Thomas Balanis1, Bernd Sanner1.
Abstract
PURPOSE: Atrial fibrillation (AF) is the most common arrhythmia and is associated with an increased risk of complications. A screening test has the potential to prevent AF-related complications. This study investigated the diagnostic accuracy of an automated device for home blood pressure (BP) monitoring, which implements an algorithm for AF detection. PATIENTS AND METHODS: A modified, automated oscillometric device for home BP monitoring (Omron BP785N (HEM-7321-Z), Omron Healthcare) with an AF detector was used to measure the BP in patients. During each BP measurements, the electrocardiogram (ECG) was recorded simultaneously. Simultaneous BP measurements and ECG recordings were obtained from 99 subjects.Entities:
Keywords: Omron BP785N; atrial fibrillation detection; blood pressure measurement
Mesh:
Year: 2021 PMID: 34285495 PMCID: PMC8285290 DOI: 10.2147/VHRM.S317859
Source DB: PubMed Journal: Vasc Health Risk Manag ISSN: 1176-6344
Characteristics of Participants with Atrial Fibrillation and Sinus Rhythm
| Characteristics | All Subjects (n=99) | Atrial Fibrillation (n=20) | Sinus Rhythm (n=79) | p value | SD |
|---|---|---|---|---|---|
| Age (years) | 70.2 | 80.2 | 67.3 | <0.001 | 14.88 |
| Weight (kg) | 74.3 | 73.5 | 74.6 | 0.781 | 11.38 |
| Arm circ (cm) | 28.2 | 28 | 28.7 | 0.258 | 2.33 |
| CVD (%) | 19 | 42 | 18.75 | 0.3581 | – |
| Arterial hypertension (%) | 60 | 100 | 54 | 0.0003 | – |
| Diabetes mellitus (%) | 11 | 0 | 11.25 | 0.3389 | – |
| Number of drugs | 3.8 | 6.2 | 3.2 | <0.001 | 2.9 |
| Systolic BP (mm Hg) | 138 | 141 | 132 | 0.096 | 20.33 |
| Diastolic BP (mm Hg) | 80 | 79 | 81 | 0.715 | 9.97 |
Abbreviations: BP, blood pressure; Circ, circumference; CVD, cardiovascular disease; SD, standard deviation.
Diagnostic Accuracy of a Home Blood Pressure Monitor to Detect Atrial Fibrillation Without Consideration of the Body Movements Detected by the Device
| Statistic | Value | 95% CI |
|---|---|---|
| Sensitivity | 100% | 83.16 to 100.00% |
| Specificity | 84.8% | 74.97% to 91.90% |
| Positive predictive value | 62.5% | 49.74% to 73.73% |
| Negative predictive value | 100% | |
| Accuracy | 87.88% | 79.78% to 93.58% |
Diagnostic Accuracy of a Home Blood Pressure Monitor to Detect Atrial Fibrillation with Excluding the Patients with Body Movements During the BP Measurement
| Statistic | Value | 95% CI |
|---|---|---|
| Sensitivity | 100% | 66.37 to 100.00% |
| Specificity | 89.5% | 79.65% to 95.70% |
| Positive predictive value | 56.25% | 38.94% to 72.16% |
| Negative predictive value | 100% | |
| Accuracy | 90.79% | 81.94% to 96.22% |