| Literature DB >> 36168371 |
Nageswar Sahu1, Madhusmita Mohanty1, Amit K Adhya2.
Abstract
Background Despite many advances in platelet counting by cell counters, the problem of falsely low or falsely high total platelet counts (TPC) is common. Many laboratories estimate platelet count on the peripheral smear to cross-check the platelet counts. However, due to the lack of a standard calculation method, discrepant results are obtained from different laboratories leading to confusion among clinicians. We aimed to formulate a standard estimation method for platelet count on peripheral smear. Methodology In the first step (in 100 blood samples), we determined the ratio of the TPC obtained by the automated cell counter and the total number of platelets per oil immersion field (filed size: 0.22 mm) of the corresponding blood smears. The mean of the ratios thus obtained was designated as the "multiplication factor" to be used for visual platelet count estimation on the peripheral blood smear. In the subsequent step, validation of the same was done on another 100 samples. TPC on the peripheral smears of these samples was estimated using the above "multiplication factor" and compared with the corresponding TPC obtained on the automated cell counter. Results The "multiplication factor" obtained was 9.4 x 103 in the first set of 100 blood samples. It was used to estimate the platelet value of the second set of 100 blood samples. Conclusions We found an excellent agreement between the platelet counts obtained by automated cell counters and the estimation method. We suggest the multiplication factor 9.4 x 103 may be used with correction for microscopic field size to estimate platelet count on peripheral smears. This method is, however, not so reliable for very low platelet counts.Entities:
Keywords: field size; multiplication factor; peripheral smear; platelet estimation; total platelet count
Year: 2022 PMID: 36168371 PMCID: PMC9506600 DOI: 10.7759/cureus.28327
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Distribution of the total platelet count in the two sets of samples
TPC: total platelet count.
| 1st set of samples (n = 100): TPC by automated cell counter (PA1) | 2nd set of samples (n = 100): TPC by estimation method (PE) | 2nd set of samples (n = 100): TPC by automated cell counter (PA2) | |
| Lowest value of TPC (x103/µL) | 7 | 9.3 | 6.8 |
| Highest value of TPC (x103/µL) | 690 | 656.3 | 689 |
| No. of samples with TPC > 400 x 103/µL | 6 | 6 | 7 |
| No. of samples with TPC = 150-400 x 103/µL | 26 | 23 | 29 |
| No. of samples with TPC < 150 x 103/µL | 68 | 71 | 64 |
Correlation between the two methods in the second set of samples (n = 100)
| Total platelet count by automated cell counter | n (100) | Pearson’s correlation coefficient “R-value” (automated cell counter method vs. estimation method) | R2 coefficient of determination | P-value |
| >400 x 103/µL | 7 | 0.9699 | 0.9407 | 0.001345 |
| 150-400 x 103/µL | 29 | 0.9677 | 0.9364 | <0.00001 |
| <150 x 103/µL | 64 | 0.9692 | 0.9393 | <0.00001 |
| Total | 100 | 0.9971 | 0.9942 | <0.00001 |
Correlation between the two methods in the second set of samples with TPC < 150 x 103/µL
TPC: total platelet count.
| Total platelet count by automated cell counter | n (64) | Pearson’s correlation coefficient “R-value” (automated cell counter method vs. estimation method) | R2 coefficient of determination | P-value |
| 50-149.9 x 103/µL | 40 | 0.927 | 0.8599 | <0.00001 |
| 20-49.9 x 103/µL | 17 | 0.7135 | 0.5091 | 0.001299 |
| <20 x 103/µL | 7 | 0.6084 | 0.3702 | 0.147 |
Figure 1Scatter diagram showing the results of Pearson’s correlation coefficient test applied to compare the platelet counts obtained by the manual calculation method and by the automated cell counter
Figure 2Diagram depicting the results of the Bland-Altman plot
Most values are concentrated around the mean and within the 2SD.
TPC: total platelet count.