RF probe recovery time reduction with a novel active ringing suppression circuit.

A S Peshkovsky1, J Forguez, L Cerioni, D J Pusiol.   

Abstract

A simple Q-damper device for active probe recovery time reduction is introduced along with a straightforward technique for the circuit's component value optimization. The device is inductively coupled to a probe through a coupling transformer positioned away from the main coil, which makes the design independent of the coil type being used. The Q-damper is a tuned circuit, which is resonant at the same frequency as the probe and can be actively interrupted. When the circuit is interrupted, it is detuned and, thereby, is uncoupled from the probe, which operates normally. Turning the device on leads to re-coupling of the circuits and causes splitting of the probe's resonance line, which can be observed through its drive port. A resistance of an appropriate value is introduced into the Q-damper circuit, resulting in smoothing of the resonance splitting into one broad line, representing the coupled system's low-Q state, in which the energy stored in the main coil is efficiently dissipated. The circuit's component values are optimized by monitoring the shape of this low-Q state. Probe recovery time reduction by, approximately, an order of magnitude has been obtained with this device. Application of the device during an NQR experiment led to an increase in the signal-to-noise ratio by a factor of 4.9.

Mesh:

Substances:

Year:  2005        PMID: 16111906      PMCID: PMC1479304          DOI: 10.1016/j.jmr.2005.07.004

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  4 in total

1.  Feedback optimization of pulse width in the SORC sequence.

Authors:  J L Schiano; T Routhier; A J Blauch; M D Ginsberg
Journal:  J Magn Reson       Date:  1999-09       Impact factor: 2.229

2.  Fast recovery, high sensitivity NMR probe and preamplifier for low frequencies.

Authors:  D I Hoult
Journal:  Rev Sci Instrum       Date:  1979-02       Impact factor: 1.523

3.  Low-field NMR determinations of the properties of heavy oils and water-in-oil emulsions.

Authors:  G A LaTorraca; K J Dunn; P R Webber; R M Carlson
Journal:  Magn Reson Imaging       Date:  1998 Jun-Jul       Impact factor: 2.546

4.  Biological magnetic resonance imaging using laser-polarized 129Xe.

Authors:  M S Albert; G D Cates; B Driehuys; W Happer; B Saam; C S Springer; A Wishnia
Journal:  Nature       Date:  1994-07-21       Impact factor: 49.962

  4 in total
  1 in total

1.  Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI.

Authors:  Yonghyun Ha; Kartiga Selvaganesan; Baosong Wu; Kasey Hancock; Charles Rogers; Sajad Hosseinnezhadian; Gigi Galiana; R Todd Constable
Journal:  PLoS One       Date:  2022-09-16       Impact factor: 3.752

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.