6 ± 7.9 mmol/mol], P < 0.001), time in range 70-180 mg/dL increased by 12.0% ± 12.5% (from 53% ± 17% to 65% ± 9%, P < 0.001), and mean glucose decreased by -18 ± 23 mg/dL (from 182 ± 32 to 164 ± 15 mg/dL, P < 0.001). The higher the baseline HbA1c level, the greater the change in HbA1c. Results were similar in the adult (N = 42) and pediatric (N = 48) cohorts. Time <70 mg/dL decreased from baseline over the 13 weeks by -0.50% ± 1.86% (P = 0.02), and time <54 mg/dL was similar (change from baseline -0.08% ± 0.59%, P = 0.24). Two severe hypoglycemia events (in same participant) and one diabetic ketoacidosis event occurred. Conclusions: Glycemic control improved after adult and pediatric participants in the SC arm in the Insulin-Only BP Pivotal Trial transitioned to use of the BP. Improvement using the BP was of similar magnitude to that observed during the RCT. ClinicalTrials.gov number, NCT04200313.

Entities:  

Keywords:  adults; artificial pancreas; automated insulin delivery; bionic pancreas; evaluation; pediatrics; type 1 diabetes

Mesh:

Substances:

Year:  2022        PMID: 36173238      PMCID: PMC9529297          DOI: 10.1089/dia.2022.0341

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   7.337


  17 in total

1.  Safety of a Hybrid Closed-Loop Insulin Delivery System in Patients With Type 1 Diabetes.

Authors:  Richard M Bergenstal; Satish Garg; Stuart A Weinzimer; Bruce A Buckingham; Bruce W Bode; William V Tamborlane; Francine R Kaufman
Journal:  JAMA       Date:  2016-10-04       Impact factor: 56.272

2.  An Evaluation of Two Capillary Sample Collection Kits for Laboratory Measurement of HbA1c.

Authors:  Roy W Beck; Laura E Bocchino; John W Lum; Craig Kollman; Victoria Barnes-Lomen; Mark Sulik; Michael J Haller; Bruce Bode; Joseph T Cernich; Anthony A Killeen; Uttam Garg; David Liljenquist; Janey G Adams; Margaret Clements; Deanna Gabrielson; Terri Johnson; Mark A Clements
Journal:  Diabetes Technol Ther       Date:  2021-04-07       Impact factor: 6.118

3.  Positive Impact of the Bionic Pancreas on Diabetes Control in Youth 6-17 Years Old with Type 1 Diabetes: A Multicenter Randomized Trial.

Authors:  Laurel H Messer; Bruce A Buckingham; Fran Cogen; Mark Daniels; Greg Forlenza; Rabab Z Jafri; Nelly Mauras; Andrew Muir; R Paul Wadwa; Perrin C White; Steven J Russell; Edward R Damiano; Firas H El-Khatib; Katrina J Ruedy; Courtney A Balliro; Zoey Li; Martin Chase Marak; Peter Calhoun; Roy W Beck
Journal:  Diabetes Technol Ther       Date:  2022-10       Impact factor: 7.337

4.  A Multicenter Randomized Trial Evaluating the Insulin-Only Configuration of the Bionic Pancreas in Adults with Type 1 Diabetes.

Authors:  Davida Kruger; Alex Kass; Jacqueline Lonier; Jeremy Pettus; Philip Raskin; Maamoun Salam; Subbulaxmi Trikudanathan; Keren Zhou; Steven J Russell; Edward R Damiano; Firas H El-Khatib; Katrina J Ruedy; Courtney Balliro; Zoey Li; Martin Chase Marak; Peter Calhoun; Roy W Beck
Journal:  Diabetes Technol Ther       Date:  2022-10       Impact factor: 7.337

Review 5.  Current Status and Emerging Options for Automated Insulin Delivery Systems.

Authors:  Gregory P Forlenza; Rayhan A Lal
Journal:  Diabetes Technol Ther       Date:  2022-03-14       Impact factor: 7.337

6.  A Randomized Trial of Closed-Loop Control in Children with Type 1 Diabetes.

Authors:  Marc D Breton; Lauren G Kanapka; Roy W Beck; Laya Ekhlaspour; Gregory P Forlenza; Eda Cengiz; Melissa Schoelwer; Katrina J Ruedy; Emily Jost; Lori Carria; Emma Emory; Liana J Hsu; Mary Oliveri; Craig C Kollman; Betsy B Dokken; Stuart A Weinzimer; Mark D DeBoer; Bruce A Buckingham; Daniel Cherñavvsky; R Paul Wadwa
Journal:  N Engl J Med       Date:  2020-08-27       Impact factor: 91.245

7.  Glucose Outcomes with the In-Home Use of a Hybrid Closed-Loop Insulin Delivery System in Adolescents and Adults with Type 1 Diabetes.

Authors:  Satish K Garg; Stuart A Weinzimer; William V Tamborlane; Bruce A Buckingham; Bruce W Bode; Timothy S Bailey; Ronald L Brazg; Jacob Ilany; Robert H Slover; Stacey M Anderson; Richard M Bergenstal; Benyamin Grosman; Anirban Roy; Toni L Cordero; John Shin; Scott W Lee; Francine R Kaufman
Journal:  Diabetes Technol Ther       Date:  2017-01-30       Impact factor: 6.118

8.  A Pooled Analysis of Clinical Pharmacology Trials Investigating the Pharmacokinetic and Pharmacodynamic Characteristics of Fast-Acting Insulin Aspart in Adults with Type 1 Diabetes.

Authors:  Tim Heise; Thomas R Pieber; Thomas Danne; Lars Erichsen; Hanne Haahr
Journal:  Clin Pharmacokinet       Date:  2017-05       Impact factor: 6.447

9.  Safety Evaluation of the MiniMed 670G System in Children 7-13 Years of Age with Type 1 Diabetes.

Authors:  Gregory P Forlenza; Orit Pinhas-Hamiel; David R Liljenquist; Dorothy I Shulman; Timothy S Bailey; Bruce W Bode; Michael A Wood; Bruce A Buckingham; Kevin B Kaiserman; John Shin; Suiying Huang; Scott W Lee; Francine R Kaufman
Journal:  Diabetes Technol Ther       Date:  2018-12-26       Impact factor: 6.118

10.  Safety and Glycemic Outcomes During the MiniMed™ Advanced Hybrid Closed-Loop System Pivotal Trial in Adolescents and Adults with Type 1 Diabetes.

Authors:  Anders L Carlson; Jennifer L Sherr; Dorothy I Shulman; Satish K Garg; Rodica Pop-Busui; Bruce W Bode; David R Lilenquist; Ron L Brazg; Kevin B Kaiserman; Mark S Kipnes; James R Thrasher; John H Chip Reed; Robert H Slover; Athena Philis-Tsimikas; Mark Christiansen; Benyamin Grosman; Anirban Roy; Melissa Vella; Richard A M Jonkers; Xiaoxiao Chen; John Shin; Toni L Cordero; Scott W Lee; Andrew S Rhinehart; Robert A Vigersky
Journal:  Diabetes Technol Ther       Date:  2021-11-16       Impact factor: 6.118