Sung-Sahn Lee1, Jaesung Park2, Dae-Hee Lee2.
Abstract
Background andEntities:
Keywords: TomoFix; anatomical conformity; high tibial osteotomy; plate position
Mesh:
Year: 2022 PMID: 36013511 PMCID: PMC9413536 DOI: 10.3390/medicina58081045
Source DB: PubMed Journal: Medicina (Kaunas) ISSN: 1010-660X Impact factor: 2.948
Figure 1(A) Schematic diagram of the new plate lateral view. While the previous plate was a 90° T-shaped plate with no left and right distinction, the TomoFix anatomical plate has a 5° posterior slope, allowing the plate to be positioned relatively backward. As a result, the new plate system (B) enables parallel screw fixation than the previous plate (C).
Demographic and preoperative data.
| Number of Cases, | 116 |
|---|---|
| Sex, M:F | 35:81 |
| Age, year | 56.9 ± 7.0 (43–77) |
| BMI, kg/m2 | 27.5 ± 3.4 (20.4–38.0) |
| Direction, R:L | 50:66 |
| Preoperative HKA angle, ° | 8.6 ± 3.0 (5–20.1) |
| Preoperative MPTA, ° | 84.6 ± 2.0 (79.5–89.3) |
| Preoperative tibial slope, ° | 10.1 ± 3.7 (1.7–20.7) |
BMI, body mass index; HKA, hip-knee-ankle; MPTA, medial proximal tibial angle.
Figure 2Measurement of the plate angle and screw angles on the CT axial scan. (A) The plate angle was defined as the angle between the anteroposterior line connecting the proximal part of the plate and the posterior tibial condylar line. (B) The A, B, and C screw angle was defined as the angle between the line drawn along the screw and the posterior tibial condylar line. (A, B, and C screws from anterior to posterior, respectively).
Figure 3Measurement of the plate-to-cortex distance at #1 hole. It was defined as the distance between the inner surface of the plate and the outer surface of the tibial cortex at #1 hole.
Comparison of the demographic and simple radiographic data between the TomoFix conventional plate and the TomoFix anatomical plate groups.
| TomoFix Conventional | TomoFix Anatomical | ||
|---|---|---|---|
| Number of cases, | 63 | 53 | |
| Sex, M:F | 17:46 | 18:35 | 0.415 |
| Age, year | 56.0 ± 7.4 | 57.8 ± 6.5 | 0.171 |
| BMI, kg/m2 | 27.4 ± 3.3 | 27.7 ± 3.5 | 0.681 |
| Direction, R:L | 28:35 | 22:31 | 0.851 |
| Preoperative HKA angle, ° | 8.3 ± 2.7 | 9.0 ± 3.3 | 0.156 |
| Preoperative MPTA, ° | 84.8 ± 2.2 | 84.3 ± 1.9 | 0.26 |
| Preoperative tibial slope, ° | 10.4 ± 3.7 | 9.7 ± 3.6 | 0.308 |
| Opening width, mm | 10.6 ± 2.6 | 11.2 ± 1.7 | 0.169 |
| Postoperative HKA angle, ° | -3.1 ± 2.5 | -2.5 ± 2.3 | 0.135 |
| Postoperative MPTA, ° | 92.8 ± 2.5 | 92.5 ± 2.5 | 0.469 |
| Postoperative tibial slope, ° | 11.8 ± 4.2 | 11.4 ± 3.8 | 0.549 |
BMI, body mass index; HKA, hip–knee–ankle; MPTA, medial proximal tibial angle.
Comparison of the fixative device position on CT scan between the two groups.
| TomoFix Conventional | TomoFix Anatomical | ||
|---|---|---|---|
| Plate angle, ° | 31.7 ± 7.0 | 39.2 ± 8.1 | 0.001 |
| A screw angle, ° | 61.4 ± 6.6 | 51.8 ± 7.4 | <0.001 |
| B screw angle, ° | 57.8 ± 7.3 | 49.6 ± 7.7 | <0.001 |
| C screw angle, ° | 54.2 ± 7.4 | 46.6 ± 7.4 | <0.001 |
| Using cortical screw on #1 hole | 16 cases (25.4%) | 8 cases (15.1%) | 0.128 |
| Plate-to-cortex distance | 3.2 ± 1.7 | 1.7 ± 1.3 | <0.001 |
| Plate-to-cortex distance | 3.9 ± 1.1 | 1.9 ± 1.2 | <0.001 |