Literature DB >> 27974955

Reproducibility of Manual Platelet Estimation Following Automated Low Platelet Counts.

Zainab S Al-Hosni1, Murtadha Al-Khabori2, Sahimah Al-Mamari2, Jamal Al-Qasabi2, Hiedi Davis2, Hatim Al-Lawati3, Arwa Z Al-Riyami2.   

Abstract

OBJECTIVES: Manual platelet estimation is one of the methods used when automated platelet estimates are very low. However, the reproducibility of manual platelet estimation has not been adequately studied. We sought to assess the reproducibility of manual platelet estimation following automated low platelet counts and to evaluate the impact of the level of experience of the person counting on the reproducibility of manual platelet estimates.
METHODS: In this cross-sectional study, peripheral blood films of patients with platelet counts less than 100 × 109/L were retrieved and given to four raters to perform manual platelet estimation independently using a predefined method (average of platelet counts in 10 fields using 100× objective multiplied by 20). Data were analyzed using intraclass correlation coefficient (ICC) as a method of reproducibility assessment.
RESULTS: The ICC across the four raters was 0.840, indicating excellent agreement. The median difference of the two most experienced raters was 0 (range: -64 to 78). The level of platelet estimate by the least-experienced rater predicted the disagreement (p = 0.037). When assessing the difference between pairs of raters, there was no significant difference in the ICC (p = 0.420).
CONCLUSIONS: The agreement between different raters using manual platelet estimation was excellent. Further confirmation is necessary, with a prospective study using a gold standard method of platelet counts.

Entities:  

Keywords:  Blood Platelets; Examination; Interpretation; Peripheral Blood Smear; Reliability

Year:  2016        PMID: 27974955      PMCID: PMC5099402          DOI: 10.5001/omj.2016.83

Source DB:  PubMed          Journal:  Oman Med J        ISSN: 1999-768X


  6 in total

1.  The most accurate platelet count on the Sysmex XE-2100. Optical or impedance?

Authors:  C Briggs; S Kunka; S J Machin
Journal:  Clin Lab Haematol       Date:  2004-04

2.  The accuracy of platelet counting in thrombocytopenic blood samples distributed by the UK National External Quality Assessment Scheme for General Haematology.

Authors:  Barbara J De la Salle; Paul N McTaggart; Carol Briggs; Paul Harrison; Caroline J Doré; Ian Longair; Samuel J Machin; Keith Hyde
Journal:  Am J Clin Pathol       Date:  2012-01       Impact factor: 2.493

3.  Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion.

Authors:  H C Segal; C Briggs; S Kunka; A Casbard; P Harrison; S J Machin; M F Murphy
Journal:  Br J Haematol       Date:  2005-02       Impact factor: 6.998

4.  Platelet count evaluation using three automated haematology analysers compared with the immunoplatelet reference method, and estimation of possible inadequate platelet transfusion.

Authors:  K H Hong; M J Kim; K W Lee; K U Park; H S Kim; J Song
Journal:  Int J Lab Hematol       Date:  2008-02-21       Impact factor: 2.877

5.  Comparison of automated platelet counts and potential effect on transfusion decisions in cancer patients.

Authors:  Steven Marionneaux; Nenita Francisco; Virgil Chan; Justin Hanenberg; Josephine Rafael; Conchita Chua; Rachel Jia; Jingqui Yao; Jean Lynch; Vicky Chan; Peter Maslak
Journal:  Am J Clin Pathol       Date:  2013-11       Impact factor: 2.493

6.  Is automated platelet counting still a problem in thrombocytopenic blood?

Authors:  Raimundo Antônio Gomes Oliveira; Maria Mariko Takadachi; Kimiyo Nonoyama; Orlando César de Oliveira Barretto
Journal:  Sao Paulo Med J       Date:  2003-07-04       Impact factor: 1.044

  6 in total
  2 in total

1.  Comparison of serological assessments in the diagnosis of liver fibrosis in bile duct ligation mice.

Authors:  Chengxia Xie; Bo Ma; Ning Wang; Lin Wan
Journal:  Exp Biol Med (Maywood)       Date:  2017-07-01

2.  Estimation of Total Platelet Count From Peripheral Blood Smear Needs a Correction Factor.

Authors:  Nageswar Sahu; Madhusmita Mohanty; Amit K Adhya
Journal:  Cureus       Date:  2022-08-23
  2 in total