Literature DB >> 36090780

Comparison of radiographic vs. computed tomography canal filling ratio measurement of stems in revision knee prostheses.

Osman Cimen1, Mehmet Akif Kaygusuz1.   

Abstract

Background: Lack of consensus on which measurement method gives the most accurate results for medullary canal filling ratio (CFR) of stems used in revision knee prostheses causes confusion when interpreting scientific studies. Therefore, the present study aims to find the most accurate measurement method of CFR on direct radiographs.
Methods: Twenty-eight femoral and 27 tibial components of 29 patients with a mean age of 72.86 ± 6.32 years were included in the study. Two different methods were used for computerized tomography CFR measurement. These methods and five different radiographic measurement methods used for CFR were compared statistically.
Results: There was no statistically significant difference between the real tomographic CFR of the femoral stem and the CFR calculation made by using ellipse area (p > 0.05). There was also no statistically significant difference between the tomographic CFR measurement calculated according to the largest stem size that can be implanted, and the method that accepted the smallest value as CFR after the measurement made according to Parsley's method on AP and lateral radiographs (p > 0.05). The results obtained by using all other tests were statistically different from each other.
Conclusion: CFR calculation made by using ellipse area, is the most accurate method for femoral stem CFR. The method that accepted the smallest value as CFR after the measurement made according to Parsley's method on AP and lateral radiographs is the most accurate method for tibial stem CFR.
© 2022 Professor P K Surendran Memorial Education Foundation. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Canal filling ratio; Femur; Modular stem; Revision total knee arthroplasty; Stem fixation; Tibia

Year:  2022        PMID: 36090780      PMCID: PMC9450069          DOI: 10.1016/j.jor.2022.08.022

Source DB:  PubMed          Journal:  J Orthop        ISSN: 0972-978X


  14 in total

1.  The effect of stem modularity and mode of fixation on tibial component stability in revision total knee arthroplasty.

Authors:  L M Jazrawi; B Bai; F J Kummer; R Hiebert; S A Stuchin
Journal:  J Arthroplasty       Date:  2001-09       Impact factor: 4.757

2.  The optimal strategy for stable tibial fixation in revision total knee arthroplasty.

Authors:  Michael A Conditt; Brian S Parsley; Jerry W Alexander; Sean D Doherty; Philip C Noble
Journal:  J Arthroplasty       Date:  2004-10       Impact factor: 4.757

3.  Outcomes and Aseptic Survivorship of Revision Total Knee Arthroplasty.

Authors:  Jeffrey J Cherian; Anil Bhave; Steven F Harwin; Michael A Mont
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2016-02

4.  Component alignment in revision total knee arthroplasty using diaphyseal engaging modular offset press-fit stems.

Authors:  Cass K Nakasone; Ayesha Abdeen; Armond G Khachatourians; Tanzo Sugimori; Kelly G Vince
Journal:  J Arthroplasty       Date:  2008-02-20       Impact factor: 4.757

5.  Optimizing Mechanical Alignment With Modular Stems in Revision TKA.

Authors:  Andrew N Fleischman; Ibrahim Azboy; Camilo Restrepo; Mitchell G Maltenfort; Javad Parvizi
Journal:  J Arthroplasty       Date:  2017-05-26       Impact factor: 4.757

6.  Effect of Stem Size and Fixation Method on Mechanical Failure After Revision Total Knee Arthroplasty.

Authors:  Andrew N Fleischman; Ibrahim Azboy; Michael Fuery; Camilo Restrepo; Hongyi Shao; Javad Parvizi
Journal:  J Arthroplasty       Date:  2017-05-04       Impact factor: 4.757

7.  Clinical & radiographic outcomes of cemented vs. diaphyseal engaging cementless stems in aseptic revision TKA.

Authors:  Jeremy M Gililland; Christian J Gaffney; Susan M Odum; Thomas K Fehring; Christopher L Peters; Walter B Beaver
Journal:  J Arthroplasty       Date:  2014-05-24       Impact factor: 4.757

8.  Stem fixation in revision total knee arthroplasty: a comparative analysis.

Authors:  Thomas K Fehring; Susan Odum; Caryn Olekson; William L Griffin; J Bohannon Mason; Thomas H McCoy
Journal:  Clin Orthop Relat Res       Date:  2003-11       Impact factor: 4.176

Review 9.  Stem Fixation in Revision Total Knee Arthroplasty: Indications, Stem Dimensions, and Fixation Methods.

Authors:  Se Gu Kang; Cheol Hee Park; Sang Jun Song
Journal:  Knee Surg Relat Res       Date:  2018-09-01

10.  Influence of extension stem length and diameter on clinical and radiographic outcomes of revision total knee arthroplasty.

Authors:  Sheng-Hsun Lee; Hsin-Nung Shih; Chih-Hsiang Chang; Tung-Wu Lu; Yu-Han Chang; Yu-Chih Lin
Journal:  BMC Musculoskelet Disord       Date:  2020-01-08       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.