| Literature DB >> 36012948 |
Dominik Adl Amini1,2, Manuel Moser1,3, Erika Chiapparelli1, Lisa Oezel1,4, Jiaqi Zhu1, Ichiro Okano1, Jennifer Shue1, Andrew A Sama1, Frank P Cammisa1, Federico P Girardi1, Alexander P Hughes1.
Abstract
BACKGROUND: Advanced glycation end products (AGEs) have been shown to accumulate in bone and are gaining interest in connective tissue research. AIMS: To investigate the intrarater reliability, two-timepoint agreement and correlations within and between two commercially available skin autofluorescence (SAF) AGE devices.Entities:
Keywords: AGE; advanced glycation end products; connective tissue; fingertip fluorescence; osteoporosis; skin autofluorescence
Year: 2022 PMID: 36012948 PMCID: PMC9410107 DOI: 10.3390/jcm11164709
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.964
Demographic characteristics (n = 40).
| Variables | |
|---|---|
| Gender, n (%) | |
| Male | 15 (37.5) |
| Female | 25 (62.5) |
| Race, n (%) | |
| White/Caucasian | 21 (52.5) |
| Others | 8 (20.0) |
| Black/African American | 6 (15.0) |
| Asian | 5 (12.5) |
| Mean Body Mass Index (BMI) (kg/m2), (SD) | 25.9 (5.2) |
| Fitzpatrick Skin Type, n (%) | |
| I | 4 (10.0) |
| II | 13 (32.5) |
| III | 9 (22.5) |
| IV | 7 (17.5) |
| V | 5 (12.5) |
| VI | 2 (5.0) |
Lifestyle and health status of study participants (n = 40). * This question was only used at timepoint two.
| Variables, n (%) | |
|---|---|
| Osteoporosis | 0 (0.0) |
| Diabetes | 1 (2.5) |
| Chronic Heart Disease | 0 (0.0) |
| Chronic Lung Disease | 1 (2.5) |
| Kidney Disease | 0 (0.0) |
| Liver Disease | 0 (0.0) |
| Mental Health Problems | 0 (0.0) |
| Smoking | |
| Never | 36 (90.0) |
| Former | 3 (7.5) |
| Current | 1 (2.5) |
| Alcohol consumption | |
| Never | 3 (7.5) |
| Social drinker | 29 (72.5) |
| More | 6 (15.0) |
| N/A | 2 (5.0) |
| Regular usage of skin moisturizer | 33 (82.5) |
| Regular usage of anti-aging products | 13 (32.5) |
| Physical activity within the last 48 h | 20 (50.0) |
| Regular physical activity | 25 (62.5) |
| Lifestyle changes since the first measurement * | 9 (22.5) |
(1). ICC for intrarater reliability.(2). ICC for test–retest reliability.
| ( | ||
|
| ||
|
|
|
|
| AGE reader | ||
| Timepoint 1 | 0.998 | (0.996, 0.999) |
| Timepoint 2 | 0.996 | (0.993, 0.998) |
| AGE sensor | ||
| Timepoint 1 | 0.942 | (0.904, 0.967) |
| Timepoint 2 | 0.923 | (0.866, 0.957) |
| ( | ||
|
| ||
|
|
|
|
| AGE reader | ||
| 1. Measurement | 0.958 | (0.919, 0.979) |
| 2. Measurement | 0.969 | (0.939, 0.984) |
| 3. Measurement | 0.965 | (0.923, 0.980) |
| Average score | 0.965 | (0.931, 0.982) |
| AGE sensor | ||
| 1. Measurement | 0.848 | (0.711, 0.919) |
| 2. Measurement | 0.814 | (0.647, 0.901) |
| 3. Measurement | 0.861 | (0.723, 0.928) |
| Average score | 0.912 | (0.834, 0.953) |
Figure 1Correlation plots within (top) and between (bottom) AGE reader and AGE sensor.