PURPOSE: The purpose of this study was to analyze the marginal fit of three-unit resin prostheses printed with the stereolithography (SLA) method in two build orientations (45°, 60°) and two layer thicknesses (50 µm, 100 µm). MATERIALS AND METHODS: A master model for a three-unit resin prosthesis was designed with two implant abutments. Forty specimens were printed using an SLA 3D printer. The specimens were printed with two build orientations (45°, 60°), and each orientation was printed with two layer thicknesses (50 µm, 100 µm). The marginal fit was measured as the marginal gap (MG) and absolute marginal discrepancy (AMD), and MG and AMD measurements were performed at 8 points per abutment, for 16 points per specimen. All statistical analyses were performed using SPSS software. Two-way analysis of variance (ANOVA) was separately performed on the MG and AMD values of the build orientations and layer thicknesses. Moreover, one-way ANOVA was performed for each point within each group. RESULTS: The margins of the area adjacent to the pontic showed significantly high values, and the values were smaller when the build orientation was 45° than when it was 60°. However, the margin did not differ significantly according to the layer thicknesses. CONCLUSION: The marginal fit of the three-unit resin prosthesis fabricated by the SLA 3D method was affected by the pontic. Moreover, the marginal fit was affected by the build orientation. The 45° build orientation is recommended.
PURPOSE: The purpose of this study was to analyze the marginal fit of three-unit resin prostheses printed with the stereolithography (SLA) method in two build orientations (45°, 60°) and two layer thicknesses (50 µm, 100 µm). MATERIALS AND METHODS: A master model for a three-unit resin prosthesis was designed with two implant abutments. Forty specimens were printed using an SLA 3D printer. The specimens were printed with two build orientations (45°, 60°), and each orientation was printed with two layer thicknesses (50 µm, 100 µm). The marginal fit was measured as the marginal gap (MG) and absolute marginal discrepancy (AMD), and MG and AMD measurements were performed at 8 points per abutment, for 16 points per specimen. All statistical analyses were performed using SPSS software. Two-way analysis of variance (ANOVA) was separately performed on the MG and AMD values of the build orientations and layer thicknesses. Moreover, one-way ANOVA was performed for each point within each group. RESULTS: The margins of the area adjacent to the pontic showed significantly high values, and the values were smaller when the build orientation was 45° than when it was 60°. However, the margin did not differ significantly according to the layer thicknesses. CONCLUSION: The marginal fit of the three-unit resin prosthesis fabricated by the SLA 3D method was affected by the pontic. Moreover, the marginal fit was affected by the build orientation. The 45° build orientation is recommended.
Digital impressions and computer-aided design (CAD)-computer-aided manufacturing (CAM) techniques have replaced traditional manufacturing methods for dental prostheses and enabled a new era of prosthesis fabrication.1 The traditional fabrication method, the lost wax technique, produces a prosthesis by shaping it in wax and casting it after investing. However, that method is time-consuming and costly, and the technician’s skill affects the quality of the prosthesis.2 Compared with the traditional method, the CAD-CAM technique has a low learning curve and can quickly and easily fabricate prostheses of a consistent high quality, so it has been widely applied in various areas of dentistry.3The CAM technique can use subtractive or additive manufacturing.4 Subtractive manufacturing (SM) uses a milling tool to sculpt a block into the desired shape. It has disadvantages such as the waste of raw materials, wear of milling burs, and the occurrence of micro-cracks in the restorations during milling.5 The additive manufacturing (AM) method, on the other hand, builds up materials into the desired shape layer by layer. As its techniques have developed, it has overcome many shortcomings of the SM method.67 Several AM methods are used, but stereolithography (SLA) and digital light processing (DLP) are the most common. SLA polymerizes a photosensitive resin using a single focused laser.8 Although the initial cost of an SLA 3D printer is high, and the printing time is longer than with a DLP 3D printer, it is widely used in dentistry because the resulting printed prostheses have high dimensional accuracy.9Various printing parameters affect the accuracy of prostheses manufactured using CAD-CAM: build orientation, layer thickness, x-y resolution, light exposure time, post-processing, printing materials, and other factors.6 Among them, many studies have examined build orientation and layer thickness. Byun and Lee10 used a genetic algorithm to find the optimal orientation and improve the prototype surface roughness and build time. Zwier and Wits11 found that the print orientation was a decisive factor in the quality of the printed product. To find the optimal print orientation, they minimized the overhanging and support structures using ray-tracing and convex hull methods. Singhal et al.
12 used an adaptive slicing algorithm with different slice thicknesses to improve their SLS prototype’s surface roughness and geometric accuracy.Many other studies have also been conducted to optimize the build orientation and layer thickness, including studies on the effects of those two parameters on prosthesis fit. Park et al.
13 assessed three-unit provisional resin prostheses printed with a DLP 3D printer in ten groups (n = 10 per group) with five build orientations (0°, 30°, 45°, 60°, 90°) and two layer thicknesses (50 µm, 100 µm). Absolute marginal discrepancy (AMD), marginal gap (MG), and internal gap volume (IGV) showed the best results at build orientations of 45° and 60°, but the layer thicknesses produced different AMD, MG, and IGV results. Jang and Kim14 printed three-unit resin prostheses on an SLA 3D printer and set ten groups (n = 10 for each group) using the same build orientations and layer thicknesses as in the above study to measure marginal fit and internal fit. They found that the build orientations of 45° and 60° and the 50-µm layer thickness showed the most desirable fit. Both those studies found that the build orientation and layer thickness affected the quality of the printed prostheses. However, it is necessary to check whether printed prostheses show clinically appropriate marginal fit under the conditions deemed optimal in those studies. Moreover, in those studies, margin values were measured at the 4 margin areas, but it is essential to measure margin values in more diverse areas to evaluate whether printed prostheses show clinically appropriate margin values.Provisional restoration is an essential part of the transition period until a final restoration is delivered. It plays a role in pulp and abutment protection, positional stability, soft tissue management, and the maintenance of function and aesthetics.15 Polymethyl-methacrylate (PMMA) is often used to fabricate provisional restorations. Fabricating directly in the oral cavity has various drawbacks,16 particularly polymerization shrinkage and exothermic reaction. Polymerization shrinkage can cause dimensional discrepancy of the provisional prosthesis, and the exothermic reaction can inflict thermal trauma on the tooth pulp.17 The indirect method for building a provisional prosthesis makes a cast from an impression of the patient’s teeth, but it has the disadvantages of being affected by the technician’s skill and not being reproducible.18The most critical factor and prerequisite for ensuring the long-term success of a fixed prosthesis is marginal fit.19 Adequate marginal fit has been studied by many scholars2021 because an improper marginal fit causes cement dissolution that leads to percolation of bacteria, secondary dental caries, and pulp necrosis.22 Moreover, it affects the supporting periodontal tissue by causing plaque accumulation and bacterial proliferation.23It is difficult to accurately determine a clinically acceptable margin for fixed prostheses. According to ADA standard No.8,24 when using type I luting cement, the margin should not exceed 25 µm, and when using type II luting cement, it should not exceed 40 µm. However, satisfying those criteria in clinical situations is difficult. After examining the marginal fit of 1,000 fixed prostheses for five years, McLean and von Fraunhofer25 reported that if the cement film thickness and marginal gap were less than 120 µm, it was considered a successful prosthesis. Many clinicians agreed with that result, and many studies have used it as a clinical criterion for a successful marginal gap.182026In many cases, the same terms for marginal gap measurement have been interpreted differently. However, in 1989, Holmes et al.
27 re-established various casting misfit terms used to measure the marginal fit of prostheses, and among them, MG and AMD were widely used. Therefore, in this study, the marginal fit was measured and evaluated using MG and AMD.The purpose of this study was to analyze the marginal fit of three-unit resin prostheses printed with the SLA method in two build orientations (45°, 60°) and two layer thicknesses (50 µm, 100 µm). The null hypothesis was that the marginal gap and absolute marginal discrepancy of the three-unit resin prostheses manufactured by the SLA method would not be different under the tested printing conditions.
MATERIALS AND METHODS
A master model for a three-unit resin prosthesis with two implant abutments was designed (Fig. 1). The abutments were assumed to be mandibular second premolar and second molar, and the first molar was assumed to be the missing tooth. Each abutment was designed considering the anatomical shape and size of the teeth.28 Goodacre et al.
29 argued that it was appropriate to set a total occlusal convergence angle between 10°and 20° when an occluso-cervical dimension of 4 mm was set during tooth preparation. Thus, the total convergence angle was set to 14.3°. The abutments were set with a 1 mm shoulder margin, and the reference points of the cone shape were designed with a diameter of 1 mm and a height of 1 mm below the margin. Reference points were set in three directions at 90° each and were considered to confirm the exact positional relationship during coronal and sagittal sectioning.
Fig. 1
The master model. (A) Occlusal view and (B) Buccal view.
The standard tessellation language (STL) file of the master model was acquired using CAD software (Rhinoceros 5.0; Robert McNeel & Associates, Seattle, WA, USA). Using the acquired STL file, a PMMA resin block (Yamahachi Dental MFG, Ochigara, Japan) was milled using a 5-axis milling machine (IDC MILL 5X; Amann Girrbach AG, Koblach, Austria). According to ISO/ASTM 52916, STL is an abbreviation of ‘Stereolithography’, but for connectivity with the previous studies,1314 ‘Standard tessellation language’ was defined as STL and ‘Stereolithography’ as SLA in this study.The master model was scanned with a model scanner (T500; Medit, Seoul, Korea). According to the manufacturer, the scanning accuracy was within 7 µm. Before scanning, anti-reflective spray (IP scan spray; IP-division, Haimhausen, Germany) was used. The scanned model was exported as an STL file. Then, the three-unit resin prosthesis was designed using the STL file and CAD software (Exocad; Darmstadt, Germany). Based on a previous study30 to find an appropriate cement space for a three-unit resin prosthesis made with an SLA 3D printer, the cement space was set to 100 µm. The shape and size of the prosthesis were designed to fit the anatomical tooth morphology.28Forty specimens were printed using an SLA 3D printer (Zenith U; Dentis, Daegu, Korea). The specifications of the SLA 3D printer are shown in Table 1. Temporary resin (ZMD-1000B temporary; Dentis, Daegu, Korea) for 3D printing was used to fabricate the specimens. The specimens were printed with two build orientations (45°, 60°) (Fig. 2), and each build orientation was used with two layer thicknesses (50 µm, 100 µm). When the build orientation was 45°, the specimens with a layer thickness of 50 µm were called Group 1, and those with a layer thickness of 100 µm were called Group 2. When the build orientation was 60°, the specimens with a layer thickness of 50 µm were called Group 3, and those with a layer thickness of 100 µm were called Group 4 (n = 10 per group). The details of the groups are summarized in Table 2.
Table 1
The specifications of the SLA 3D printer used in these experiments
Scanning method
Galvanometer
Light source
Blue laser (405 nm)
Layer thickness
16 μm, 50 μm, 100 μm
Dimension / Weight
354 × 366 × 483 mm / 17.5 kg
Working area
110 × 110 × 150 (X, Y, Z / mm)
Electrical consumption
120W
Country
Daegu, Korea
Company
Dentis
Fig. 2
Prosthesis design with two build orientations (45° and 60°). The connection elements of the supporting structures varied by build orientation. For example, when the build orientation was 0°, the supporting structures were vertically attached to the occlusal surface of the prosthesis, and when the build orientation was 90°, the supporting structures were vertically attached to the lingual surface of the prosthesis. (A) Distal view and (B) Buccal view.
Table 2
The main features of the groups designed for this experiment
Manufacturing method
Manufacturer
Material
Build orientation
Layer thickness
Group 1
SLA method
Dentis
ZMD-1000B temporary
45o
50 μm
Group 2
SLA method
Dentis
ZMD-1000B temporary
45o
100 μm
Group 3
SLA method
Dentis
ZMD-1000B temporary
60o
50 μm
Group 4
SLA method
Dentis
ZMD-1000B temporary
60o
100 μm
According to the studies of Unkovskiy et al.
31 and Osman et al.,32 the exposure time required for the laser to reach the specimens can be affected by their platform position, causing volume discrepancy. Therefore, four specimens were printed simultaneously to keep them the same distance from the center of the platform and facilitate printing. Supporting structures were applied automatically using Zenith printer software (Zenith S/W; Dentis, Daegu, Korea). After printing, the remaining resin was removed, and the prostheses were cleaned with 99.8% ethanol (Absolute ethanol; Koryo Chemical Eng., Seoul, Korea) for 5 minutes. Next, post-curing was performed in an ultraviolet curing unit (LC-3D Print Box; Nextdent, Utrecht, Netherlands) for 5 minutes according to the manufacturer’s instruction. Then, the supporting structures were carefully removed using denture burs.After it was confirmed that each specimen was seated on the master model without interference, the model was fixed on the jig for micro-CT scanning using a laboratory wrapping film (Parafilm; Bemis, Neenah, WI, USA) without cementation. Then, CT scanning was performed using a micro-CT scanner (Skyscan 1172; Bruker Micro CT, Billerica, MA, USA). Scanning was done using parameters based on previous studies1314: 60 kVp and 167 µm, with an exposure time of 1475 ms. An aluminum filter (5 mm) was used, and the resolution of the CT scan was 15.43 µm. Each specimen was rotated 180° with 0.7° rotational steps and 3-frame averaging.The CT data were reconstructed using NRecon software (NRecon 1.7.4.2 version; Bruker Micro CT, Billerica, MA, USA). The threshold for the defect pixel mask was set to 3%. Ring artifact reduction was set to 8, and smoothening was set to 3. ImageJ software (ImageJ 1.52 version; NIH, Bethesda, MD, USA) was used to measure the marginal fit as the marginal gap (MG) and absolute marginal discrepancy (AMD). According to Holmes et al.,27 MG is “the perpendicular measurement from the margin of the casting to the axial wall of the preparation,” and AMD is “the angular combination of the vertical marginal discrepancy and the horizontal marginal discrepancy” (Fig. 3). Three reference points per abutment in the master model (A to F) were used for sectioning (Fig. 4). To find the desired plane, the coronal and sagittal planes were sectioned with DataViewer software (DataViewer 1.5.6.2 version; Bruker Micro CT, Billerica, MA, USA) based on the reconstructed data.
Fig. 3
(A) Marginal gap (MG) and (B) Absolute marginal discrepancy (AMD). The marginal fit at each margin point of the prostheses was measured using both MG and AMD.
Fig. 4
Reference points from A to F. For accurate margin measurement, the reference points were used for sectioning the reconstructed CT data in the desired planes.
MG and AMD measurements were taken at 8 points per abutment (16 points per specimen). In the premolar, the margin on the buccal side, which was the closest to reference point C, was set as P1, and as the specimen was rotated clockwise by 45°, the margin was set as P2, P3,…, P8. In the molar, the margin on the buccal side, which was the closest to reference point D, was set as M1, and when the specimen was rotated counter-clockwise by 45°, the margin was set as M2, M3,…, M8 (Fig. 5).
Fig. 5
Margin points for each abutment. Sixteen margins per specimen were measured. Based on the center of the abutment, the margin was set to be 45° different from the adjacent margin. Margin points in the premolar were set in clockwise order, and margin points in the molar were set in counter-clockwise order.
To make the MG and AMD measurements at 16 points, seven sections per specimen had to be made. The plane passing through reference points A and B perpendicular to the master model was defined as the coronal section. The plane passing through reference points C and E perpendicular to the master model was defined as the premolar sagittal section. The plane passing through reference points D and F perpendicular to the master model was defined as the molar sagittal section. The remaining four sections were obtained by rotating the specimen 45° counter-clockwise. In the premolar, the plane connecting the longest axis to the coronal was defined as the premolar 45° coronal section. The plane connecting the longest axis to the sagittal was defined as the premolar 45° sagittal section. Likewise, in the molar, the plane connecting the longest axis to the coronal was defined as the molar 45° coronal section, and the plane connecting the longest axis to the sagittal was defined as the molar 45° sagittal section. The margin points measured in each section are summarized in Table 3.
Table 3
Margin points measured in each section
Section
Premolar
Molar
Coronal section
P3, P7
M3, M7
Premolar sagittal section
P1, P5
Molar sagittal section
M1, M5
Premolar 45° coronal section
P4, P8
Premolar 45° sagittal section
P2, P6
Molar 45° coronal section
M2, M6
Molar 45° sagittal section
M4, M8
All measurements were taken under magnification of x100. Each measurement was repeated three times, and the average value was used.For the printed prostheses, the MG and AMD values were each measured for 16 margin points. Each group was tested for normality using the Shapiro-Wilk test and for homogeneity of variance using Levene’s test. All statistical analyses were performed using SPSS software (SPSS version 25; IBM, Armonk, NY, USA).Two-way analysis of variance (ANOVA) was separately performed on the MG and AMD values for the build orientations and layer thicknesses to check the interaction effect and significance of each factor. Moreover, one-way ANOVA was performed for the for MG and AMD values at each point within each group. Post hoc testing was performed using the Tukey test. The confidence level was set at 95% (α = 0.05).
RESULTS
The MG values are shown in Figure 6. In the premolar, the smallest value was 51.8 ± 10.6 µm at P6 (Group 1), and the largest value was 126.3 ± 16.1 µm at P4 (Group 3). The points at which the value exceeded 120 µm were P4 in Group 3 and P4 in Group 4. In the molar, the smallest value was 48.9 ± 6.3 µm at M7 (Group 1), and the largest value was 127.4 ± 12.6 µm at M4 (Group 3). The only value that exceeded 120 µm was M4 in Group 3. The AMD values are also shown in Figure 6. In the premolar, the smallest value was 54.6 ± 9.7 µm at P6 (Group 1), and the largest value was 127.3 ± 16.0 µm at P4 (Group 3). The points at which the value exceeded 120 µm were P4 in Group 3 and P4 in Group 4, which is the same pattern seen with the MG. In the molar, the smallest value was 60.1 ± 6.8 µm at M7 (Group 4), and the largest value was 128.5 ± 14.2 µm at M2 (Group 3). Points at which the value exceeded 120 µm were present in all groups: M3 and M4 in Group 1; M4 in Group 2; M2, M3, and M4 in Group 3; and M3 in Group 4. Furthermore, The MG and AMD values were analyzed for each point within each group. Regardless of the group and margin (MG or AMD), the results were somewhat consistent. Overall, P2, P3, and P4 in the premolar had significantly larger values than the other points, and M2, M3, and M4 in the molar had significantly larger values than the other points (Fig. 6). Although it did not always show a statistically significant difference, P4 tended to show larger values than P2 and P3, and M4 tended to show larger values than M2 and M3.
Fig. 6
Comparison of MG and AMD values by point within the groups. (A) Group 1 (build orientation: 45°, layer thickness: 50 µm), (B) Group 2 (build orientation: 45°, layer thickness: 100 µm), (C) Group 3 (build orientation: 60°, layer thickness: 50 µm), (D) Group 4 (build orientation: 60°, layer thickness: 100 µm). Although there were some exceptions, P2, P3, and P4 showed significantly larger values than the other points in the premolar, and M2, M3, and M4 showed significantly larger values than the other points in the molar. MG and AMD denote Marginal gap and Absolute marginal discrepancy, respectively.
The MG values differed significantly according to the build orientation. Smaller MG values were shown when the build orientation was 45° than when it was 60° (P = .002). However, the MG values did not differ significantly according to the layer thickness (P = .681), though there was an interaction effect between the build orientations and layer thicknesses (P < .001) (Fig. 7A). The AMD values showed the same pattern as the MG values. Smaller AMD values were found when the build orientation was 45° than when it was 60° (P = .042), the AMD values did not differ significantly according to the layer thicknesses (P = .479), and there was an interaction effect between the build orientations and layer thicknesses (P < .001) (Fig. 7B).
Fig. 7
Comparison of the effects of the build orientations and layer thicknesses. (A) MG values and (B) AMD values. Both the MG and AMD values showed significant differences according to the build orientation, with the 45° build orientation showing significantly smaller values than 60°. However, there were no differences according to the layer thickness. P values are listed on the graph for P < .05 (*). The error bars represent the standard deviations (SD). MG and AMD denote Marginal gap and Absolute marginal discrepancy, respectively.
DISCUSSION
This study analyzed the marginal fit of three-unit resin prostheses fabricated with an SLA 3D printer using two build orientations and two layer thicknesses. The marginal fit was measured separately at 8 points in the premolar and molar, for a total of 16 points. Based on McLean’s study,25 120 µm was set as a clinically acceptable margin for prostheses, and based on Holmes et al.’s study,27 marginal fit was evaluated using the MG and AMD. The null hypothesis was rejected because points in the margins of the three-unit resin prostheses showed differences in both MG values and AMD values.No previous study analyzed the various margin points of 3D-printed three-unit resin prostheses from both the MG and AMD aspects. In this study, the MG values ranged from 51.8 µm to 126.3 µm in the premolar and 48.9 µm to 127.4 µm in the molar, and the AMD values ranged from 54.6 µm to 127.3 µm in the premolar and 60.1 µm to 128.5 µm in the molar. Points exceeding 120 µm were observed in both the premolar and molar, and the frequency was higher in the molar than the premolar. In the premolar, it exceeded 120 µm only at P4, but in the molar, it exceeded 120 µm at M2, M3, and M4.Although there were some exceptions, P2, P3, P4 in the premolar and M2, M3, M4 in the molar showed significantly larger margin values than the other points. It should be noted that these points were adjacent to the pontic. Furthermore, P4 and M4 tended to show larger values than P2 and P3 and M2 and M3, respectively, although those differences were not statistically significant.The significantly large values found in the margins adjacent to the pontic seem to be due to polymerization shrinkage of the resins used in this experiment. The amount of resin used for the pontic was larger than the amount used for each abutment, and as the amount of resin increases, the polymerization shrinkage also increases.33 Moreover, referring to previous experimental results showing that the polymerization shrinkage of the resin occurred in the inward direction,34 it could be expected that polymerization shrinkage in the pontic might have affected the marginal fit of the prosthesis in the area adjacent to it.Polymerization shrinkage occurs in resins used to fabricate provisional crowns, and casting shrinkage occurs in the metal alloys often used as the final restoration materials. The casting shrinkage of metal alloys can compensate for shrinkage caused by investment expansion, but the shrinkage of resins cannot. Although the degree of casting shrinkage varies with the metal’s composition, it is about 2.1% for regular gold inlays, 2.0% for crowns, and 1.9% for MOD inlays.35 In non-precious metal alloys, casting shrinkage of about 3.4% occurs because casting is performed at high temperatures.36 Although polymerization shrinkage in resin depends on the main component, filler content, and polymerization method, PMMA, which was is the most common material used in temporary restorations, has polymerization shrinkage of 4.5 - 6.4%.37 Urethane dimethacrylate (UDMA) was the main component of the ZMD-1000B temporary resin used in this experiment, and it offers better marginal discrepancy than PMMA because it generates less heat and polymerization shrinkage.38 Nonetheless, the UDMA resin still had more polymerization shrinkage than the casting shrinkage of the alloys,39 which might have caused relatively more marginal discrepancy. Thus, the effect of the polymerization shrinkage of the resin on marginal discrepancy was significant. Moreover, as the amount of resin increases, the degree of polymerization shrinkage also increases,33 which likely explains why the margin near the pontic in this experiment often exceeded 120 µm.Meanwhile, the influence of the supporting structures is apparently why P4 and M4 tended to have larger MG and AMD values than P2 and P3 and M2 and M3, respectively. The resin on the 3D-printer platform was light-cured beginning with the part closest to the platform. Because the supporting structures were directly connected to the specimen, the polymerization shrinkage in those supporting structures might have affected the specimen. However, the resin volumes in the supporting structures adjacent to the margin were a little smaller than those in the pontic, which could explain why those differences were not statistically significant. Depending on the build orientation set in this experiment (45°, 60°), many of the supporting structures were located on the lingual, occlusal surfaces. The marginal points close to the lingual surface were P4, P5, P6 in the premolar and M4, M5, M6 in the molar. Considering that those points did not differ significantly from the other points, the polymerization shrinkage of the supporting structures probably had a minor effect compared with that of the pontic. Therefore, the polymerization shrinkage of the pontic had a major effect on the margin adjacent to it, and the shrinkage of supporting structures had a minor effect on the margin adjacent to them. In the end, the polymerization shrinkage of the pontic and supporting structures caused a synergistic effect at P4 and M4.The effect of the supporting structures was also found in other studies. Yu et al.
40 reported that the margin quality of the area near the support attachment was poor, and rough edges tended to occur in resin prostheses printed with an SLA 3D printer. In Osman’s study,32 which investigated the dimensional accuracy of the build orientation of dental crowns fabricated with a DLP 3D printer, the root mean square error (RMSE) was higher in the area with the support structure. In addition, Alharbi et al.
41 investigated the effect of the build angle and support thickness on SLA 3D-printed dental restorations. The smallest RMSE was obtained with a build angle of 120° (60° in this study) and thin support, and the deviation was significant when the support was thick. Further studies are needed to examine how various support structures affect the margins of 3D-printed prostheses.Fig. 7 shows how the build orientation and layer thickness affected the MG and AMD values. Both values showed the same results, and the layer thickness caused no significant differences. However, the build orientation did cause significant differences, with significantly smaller MG and AMD values at 45° than at 60°. In other words, the SLA 3D printer used in this experiment produced a better marginal fit when the build orientation was 45°. Similar studies of build orientation have been conducted on both DLP and SLA 3D printers. Osman et al.
32 evaluated the effect of 9 build orientations on full coverage crowns printed with a DLP 3D printer, and the dimensional accuracy was best when the build orientation was 135° (45° in this study). Unkovskiy et al.
31 evaluated bar-shaped specimens printed with an SLA 3D printer using three build orientations and reported that the printing accuracy was best at 45°. The finding of a better marginal fit at a build orientation of 45° reported here was expected for two reasons.First, the supporting structures were positioned closer to the lingual margin when the build orientation of the prosthesis was at 60° than when it was at 45°. Therefore, polymerization shrinkage in the supporting structures was more likely to have a greater effect when the build orientation was 60°, and that was assumed to adversely affect the marginal fit.40Second, the build orientation affected the number of slices (layers) that were polymerized,42 with more layers polymerized at 60° than at 45°. Errors between layers accumulate as the number of layers increases, which could affect the accuracy of the final printed prosthesis. A result similar to this experiment was confirmed in another study14 that examined the fit of three-unit resin prostheses made on an SLA 3D printer and found a significantly poorer marginal fit at 60° than at 45°.Unlike the build orientation, the layer thickness did not cause significant differences in marginal fit. Although many studies have examined the layer thickness of 3D printers, they have often shown conflicting results. Favero et al.
43 measured the printing accuracy of an orthodontic model using three layer thicknesses (25 µm, 50 µm, and 100 µm) on an SLA 3D printer. The largest average deviation was found at 25 µm, and the smallest was found at 100 µm. However, Zhang et al.
44 reported that the average absolute deviation of orthodontic models printed with SLA and DLP 3D printers was the lowest when the layer thickness was 25 µm with SLA and 50 µm with DLP. These conflicting results indicate the need for additional, precise experiments on the effect of layer thickness.The interaction effect between the build orientation and layer thickness was significant in both the MG and AMD values. Group 1 (build orientation: 45°, layer thickness 50 µm) and Group 4 (build orientation: 60°, layer thickness : 100 µm) had the best marginal fit, which was similar to Park’s study.13 This might be seen to indicate that a layer thickness offering a better marginal fit for each build orientation exists. However, because these fragmentary results alone are insufficient to derive that conclusion, studies on various combinations of build orientation and layer thickness with more diverse types of prostheses and 3D printers are needed.Marginal fit measurements have been performed using various methods. According to a review paper45 on the margin measurement method, the most-used method is the direct view technique, followed by the cross-sectioning method, and then the replica technique. The direct view technique measures the margin using a microscope or scanning electron microscope and has the advantage of being relatively simple and not requiring additional steps.46 However, it is difficult to measure a rounded margin and find an exact point.47 The cross-sectioning method enables accurate margin measurement by sectioning the sample at the desired angle, but the sample must be sacrificed, and the number of sectioning planes is limited.48 The replica technique obtains a fragile cement space by seating the crown using light body silicone and fixing the light body silicone using heavy body silicone. Then, the margin of the desired area is measured by sectioning. The drawback of this method is that the replica can be torn or deformed when the silicone materials are removed from the crown.49 In this experiment, micro-CT was used because it can measure the margin through a three-dimensional high-resolution image without causing any damage to the specimen.26 Moreover, unlike the cross-sectioning method, it is possible to use as many sections as desired from various angles to measure many margin points.16Instead of cementing the prosthesis to the abutment, it was fixed to the micro-CT jig using wrapping film. Gonzalo et al.
19 reported a slight increase in the margin after cementation, but it was not a statistically significant difference. However, the use of cement can result in improper prosthesis seating and worsen the marginal fit.50 In addition, it was reported that the radiopacity of the luting agent could interfere with the margin measurement.1
CONCLUSION
Within the limitations of this study, the following conclusions are drawn. The marginal fit was affected by the build orientation, and a build orientation of 45° is recommended. Specifically, the marginal fit of three-unit resin prostheses fabricated by the SLA 3D-printing method was poor in the marginal areas adjacent to the pontic. The comparison of different layer thicknesses did not show a significant result, so an additional research is needed.