Amanda K Kussrow1, Michael N Kammer2, Pierre P Massion2, Rebekah Webster1, Darryl J Bornhop1. 1. Department of Chemistry and The Vanderbilt Institute for Chemical Biology, Vanderbilt University, Nashville, Tennessee 37235, United States. 2. Division of Allergy, Pulmonary and Critical Care Medicine and Vanderbilt-Ingram Cancer Center, Vanderbilt University, Nashville, Tennessee 37235, United States.
Abstract
CYFRA 21.1, a cytokeratin fragment of epithelial origin, has long been a valuable blood-based biomarker. As with most biomarkers, the clinical diagnostic value of CYFRA 21.1 is dependent on the quantitative performance of the assay. Looking toward translation, it is shown here that a free-solution assay (FSA) coupled with a compensated interferometric reader (CIR) can be used to provide excellent analytical performance in quantifying CYFRA 21.1 in patient serum samples. This report focuses on the analytical performance of the high-sensitivity (hs)-CYFRA 21.1 assay in the context of quantifying the biomarker in two indeterminate pulmonary nodule (IPN) patient cohorts totaling 179 patients. Each of the ten assay calibrations consisted of 6 concentrations, each run as 7 replicates (e.g., 10 × 6 × 7 data points) and were performed on two different instruments by two different operators. Coefficients of variation (CVs) for the hs-CYFRA 21.1 analytical figures of merit, limit of quantification (LOQ) of ca. 60 pg/mL, B max, initial slope, probe-target binding affinity, and reproducibility of quantifying an unknown were found to range from 2.5 to 8.3%. Our results demonstrate the excellent performance of our FSA-CIR hs-CYFRA 21-1 assay and a proof of concept for potentially redefining the performance characteristics of this existing important candidate biomarker.
CYFRA 21.1, a cytokeratin fragment of epithelial origin, has long been a valuable blood-based biomarker. As with most biomarkers, the clinical diagnostic value of CYFRA 21.1 is dependent on the quantitative performance of the assay. Looking toward translation, it is shown here that a free-solution assay (FSA) coupled with a compensated interferometric reader (CIR) can be used to provide excellent analytical performance in quantifying CYFRA 21.1 in patient serum samples. This report focuses on the analytical performance of the high-sensitivity (hs)-CYFRA 21.1 assay in the context of quantifying the biomarker in two indeterminate pulmonary nodule (IPN) patient cohorts totaling 179 patients. Each of the ten assay calibrations consisted of 6 concentrations, each run as 7 replicates (e.g., 10 × 6 × 7 data points) and were performed on two different instruments by two different operators. Coefficients of variation (CVs) for the hs-CYFRA 21.1 analytical figures of merit, limit of quantification (LOQ) of ca. 60 pg/mL, B max, initial slope, probe-target binding affinity, and reproducibility of quantifying an unknown were found to range from 2.5 to 8.3%. Our results demonstrate the excellent performance of our FSA-CIR hs-CYFRA 21-1 assay and a proof of concept for potentially redefining the performance characteristics of this existing important candidate biomarker.
The quantification of protein biomarkers
at physiologically relevant
levels in patient serum samples is essential across the continuum
of patient care, from diagnostics to response to therapy. Among the
major contributors to the clinical translation bottleneck for biomarker
assays is the intrinsic biological variability in large cohorts of
samples for systemic biomarkers, the relatively long development time
for assays, and the need for assays with improved sensitivity.[1]Many approaches exist to quantify
biomarkers at high sensitivity,
with most depending heavily on the ability to quantify molecular interactions.
These include enzyme-linked immunosorbent assay (ELISA), bead array
technologies, and mass spectrometry (MS).[2−8] Some new methods report single-molecule sensitivity[9] and, as we have shown,[10] have
the potential to impact clinical practice due to their low limit of
quantification (LOQ).[11] While the free-solution
assay (FSA), combined with the compensated interferometric reader
(CIR), is not a single-molecule approach, it does represent a simpler
(mix-and-read, label-free) approach to obtain picogram/mL sensitivity.[10] “Single molecule” techniques employ
a fluorescence assay based on multiple chemical steps and relatively
complicated optical readers. In Singulex, the probe volume is limited
to a few femtoliters in a manner similar to confocal microscopy. Combining
this optical approach with a fluorescent sandwich assay that uses
a capture and separation step provides excellent signal-to-noise (S/N).[9] The Simoa (Single-Molecule Array) from Quanterix
technology exploits the advantages of digital assays by employing
fluorescence-labeled sandwich beads, which are each collected in wells
formed at the end of a coherent fiber optic or similar small volume
receptacle.[12,13] Both of these techniques are
promising but still are somewhat limited by speed, reproducibility,
cost, and/or accessibility.SOMALOGIC has taken a different
approach to quantify serum proteins
by employing aptamers (strands of DNA or RNA).[14] Their detection approach capitalizes on a slow “off-rate”
for one of the protein–aptamer complexes formed to help separate
the sample from the background.[15] This
aptamer-probe method has shown promise for high-sensitivity (hs) protein
quantification,[16] yet multiple (ca. 10)
sample handling and labeling steps, combined with relatively complicated
instrumentation, have impeded the wide dissemination of the technology
for biomarker quantification.Label-free techniques such as
surface plasmon resonance (SPR),
quartz crystal microbalance, wave-guided interferometry, and mass
spectrometry (MS) have been used for biomarker quantification.[17−19] While MS has been exceedingly valuable in the biomarker discovery
phase,[20−23] relative instrument complexity and difficulty with quantification
make its use in clinical validation unattractive. Multiplexed MRM/MS
targeted assays using stable isotope-labeled peptide standards for
quantitation show promise as clinical diagnostic assays, yet low-throughput
remains a challenge.[23−25]In label-free methods, surface immobilization
can negatively influence
assay development time and sensitivity in complex matrices.[26−29] Actually, the same is true for the most common methods used today
for biomarker analysis, such as ELISA and/or bead arrays, which also
require chemical modification for surface immobilization and/or labeling
steps that can influence assay performance. Significant strides have
been made toward miniaturizing, multiplexing, and improving ELISA-like
assays.[13,30] One example is the electrochemiluminescent
(ECL) assay, which is typically 10-fold more sensitive than the standard
fluorescent analogue.[31] Yet, reaching low
picomolar or high femtomolar detection limits often requires amplification
chemistries that can be costly and require lengthy development times.
These limitations tend to extend the interval between biomarker discovery
and clinical validation. Furthermore, the large sample consumption
associated with some of these methods impedes validation of promising
biomarkers due to the precious nature of banked samples on relevant
patient populations. Therefore, the volume-constrained, FSA method
represents an attractive alternative to ELISA.Sensitivity and
specificity of all interaction-based determinations
are influenced by numerous factors. These include the equilibrium
binding affinity, the transduction mechanism, and the level of background
(matrix and/or spectroscopic). In general, the higher the magnitude
of the equilibrium binding affinity (the lower the true-KD), the lower the potential limit of detection. In fluorescence
and absorbance determinations, performance is also impacted by quantum
efficiencies, molar absorptivity, Stokes shifts, and other nonspectroscopic
metrics for the analyte. FSA is therefore a complementary approach,
with the signal transduction parameter dictating assay performance
and the magnitude of change in dipole moment (e.g., refractive index,
RI) due to conformation/hydration changes during the binding event.Since lung cancer is the leading cause of cancer-related deaths
in the United States, there is a growing movement to improve diagnostics.
One approach is to implement low-dose chest computed tomography (CT)
screening into routine practice. CT screening has been endorsed by
U.S. Preventive Services Task Force, the American Thoracic Society,
the American College of Chest Physicians, and reimbursement payers.
While these programs can reduce lung cancer-specific mortality by
20% and overall mortality by 6.9% in high-risk individuals, numerous
challenges still exist to realize early detection and better outcomes
for the majority of lung cancer patients. Improved biomarker assays
could address these challenges. So, (a) how to position biomarker
use prior to or alongside chest CT screening to decrease the cost
and rates of false-positive tests, and (b) how to diagnose lung cancer
patients with indeterminate pulmonary nodules (IPNs)?Recently,
as with the development of the high-sensitivity (hs)-CRP
assay for C-reactive protein (CRP),[32] we
postulated that lowering the LOQ for CYFRA 21.1 can improve risk stratification
within the context of noninvasive diagnosis of IPNs.[10,33] FSA-CIR represents the only solution-phase, label-free molecular
interaction measurement methodology with sensitivity comparable to
or better than fluorescence[10,34,35] that is compatible with a wide range of complex matrices,[36−38] and that offers good throughput and constrained sample consumption.[10,34] FSA-CIR can be used to address numerous previously challenging or
intractable problems, ranging from poor in vitro in vivo correlations for first in human dosing[38] to the rapid quantification of low abundance chemicals in human
samples, including neonatal opioids in urine and chemical nerve agents
in urine and serum.[34,35] Capitalizing on the unique properties
of FSA-CIR, we demonstrated a ∼10-fold LOQ improvement over
ECL for the biomarker CYFRA 21.1.[10] This
observation led to a larger study that showed there is promise for
improving management of patients with IPNs using a CYFRA 21.1 assay,
with improved LOQs.[33]Here, we report
on the long-term analytical performance of using
the FSA and CIR[34,35,39] for quantifying CYFRA 21.1 in serum. We illustrate that CVs for
the analytical figures of merit of the CYFRA 21.1 assay, LOQ, Bmax, initial slope, probe–target apparent
binding affinity, and reproducibility of quantifying an unknown range
from 2.5 to 8.3%. These metrics of performance were obtained over
the course of 16 days, with experiments run on 5 days while constructing
10 calibration curves and analyzing 2 unknowns (operator-blinded spiked
samples) per calibration curve. The calibration tests were run in
the context of studying two IPN patient cohorts, totaling 179 patients.
Interestingly, our results came from a nanoliter-volume, universal,
mix-and-read, solution-phase, enzyme amplification-free assay method
that exhibits no relative mass sensitivity on serum and uses a reader
with a simple optical train that does not require high-resolution
temperature control.[39] These collective
properties could pave the way for the use of FSA-CIR in the management
of IPN patients in a clinical lab or the near-patient setting.
Materials and Methods
The Assay methodology has been
described in detail previously[10] and is
described briefly here. The measurements
were performed on the compensated interferometer, as described previously.[34,39−41] All lab disposables were purchased from Thermo Fisher
Scientific, and reagents were purchased from Sigma-Aldrich. All aqueous
solutions were prepared using deionized water. Pooled human serum
was purchased from Valley Biomedical (Winchester, VA). The protein
standard (CYFRA 21-1) was purchased from DRG International (Springfield,
NJ). A CYFRA 21-1 monoclonal antibody (clone XC4, product #MBS850246)
obtained from MyBioSource (San Diego, CA) was used as the probe.
Assay Preparation
Calibration solutions were prepared
by spiking several aliquots of a 50% pooled human serum/50% phosphate-buffered
saline (PBS) solution with concentrations of the biomarker target
CYFRA 21-1 ranging from 0.08 to 25 ng/mL. To produce the binding sample,
20 μL of each serum aliquot was combined with 20 μL of
a solution containing 2 μg/mL of the probe antibody. To produce
an RI-matched, nonbinding reference sample, 20 μL of each serum
aliquot was combined with 20 μL of a blank solution containing
PBS devoid of antibodies. Final sample compositions were 25% serum
in PBS, with CYFRA 21-1 concentrations of 0, 0.04, 0.10, 0.50, 2.50,
and 12.50 ng/mL. The binding samples contained 1 μg/mL of the
probe antibody, and the reference samples were devoid of the probe.Following a 1 h incubation at ambient temperature (22 °C)
on a shaker (300 rpm), the sample and reference solutions were loaded
onto the droplet generator in the order of increasing concentration
and the phase shift between reference and binding samples was measured
by the CIR. A calibration curve was created by fitting the resulting
phase shift to a single site saturation isotherm according to the
equation Y = Bmax × X/(KD + X), where Y is the phase shift at the concentration of X (in ng/mL). Bmax and KD are fitted
to the data by nonlinear regression, where Bmax is the maximum signal at saturation and KD is the apparent dissociation constant. The initial slope
of the calibration curve was determined by dividing Bmax by KD. The initial slope
was used to calculate the assay limit of detection by 3 × (standard
deviation of 5 s instrument baseline)/(initial slope), and the assay
limit of quantification by 3 × (standard deviation of replicate
measurements of the same sample)/(initial slope). The value used for X was taken as the final CYFRA 21-1 concentration in 25%
serum (listed above) multiplied by 4 to recover the concentration
of CYFRA 21-1 in 100% serum. This calibration curve was prepared from
stock reagents independently for each assay.Recovery was assessed
by preparing spiked “unknowns.”
These test samples were prepared by adding a known quantity of the
CYFRA 21-1 to pooled serum and then processing the samples, as described
above. The biomarker concentration was recovered by fitting the signal
to the calibration curve.
Results and Discussion
The high-sensitivity analysis
of serum protein reported here is
enabled by marrying two synergistic technologies, a solution-phase
assay and a compensated interferometer. CIR, shown in Figure and described in detail elsewhere,[34,35,39,41] consists of three components, an optical engine, a droplet generator,
and a syringe pump. The optical engine has a diode laser, an object
(polyimide-coated fused silica capillary tube, 250 × 350 μm),
and a camera (Figure ). Two nearly identical interferometers[41] result from spreading out the interrogation beam along the capillary
and reading adjacent regions of the resulting interference fringes.
The difference measurement facilitates high-resolution, low-volume
refractive index (RI) measurements. Proper laser–capillary
alignment and fringe pattern selection allow the elimination of the
high-resolution temperature controller typically needed for such devices.[41] The relative RI measurement of reference-and-sample
droplet pairs, separated by a 40 nL oil droplet, is accomplished by
measuring positional fringe shifts in adjacent windows with a Fourier
transform.[40,42] The fast Fourier transform (FFT)
readout reports the positional fringe shift as a phase change. The
polyimide-coated capillary tube serves as the main optical component
of the optical train while providing seamless droplet train transfer.
Use of a capillary is noteworthy because it provides improved sensitivity
and S/N over chip-based optics.[43,44]
Figure 1
Block diagram CRI, illustrating
sample flow from the droplet generator
(top) through the laser and fringe detector (center) and to the syringe
pump (bottom).
Block diagram CRI, illustrating
sample flow from the droplet generator
(top) through the laser and fringe detector (center) and to the syringe
pump (bottom).Use of a modified droplet generator (Mitos Dropix)
upstream and
a syringe pump (Harvard Apparatus, Holliston, Massachusetts) downstream
from the compensated interferometer allows for the smooth introduction
of the sample–reference droplet train. To mitigate nonspecific
adsorption and reduce noise due to plastic leaching, sample well plates
were made of highly biocompatible poly(ethyl ethyl ketone) (PEEK)
resin.Accurate patient CYFRA 21.1 quantification is contingent
on having
high-quality, reproducible, free-solution assay calibrations. These
calibration curves provide analytical figures of merit, including
the response function of the instrument and the quality of the assay
chemistry. These standards are performed throughout the day intermixed
with the patient’s CYFRA determinations. Such a procedure ensures
performance stability throughout determinations, mitigates bias, and
has been described previously.[10]Sensing of adjacent sample–reference droplets in a train
provides numerous advantages, including complex matrix compatibility
(specificity) and sensitivity that rival or exceed fluorescent assays.[41,45]Figure illustrates
the FSA biomarker workflow. A small volume of serum is split into
two aliquots and then processed to provide “binding”
and “reference” solutions. To quantify a biomarker target,
we add an excess antibody probe to one of these aliquots, giving the
“binding/test” sample. To the other aliquot, we add
an RI matching solution without the probe. These solutions equilibrate
for ∼1 h and are then introduced into adjacent wells of the
droplet generator for analysis by the interferometer (as pairs separated
by an oil droplet). The difference in signal between the sample–reference
pair provides a quantitative measure of the antibody-target complex
while allowing the matrix signal to be nullified. As previously described,[40] the scientific principle for FSA depends on
binding-induced changes in molecular conformation and hydration, producing
predictable and reproducible changes to the solution RI.
Figure 2
Schematic of
the free-solution assay method.
Schematic of
the free-solution assay method.Our evaluation of the analytical performance of
the CYFRA 21-1
assay in 25% serum is presented in the context of evaluating IPN patient
samples.[33] We employed a commercial antibody
(Ab) to the target biomarker, CYRFA 21.1 (MyBioSource, San Diego).
Characterization and selection of this Ab involved a simple screening
experiment with several Ab’s, as described in detail elsewhere.[10,34,35,46] This experiment allows confirmation of antibody-target binding and
signal-to-noise (S/N) maximization. Once selected, the Ab is further
characterized with a saturation isotherm determination to evaluate
the quality of the molecular interaction and the binding affinity
and obtain an estimate of the LOQ. While true-KD determinations are used to “guide chemistry selection”
during the assay development phase, our binding assay calibration
curve represents an apparent KD (app-KD). This parameter serves both to define the
response function for the assay and as a proxy for the fidelity of
the molecular interaction or KD (affinity).
Hence, the app-KD is obtained from calibrations
and used as a key analytical figure of merit for the assay chemistry
performance. Unlike fluorescence or absorption measurements, FSA transduces
the molecular interaction between the probe and the target directly.
Therefore, any variation or reduction in apparent binding affinity
can influence the S/N for quantifying the target. Thus, the app-KD obtained from response curve measurements
provides a metric of probe Ab and target/standard quality. A full
calibration curve (including unknowns) is performed with each batch
of about 20 patient serum samples that are processed.Accurate
quantification of any analyte, regardless of the method,
requires the determination of the instrumental response as a function
of target concentration, preferably in the matrix of interest. FSA
is no different. In the case of FSA-CIR, calibration curves are built
in the target matrix and analyzed as sample and reference droplet
trains. Specifically, samples of 25% serum containing 80 pg/mL to
50 ng/mL of the target protein (CYFRA 21-1) and the probe antibody
are analyzed by comparing the response vs matrix-matched reference
droplets and those consisting of identical concentrations of CYFRA
21-1 without the Ab (Figure ).Figure presents
a summary of 10 calibration determinations performed in 25% serum
using CYFRA 21.1 standards and an excess of antibody probe. Figure illustrates the
high quality and reproducibility of CYFRA 21.1 calibration curves
run on FSA-CIR over multiple days with two different operators working
with two different instruments. In the context of quantifying the
CYFRA 21.1 protein biomarker in two patient cohorts, the FSA-CIR calibration
methodology reported an average LOQ of 61 pg/mL with a standard deviation
of 3.0 pg/mL for 10 determinations with 6 droplets at 6 concentrations.
Correlation coefficients for these calibration curves ranged from
an R2 = 0.981 to 0.999.
Figure 3
Representative replicate
response function/app-KD curve. Antibody
at 1 μg/mL. Error bars representing
the standard deviations for 6 determinations are present but not legible.
Representative replicate
response function/app-KD curve. Antibody
at 1 μg/mL. Error bars representing
the standard deviations for 6 determinations are present but not legible.Assay accuracy/recovery is validated by preparing
two unknowns
with every calibration curve. These samples are made by spiking serum
with CYFRA 21-1, blinding the instrument operator to sample identity
and experimentally determining the biomarker concentration. Figure shows that FSA-CIR
provides excellent recovery at both 500 pg/mL and 1.5 ng/mL. The results
demonstrate the method is highly accurate and quantitative, with the
error in correctly determining the CYFRA 21.1 concentration ranging
from 0.25 to 13.6% for unknowns. Overall, within the operating range
of the calibration curve (80 pg/mL to 10 ng/mL), FSA-CIR provided
an average percent difference between actual and determined concentration
(recovery level) of 6.02%.
Figure 4
Plot (A) and (B) present the recovery values
and, (C) and (D),
the accuracy (percent error) for quantifying CYFRA 21-1 with FSA-CIR.
Plot (A) and (B) present the recovery values
and, (C) and (D),
the accuracy (percent error) for quantifying CYFRA 21-1 with FSA-CIR.Further evidence of the performance of FSA-CIR
is seen in the LOQ
data. As shown in Figure A, the LOQ of the 10 calibration determinations, run in the
context of patient sample determinations, ranged from 57.0 to 66.7
pg/mL, giving a CV of 4.9%. These LOQ values are ∼1.5-fold
better than the LOQ of 80 pg/mL,[47] published
recently for a somewhat complicated, chemically intensive immunoassay,
and ∼10-fold better than the ∼500 pg/mL LOQ for the
commercial, gold standard Roche Cobas electro-chemiluminescence assay
(ECLIA) for CYFRA 21-1.[48]
Figure 5
Summary of the assay
figures of merit: (A) Limit of quantification,
(B) saturation binding signal, (C) apparent KD, (D) pooled standard deviation, (E) initial slope of the
saturation isotherm, and (F) correlation coefficient.
Summary of the assay
figures of merit: (A) Limit of quantification,
(B) saturation binding signal, (C) apparent KD, (D) pooled standard deviation, (E) initial slope of the
saturation isotherm, and (F) correlation coefficient.Additional analytical FOM measurements for the
10 calibrations
(two independent instruments and operators) performed on 5 subsequent
days in 25% serum are extracted from the calibration curves. These
metrics include the Bmax (Figure B), app-KD (Figure C),
pooled standard deviation (Figure D), initial slope (Figure E), and the correlation coefficient (Figure F). Here, for ten
discrete determinations, maximum response, Bmax, ranged from 6141 milliradians to 7044 milliradians, exhibiting
a CV of 4.7%. The initial slope for the FSA-CIR calibration curves
was found to be 2582 mrad/ng/mL to 2948 mrad/ng/mL, with a CV of 4.1%,
and the app-KD spanned from 2.12 to 2.73
ng/mL, with a CV of just 8.7%. The pooled standard deviation of the
calibration measurements was determined to be 56.4–61.0 mrad,
with a CV of only 2.5%. It is noteworthy that each data point displayed
in Figure represents
an entire calibration run consisting of 6 sample and reference pair
measurements at each of 6 concentrations (including a zero value).
Discussion
In this report, it was shown that a mix-and-read,
free-solution
assay combined with a compensated interferometer enables the quantification
of CYFRA 21.1 in serum samples by two independent instruments/operators
over the course of 16 days, with 5 days of experimentation. The result
was excellent LOQs, precision, and accuracy. The assay employs a single
monoclonal antibody (IgG1 anti-CK19 clone XC4 (MyBioSource, San Diego,
CA)) and capitalizes on a unique label-free, signal transduction method
based on binding-induced solution-phase conformation and hydration
changes the free-solution response function (FreeSRF).[40] FSA is label-free (no fluorescence or radiolabeling),
making it rapid, cost-effective, and allowing the use of unaltered
or minimally processed patient samples. It is also assay agnostic,
allowing for quantitation of a wide array of interactions (antibodies
to DNA to small molecules) in a mass-independent manner.The
compensated interferometer is a unique biosensor that capitalizes
on an adjacent sample–reference configuration for matrix-insensitive
operation and assay specificity. Analysis can be effectively performed
on constrained volumes (<10 μL of patient serum), facilitating
multiple replicates to be performed on quantity-limited samples. The
optical engine of CIR is simple, consisting of a diode laser, a capillary
tube, and a camera. CIR is also among the most sensitive nanoliter-volume
universal sensors. Under conditions reported here, the instrument
performed at a baseline noise level of <10–7 RIU,
enabling a LOQ for CYFRA 21.1 of 61 pg/mL in 25% serum. At this LOQ,
in the probe volume of 40 nL defined by the laser–capillary
interaction, there is just 610 attograms (∼4 zeptomoles) of
the target-probe present. Overall, FOM CVs range from 2.5 to 8.7%,
with no CV exceeding 9%.The data shown here was generated in
the context of an ongoing
multisite prospectively collected, retrospective blinded evaluation
(PROBE) design trail[49] to determine the
clinical utility of adding hs-CFRA 21.1 by FSA-CIR as a biomarker
to the clinical problem of discrimination of cases from controls in
lung cancer patients with indeterminate pulmonary nodules (IPNs).[33] Here, our focus was on the analytical performance
of the assay methodology to demonstrate that a manual, first-generation
prototype instrument provides excellent performance over multiple
days of biomarker analysis research. We are acutely aware that to
have an ongoing impact in the clinical setting, refinement of our
instrument and assay will be necessary. Thus, to enable FSA-CIR translation,
we are currently working on a next-generation CIR with full automation,
from alignment to data analysis. Should our approach withstand more
stringent validation, FSA-CIR could provide improved noninvasive testing
for the clinical management of patients with lung cancer and other
diseases.The unique nature of FSA also represents numerous
opportunities
to rapidly develop and characterize new assays. For example, preliminary
validation tests look promising for using FSA-CIR to quantify the
serum biomarkers HE-4 and CEA. Further, because FSA is assay agnostic,
various types of probes, including DNA/RNA aptamers and small molecules,
all represent opportunities.[34,35] The unique transduction
method of FSA could represent a way forward in reducing the biomarker
validation bottleneck and expediting the clinical translation of new
biomarkers of disease.
Authors: Michael N Kammer; Ian R Olmsted; Amanda K Kussrow; Mark J Morris; George W Jackson; Darryl J Bornhop Journal: Analyst Date: 2014-11-21 Impact factor: 4.616
Authors: Neil N Trivedi; Mehrdad Arjomandi; James K Brown; Tess Rubenstein; Abigail D Rostykus; Stephanie Esposito; Eden Axler; Mike Beggs; Heng Yu; Luis Carbonell; Alice Juang; Sandy Kamer; Bhavin Patel; Shan Wang; Amanda L Fish; Zaid Haddad; Alan Hb Wu Journal: Biomed Res Clin Pract Date: 2018-10-29
Authors: David M Rissin; Cheuk W Kan; Todd G Campbell; Stuart C Howes; David R Fournier; Linan Song; Tomasz Piech; Purvish P Patel; Lei Chang; Andrew J Rivnak; Evan P Ferrell; Jeffrey D Randall; Gail K Provuncher; David R Walt; David C Duffy Journal: Nat Biotechnol Date: 2010-05-23 Impact factor: 54.908
Authors: Theresa M Russell; Louis S Green; Taylor Rice; Nicole A Kruh-Garcia; Karen Dobos; Mary A De Groote; Thomas Hraha; David G Sterling; Nebojsa Janjic; Urs A Ochsner Journal: J Clin Microbiol Date: 2017-08-09 Impact factor: 5.948