Impact of tricuspid valve insufficiency on the performance of left ventricular assist devices.

Claudio J R Gomez Hamacher1, Carolin Torregroza2, Najla Sadat1, Daniel Scheiber3, Jil-Cathrin von der Beek4, Ralf Westenfeld3, Ivonne Jeanette Knorr4, Martin Sager4, Artur Lichtenberg1, Diyar Saeed1,5.   

Abstract

Objective: To evaluate the impact of severe tricuspid valve insufficiency (TVI) at the time of left ventricular assist device (LVAD) implantation on the hemodynamic and LVAD parameters in an acute ovine model.
Methods: Stable heart failure (HF) was induced in 10 ovines through the application of 3 ± 1 coronary ligations. Once stable HF was obtained (after 15 ± 5 days), the animals were supported with an LVAD. Hemodynamic data and pump parameters were obtained and compared in 2 settings; first with LVAD in place after weaning from the cardiopulmonary bypass machine (no TVI condition) and second following the induction of severe TVI through resection of the tricuspid valve (TVI condition).
Results: There were no statistically significant differences in the hemodynamic and pump parameters between TVI condition and no TVI conditions except for lower cardiac output in the TVI condition (2 [1.38-2.8] L/min vs 3.2 [1.55-3.7] L/min, P = .027) and the expected greater central venous pressure in the TVI condition (26 [24-31] mm Hg vs 15 [13-25] mm Hg, P = .020). A median pump flow of 2.8 (2.45-3.75) L/min versus 2.9 (2.75-3.8) L/min in the TVI condition and no TVI condition was documented (P = .160). Conclusions: Results from this acute animal study suggest that severe TVI in HF with preserved right ventricular function does not have significant impact on the LVAD pump parameters. The observed reduction in cardiac output may warrant further investigations, especially under loading conditions.
© 2020 The Authors.

Entities:  

Keywords:  CO, cardiac output; CPB, cardiopulmonary bypass; CVP, central venous pressure; HF, heart failure; HR, heart rate; LVAD, left ventricular assist device; MAP, mean arterial pressure; PAPi, pulmonary artery pulsatility index; RV, right ventricular; RVF, right ventricle failure; TVI, tricuspid valve insufficiency; TVR, tricuspid valve reconstruction; animal model; bpm, beats per minute; coronary ligation; heart failure; mPAP, mean pulmonary artery pressure; ovine model; tricuspid regurgitation; tricuspid valve insufficiency; ventricular assist device

Year:  2020        PMID: 36004302      PMCID: PMC9390470          DOI: 10.1016/j.xjon.2020.09.001

Source DB:  PubMed          Journal:  JTCVS Open        ISSN: 2666-2736


LVAD parameters in both settings; with and without TVI. In our acute ovine model with preserved RV function, tricuspid valve insufficiency did not negatively impact ventricular assist device parameters, although cardiac output was significantly reduced. The impact of tricuspid valve insufficiency in patients with VAD is controversial. In this acute study, animals with heart failure supported with VAD and data were documented with and without tricuspid valve insufficiency. In the absence of RV dysfunction, pump parameters were similar in both settings. See Commentary on page 24. Heart failure (HF) is still one of the leading causes of death, affecting more than 26 million people worldwide. After the exhaustion of symptomatic therapies, heart transplantation and mechanical support systems are often ultimate resorts for the treatment of patients with end-stage HF. The increasing prevalence of patients with HF exceeds the number of organ donations, which has led to an increasing number of implantations of left ventricular assist devices (LVADs) in recent years as either bridge to transplantation or destination therapy, improving patients' survival and quality of life.2, 3, 4, 5 Up to 50% of patients with end-stage HF may present with simultaneous tricuspid valve insufficiency (TVI). Due to a progressive failing of the left ventricle and the consequent congestion, the afterload of the right ventricle increases, resulting in TVI. Because TVI is associated with right HF and worse overall outcome, the presence of TVI raises the question whether tricuspid valve reconstruction (TVR) should be performed at the time of LVAD implantation., The current International Society of Heart and Lung Transplantation guidelines state that moderate-to-greater TVI should prompt consideration of surgical repair at implantation. However, recent studies regarding treatment and outcome of TVI in patients with LVAD provide diverging results.9, 10, 11, 12, 13, 14, 15, 16 Therefore, surgical approach varies among surgeons, and a consistent treatment strategy according to guidelines is missing. The aim of this study was to examine the direct impact of acute TVI at the time of LVAD implantation on the hemodynamic and LVAD parameters using an acute ovine model with stabile left-sided HF and preserved right ventricular (RV) function.

Methods

All animals received humane care in compliance regarding “the EU Directive 2010/63,” “the German Animal Welfare Act,” and “the Animal Welfare Regulation Governing Experimental Animals,” all of which are on par with “the Principles of Laboratory Animal Care” formulated by the National Society for Medical Research and the “Guide for the Care and Use of Laboratory Animals,” prepared by the Institute of Laboratory Animal Resources and published by the National Institutes of Health (publication No. 86-23, revised 1996). The study protocol was approved by the corresponding regional authorities.

Anesthesia and Surgical Procedures

All animals received midazolam and atropine intramuscularly as premedication. After the induction of anesthesia using xylazine and ketamine (intramuscularly), endotracheal intubation was implemented, and animals were provided with a gastric tube. General anesthesia was continued using 1.3% isoflurane with oxygen air mixture, fentanyl, and pancuronium intravenously. Antiarrhythmic prophylaxis during induction of HF as well as LVAD implantation was administered intravenously before skin incision, before and after both coronary ligation, and LVAD insertion. The administered antiarrhythmic prophylaxis consisted of intravenous application of amiodarone, lidocaine, and magnesium. A continuous amount of glucose 5% or sodium chloride 0.9% solution at rates of 80 to 100 ml/h was given during the procedures. No blood products were given and no inotropic agents were administered to avoid arrhythmias. When necessary, intravenous norepinephrine was administered in moderate dose as an infusion to maintain a stable blood pressure. Notably, in all experiments, data were only obtained when stable conditions over several minutes were achieved to avoid variations due to fluid managements and/or vasoactive medications.

Surgical Induction of Stable HF

The study was carried out in 2 main stages. The first stage included induction of stable HF through coronary artery ligation. All animals underwent open-heart surgery using the left thoracotomy approach. Ovines were equipped with an arterial catheter in the left carotid artery, a central venous catheter in the left jugular vein, a pulmonary artery catheter, and an ultrasonic perivascular flow probe (AD Instruments, Dunedin, New Zealand) placed around the pulmonary artery for continuous cardiac output (CO) measurement. Hemodynamic and echocardiographic data were obtained at baseline, after coronary ligation, and before LVAD implantation. Notably, stable HF was established through transmural ligation of diagonal and marginal branches only. The hemodynamic data consisted of mean arterial pressure (MAP), central venous pressure (CVP), mean pulmonary artery pressure (mPAP), heart rate (HR), and CO. Once animals showed echocardiographic, hemodynamic, and clinical signs of HF, the animals were brought back to the operating room and LVAD implantation surgery was conducted as the second-stage procedure. Notably, the actual LVAD flow was not measured with flow probes. The pump flow measurements were calculated flows (as it clinically used in patients with HeartWare HVAD [HeartWare Inc, Framingham, Mass]). However, the hematocrit values were adjusted every 10 minutes to achieve accurate calculated flows. Similar to previous chronic animal studies of HF, HF in our studies was predefined as a 20% increase in pulmonary artery pressure and/or HR, and/or 20% reduction in CO as well as echocardiographic findings of dilation of the left ventricle and/or 20% reduction of left ventricular ejection fraction.

Surgical Procedure of LVAD Implantation and TVI Induction

The second surgical procedure consisted of LVAD implantation and induction of TVI in animals with stable HF. All animals received heparin (300 IU/kg) for anticoagulation throughout the second surgery. Right-sided thoracotomy was performed in the fifth intercostal space and the pericardial sac was opened. The carotid artery and internal jugular vein were prepared for cannulation of cardiopulmonary bypass (CPB). A second venous cannula for CPB was inserted in the inferior vena cava. CPB was initiated and the LVAD was implanted. The correct placement of the LVAD was verified by epicardial echocardiography. All animals were supported with HeartWare LVAD (HeartWare Inc). Figure 1 shows the LVAD in place. The aforementioned hemodynamic data and LVAD parameters (pump flow, speed, and power consumption) were obtained in 2 conditions, first, after weaning of CPB (no TVI condition) with LVAD in place and second, after reinitiating CPB, induction of TVI through surgical resection of the tricuspid valve (TV), and weaning off CPB (TVI condition). The pump speed was kept unchanged to allow reasonable comparison between both conditions. Figure 2 shows TV resection procedure. Once data were obtained, the animals were killed humanely in the operating room after completion of surgical procedure using pentobarbital–sodium. For the first stage of the study, hemodynamic and echocardiographic data were compared between all animals at baseline, post-ligation, and before LVAD implantation. In the second stage of the study, the hemodynamic and LVAD parameters at TVI and no TVI condition were compared within the same animal.
Figure 1

Intraoperative setting, left ventricular assist device in place. Dr, Driveline; Og, outflow graft; LV, left ventricle; PAC, pulmonary artery catheter.

Figure 2

Resection of the tricuspid valve. RA, Right atrium; RV, right ventricle.

Intraoperative setting, left ventricular assist device in place. Dr, Driveline; Og, outflow graft; LV, left ventricle; PAC, pulmonary artery catheter. Resection of the tricuspid valve. RA, Right atrium; RV, right ventricle.

Statistical Analysis

Statistical analysis was performed using SPSS for Mac (Version 25; IBM Corp, Armonk, NY). As not all variables were distributed normally (tested by Shapiro–Wilk test), data evaluation was achieved through a Wilcoxon signed-rank test (nonparametric paired t test). Data are reported as median value (interquartile range).

Primary Outcomes

Primary outcomes are hemodynamic changes between LVAD no TVI and LVAD with TVI setting, specifically CVP, MAP, CO, and mPAP, as well as LVAD parameters, including flow, pump speed, and power consumption.

Results

Ten female ovines (mean weight 61 ± 7 kg) underwent surgical induction of HF through ligation of diagonal and marginal branches. A total of 3 ± 1 transmural ligations were used. Stable HF was established after 15 ± 5 days. Animals showed clinical signs of HF, including reduced general conditions, wheezing, tachypnea, and dyspnea, on day 8 ± 3 post-ligation. Main findings in hemodynamic data were a significant increase in HR (72 ± 9 beats per minute [bpm] and 100 ± 28 bpm, P = .01) and mPAP (15 ± 4 mm Hg and 18 ± 5 mm Hg, P = .034) at baseline and before LVAD implantation. Further, diastolic pulmonary pressure increased from 9 ± 4 mm Hg to 15 ± 3 mm Hg (P = .018). Repeated echocardiography showed a drastic decrease in the left ventricular ejection fraction from baseline to latest measurements as well as a significant dilation in the left ventricle shown through an increase in left ventricular end-systolic and left ventricular end-diastolic volume. The left ventricular ejection fraction decreased significantly from 63 ± 13% (baseline) to 43 ± 6% (latest) (P = .012), whereas left ventricular end-diastolic volume and left ventricular end-systolic volume both increased from 55 ± 12 mL (baseline) to 70 ± 22 mL (latest) (P = .03) and 22 ± 5 mL (baseline) to 42 ± 16 mL (latest) (P = .022), respectively. Meanwhile, right ventricle function was assessed using echocardiography and through calculating the pulmonary artery pulsatility index (PAPi), a marker for RV dysfunction, at baseline and previous LVAD implantation. The right ventricle function was described in echocardiography as normal in all animals, and PAPi data revealed no significant change (0.91 ± 0.63 at baseline vs 1.15 ± 0.56 in stable HF; P = .39) indicating that right ventricle function was not impaired at the time of LVAD implantation. All 10 animals underwent second surgery for LVAD implantation and TVI induction. The animals were connected to CPB for a total of 101 ± 22 minutes and the total LVAD support time was 41 ± 20 minutes. Results comparing the hemodynamic parameters in no TVI condition to the parameters in TVI condition are shown in Figures 3 and 4. Changes in HR, MAP, and mPAP were not significant when comparing the 2 groups. HR showed no changes with 95 (82-127) bpm in no TVI and 98 (87-122) bpm in TVI condition (P = .953). MAP also remained stable at 64 (60-74) mm Hg (no TVI) and 68 (63-71) mm Hg (TVI) (P = .735). Similarly, no major changes in the mPAP was observed (35 [26-43] mm Hg in no TVI and 32 [27-37] mm Hg in TVI condition, P = .352). As expected, a statistically significant increase of the CVP was observed from 15 (13-25) mm Hg in no TVI to 26 (24-31) mm Hg in the TVI condition (P = .020). Furthermore, the CO decreased from 3.2 (1.55-3.7) L/min in no TVI to 2 L/min (1.38-2.8) in the TVI condition (P = .027).
Figure 3

Changes in hemodynamic data in both conditions (no TVI versus TVI). LVAD, Left ventricular assist device; TVI, tricuspid valve insufficiency; MAP, mean arterial pressure; mPAP, mean pulmonary artery pressure; CVP, central venous pressure.

Figure 4

Changes in hemodynamic data. CO and HR in both conditions. HR, Heart rate; CO, cardiac output; LVAD, left ventricular assist device; TVI, tricuspid valve insufficiency.

Changes in hemodynamic data in both conditions (no TVI versus TVI). LVAD, Left ventricular assist device; TVI, tricuspid valve insufficiency; MAP, mean arterial pressure; mPAP, mean pulmonary artery pressure; CVP, central venous pressure. Changes in hemodynamic data. CO and HR in both conditions. HR, Heart rate; CO, cardiac output; LVAD, left ventricular assist device; TVI, tricuspid valve insufficiency. There were no significant changes displayed in the LVAD parameters (Figure 5). A median LVAD pump speed of 2300 (2200-2400) revolutions per minute (no TVI condition) and 2300 (2200-2400) revolutions per minute (TVI condition) was documented (P = .317). Meanwhile, a median LVAD flow of 2.9 (2.75-3.8) L/min (no TVI condition) and 2.8 (2.45-3.75) L/min (TVI condition) was obtained (P = .160). The median LVAD power consumption was 2.3 (2.05-2.45) Watt (W) (no TVI condition) and 2.2 (2.1-2.45) W (TVI condition) (P = .345) (Figure 6, A-D). A summary of the hemodynamic and LVAD parameters is portrayed in Table 1.
Figure 5

Changes in LVAD parameters in both settings; TVI versus no TVI. LVAD, Left ventricular assist device; TVI, tricuspid valve insufficiency.

Figure 6

A-D, Intraoperative photographs of left ventricular assist device (LVAD) monitors depicting LVAD parameters and pulsatility in tricuspid valve insufficiency condition. Pulsatility was not analyzed qualitatively or included in the results. L/min, Liters per minute; U/min, revolutions per minute; Fluss, flow; Zeit, time; s, seconds; VAD, ventricular assist device.

Table 1

Comparison of hemodynamic and LVAD parameters in both settings (no TVI vs TVI condition)

VariableLVADLVAD + TVIP value
HR, bpm95 (82-127)98 (87-122).953
CO, L/min3.2 (1.55-3.7)2 (1.38-2.8).027
MAP, mm Hg64 (60-74)68 (63-71).735
mPAP, mm Hg35 (26-43)32 (27-37).352
CVP, mm Hg15 (13-25)26 (24-31).020
Flow, L/min2.9 (2.75-3.8)2.8 (2.45-3.75).160
Power, W2.3 (2.05-2.45)2.2 (2.1-2.45).345
Speed, rpm2300 (2200-2400)2300 (2200-2400).317

LVAD, Left ventricular assist device; TVI, tricuspid valve insufficiency; HR, heart rate; CO, cardiac output; MAP, mean arterial pressure; mPAP, mean pulmonary artery pressure; CVP, central venous pressure; rpm, revolutions per minute.

Data are expressed as median (interquartile range).

Wilcoxon signed-rank test.

P < .05 statistically significant.

Changes in LVAD parameters in both settings; TVI versus no TVI. LVAD, Left ventricular assist device; TVI, tricuspid valve insufficiency. A-D, Intraoperative photographs of left ventricular assist device (LVAD) monitors depicting LVAD parameters and pulsatility in tricuspid valve insufficiency condition. Pulsatility was not analyzed qualitatively or included in the results. L/min, Liters per minute; U/min, revolutions per minute; Fluss, flow; Zeit, time; s, seconds; VAD, ventricular assist device. Comparison of hemodynamic and LVAD parameters in both settings (no TVI vs TVI condition) LVAD, Left ventricular assist device; TVI, tricuspid valve insufficiency; HR, heart rate; CO, cardiac output; MAP, mean arterial pressure; mPAP, mean pulmonary artery pressure; CVP, central venous pressure; rpm, revolutions per minute. Data are expressed as median (interquartile range). Wilcoxon signed-rank test. P < .05 statistically significant.

Comment

In this acute animal setting of stable HF and severe iatrogenic TVI, no significant differences in the pump parameters were observed between TVI condition and no TVI condition. Further, the hemodynamic parameters stayed stable in both conditions except for a decrease in CO without negative impact on the pump parameters. Functional or secondary TVI is the most frequent form of TVI., It can develop as a consequence of either single or biventricular failure or, although less frequently, of high pulmonary vascular pressure, all resulting in dilation of the annular ring. Further, implantation of an implantable cardioverter/defibrillator or pacemaker can also lead to a hindrance of valve leaflet coaptation., All of the aforementioned factors can be observed in patients with end-stage HF requiring LVAD support. It has been described that LVAD implantation alone improves RV function, thereby leading to reduced TVI severity, which would render TVR obsolete.19, 20, 21 The improvement of the RV function is mostly due to better unloading of the left ventricle after LVAD implantation, resulting in a decrease of pulmonary vascular pressure and therefore of the RV afterload. RV dysfunction may improve over time after LVAD implantation and better tricuspid valve competence may be achieved., Meanwhile, it has also been reported that LVAD implantation may also lead to a leftward deviation of the interventricular septum, particularly if run at greater speed mode, which might possibly cause an exacerbation of the TVI. Considering the association between TVI and right ventricle failure (RVF), the presumption that concomitant tricuspid valve procedures along with LVAD implantation could possibly contribute to minimizing the chance of RVF is postulated. While it has been demonstrated that concomitant tricuspid valve procedure does not increase the risk of operative mortality and early postoperative morbidity, performing either TVR or replacement does not affect the clinical outcome in patients with LVAD., Piacentino and colleagues showed that TVR reduces right HF.27, 28, 29 Furthermore, they describe a prolonged CPB time in the TVR cohort although shorter hospitalization. However, a difference in mortality between the 2 groups was not detected. The question whether to perform TVR is heavily debated and controversial. Recently, mostly single-center studies have come to disparate conclusions.9, 10, 11, 12, 13, 14, 15, 16 Saeed and colleagues conducted a retrospective analysis of patients undergoing LVAD implantation with or without TVR. The main finding of this study was that TVR led to prolonged CPB time and greater requirements of blood products, thus concluding that omitting TVR is reasonable in VAD candidates. Veen and colleagues have recently carried out a meta-analysis of 8 publications. They showed that concomitant TVR during LVAD implantation provides similar results in late mortality and RVF when compared with LVAD implantation alone. Furthermore, they failed to show any significant differences in early mortality, RVF, acute kidney failure, hospital stay, and need for RV assist device. The systematic review by Dunlay and colleagues focused on early outcomes supports the findings of Veen and colleagues. The database analyses of INTERMACS (Interagency Registry for Mechanically Assisted Circulatory Support), EUROMACS (European Registry for Patients with Mechanical Circulatory Support), and the Society of Thoracic Surgeons have led to comparable results.,, These database analyses show that TVR is associated with worse early postoperative outcomes but do not affect surgical mortality. Furthermore, TVR led to significantly improved TVI shortly after TVR, but differences in probability of TVI disappeared between the TVR and the non-TVR group over a period of time., In addition, TVR does not show a survival benefit; thus, TVR may not be beneficial. Meanwhile, Robertson and colleagues found longer hospital stays and a significantly greater risk of renal failure in the TVR group and raised the question that the decision for TVR should not solely be based on echocardiographic findings. In summary, the majority of currently available literature show that TVR at the time of LVAD implantation may not be necessary. Our study is the first animal and translational research study dealing with this controversial topic (Video 1). We have shown that the induction of severe iatrogenic TVI in stable HF ovines with preserved right ventricle function is feasible in acute settings. We appreciate that establishing an animal model of HF that exactly simulates patients with end-stage HF is challenging. We intentionally chose a model that preliminarily causes left ventricular failure without major impact on the RV function. We believe that severely reduced RV function and functional TVI may rather benefit from better preconditioning before the pump implantation procedure (aggressive diuresis, etc.) and/or biventricular support strategy in emergent cases than TVR. That being said, we hypothesize that moderate TVI without severe RV dysfunction, such as those caused by defibrillator leads, may eliminate the need for TVR. We postulate that adding TVR procedure may not provide any advantage. The senior author explains the concept, importance, and relevance of this study. Video available at: https://www.jtcvs.org/article/S2666-2736(20)30107-8/fulltext. In this study, we were able to establish stable HF with hemodynamic and echocardiographic parameters mimicking HF. In the second-stage procedure, severe acute TVI was induced through surgical resection of the TV. All hemodynamic parameters remain stable in the group with LVAD alone or LVAD with TVI except for expected greater CVP value in the TVI group (15 [13-25] mm Hg vs 26 [24-31] mm Hg, P = .020) and lower CO in the TVI group (3.2 [1.55-3.7] L/min vs 2 [1.38-2.8] L/min, P = .027). The increase of CVP was expected as a result from the TV resection. The decrease of the CO could possibly be due to the second initiation of the CPB, which was conducted to perform the resection of the valve resulting in prolonged CPB time. The duration and re-initiation of the CPB cause additional immunologic and operative trauma, which may explain this finding. Another explanation of the reduction in CO may be the fact that left ventricular filling may be better in the absence of TVI leading to more CO. Notably, the RV output, which was measured using ultrasonic flow probes, represents total CO whereas pump flow only reflects the flow through the pump. That being said, we assume that in the TVI condition, the majority of the CO went through the pump. Meanwhile, in the no TVI condition the remaining CO went through the native aortic valve. Interestingly, the performance of the LVAD remains similar in both conditions. An average pump flow of 2.9 (2.75-3.8) L/min in the no TVI group versus 2.8 (2.45-3.75) L/min in the TVI group was documented (P = .160). This finding may indicate that having TVI per se without major RV dysfunction may not negatively impact the pump performance. Needless to say, the duration of the experiments was not long enough to allow investigating the impact of chronic TVI on the outcome. Limitations of this study are similar to any other animal study of HF, as a one-to-one model simulating patients with end-stage HF is challenging. However, the changes in the diastolic pulmonary pressures and echocardiography findings indicated severe HF in our study. Further, we believe that our model was representative enough to show the effect of severe TVI in the setting of continuous flow pumps. Notably, all of these animals had preserved RV function. Therefore, our results may be limited to those VAD candidates with TVI and preserved RV function such as those caused by pacing leads. Furthermore, the validity of this acute model is limited to the short period of observation time as well as the rather short follow-up time after induction of HF. In addition, the setting of an acute iatrogenic TVI might not reflect the pathophysiological changes occurring in patients with chronic TVI over time, which makes drawing a conclusion about clinical practice difficult. Therefore, we plan to repeat this study in a chronic model with an increased number of animals and randomized in comparable TVI and no TVI condition cohorts in the near future. In fact, we already performed 2 successful chronic experiments. One ovine survived LVAD implantation. However, the study was terminated after 5 days due to LVAD pump thrombosis. Another ovine survived LVAD implantation with TV resection but died afterwards due to respiratory insufficiency. Additional limitation includes the fact that no detailed echocardiographic measurement of the right ventricle was performed before LVAD implantation. Nevertheless, all animals had a good right ventricular function based on PAPi data and echocardiographic observation at the time of LVAD procedure. The induced HF was limited to the left ventricle in these animals. We hypothesize that presence of TVI with preserved RV function would not impact the VAD parameters. Another limitation is the fact that we were comparing data of the same animals in 2 different conditions. The presumably negative impact of second round of CPB on CO might have been eliminated if we would have had performed either LVAD or LVAD with TVI. A true matched control would have had a re-initiation of CPB in a second group of animals and sham right atriotomy in the no TVI condition. However, this would have necessitated more animals and more resources. Needless to say, another limitation is the fact that the pump flows were calculated flows and even updated hematocrit values may not guarantee real-time flow measurements. In conclusion, our data suggest that TVI in subjects with HF with preserved RV function did not negatively impact LVAD parameters. The observed reduction in CO may warrant further investigation, especially considering loading conditions. Further chronic experiments as well as clinical trials are needed to conclusively clarify the necessity of TVR in patients with LVAD. However, results from this study could possibly encourage omitting the TVR procedure, particularly in LVAD candidates with preserved RV function and moderate TVI and/or TVI due to pacing leads.

Conflict of Interest Statement

The authors reported no conflicts of interest. The Journal policy requires editors and reviewers to disclose conflicts of interest and to decline handling or reviewing manuscripts for which they may have a conflict of interest. The editors and reviewers of this article have no conflicts of interest.
  34 in total

1.  Effects of the HeartMate II continuous-flow left ventricular assist device on right ventricular function.

Authors:  Sangjin Lee; Forum Kamdar; Richard Madlon-Kay; Andrew Boyle; Monica Colvin-Adams; Marc Pritzker; Ranjit John
Journal:  J Heart Lung Transplant       Date:  2010-02       Impact factor: 10.247

2.  Surgical treatment of tricuspid valve insufficiency promotes early reverse remodeling in patients with axial-flow left ventricular assist devices.

Authors:  Simon Maltais; Yan Topilsky; Vakhtang Tchantchaleishvili; Stephen H McKellar; Lucian A Durham; Lyle D Joyce; Richard C Daly; Soon J Park
Journal:  J Thorac Cardiovasc Surg       Date:  2011-08-11       Impact factor: 5.209

Review 3.  Heart Disease and Stroke Statistics-2018 Update: A Report From the American Heart Association.

Authors:  Emelia J Benjamin; Salim S Virani; Clifton W Callaway; Alanna M Chamberlain; Alexander R Chang; Susan Cheng; Stephanie E Chiuve; Mary Cushman; Francesca N Delling; Rajat Deo; Sarah D de Ferranti; Jane F Ferguson; Myriam Fornage; Cathleen Gillespie; Carmen R Isasi; Monik C Jiménez; Lori Chaffin Jordan; Suzanne E Judd; Daniel Lackland; Judith H Lichtman; Lynda Lisabeth; Simin Liu; Chris T Longenecker; Pamela L Lutsey; Jason S Mackey; David B Matchar; Kunihiro Matsushita; Michael E Mussolino; Khurram Nasir; Martin O'Flaherty; Latha P Palaniappan; Ambarish Pandey; Dilip K Pandey; Mathew J Reeves; Matthew D Ritchey; Carlos J Rodriguez; Gregory A Roth; Wayne D Rosamond; Uchechukwu K A Sampson; Gary M Satou; Svati H Shah; Nicole L Spartano; David L Tirschwell; Connie W Tsao; Jenifer H Voeks; Joshua Z Willey; John T Wilkins; Jason Hy Wu; Heather M Alger; Sally S Wong; Paul Muntner
Journal:  Circulation       Date:  2018-01-31       Impact factor: 29.690

4.  Influence of left ventricular assist on valvular regurgitation.

Authors:  W L Holman; R C Bourge; P Fan; J K Kirklin; A D Pacifico; N C Nanda
Journal:  Circulation       Date:  1993-11       Impact factor: 29.690

Review 5.  Impact of tricuspid valve surgery at the time of left ventricular assist device insertion on postoperative outcomes.

Authors:  Shannon M Dunlay; Salil V Deo; Soon J Park
Journal:  ASAIO J       Date:  2015 Jan-Feb       Impact factor: 2.872

6.  Utility of concomitant tricuspid valve procedures for patients undergoing implantation of a continuous-flow left ventricular device.

Authors:  Valentino Piacentino; Asvin M Ganapathi; Mark Stafford-Smith; Michael K Hsieh; Chetan B Patel; Alan A Simeone; Joseph G Rogers; Carmelo A Milano
Journal:  J Thorac Cardiovasc Surg       Date:  2012-08-29       Impact factor: 5.209

7.  Impact of tricuspid valve regurgitation in patients treated with implantable left ventricular assist devices.

Authors:  Valentino Piacentino; Matthew L Williams; Tim Depp; Karla Garcia-Huerta; Laura Blue; Andrew J Lodge; G Burkhard Mackensen; Madhav Swaminathan; Joseph G Rogers; Carmelo A Milano
Journal:  Ann Thorac Surg       Date:  2011-04-01       Impact factor: 4.330

8.  Durability and clinical impact of tricuspid valve procedures in patients receiving a continuous-flow left ventricular assist device.

Authors:  Jiho Han; Koji Takeda; Hiroo Takayama; Paul A Kurlansky; Christine M Mauro; Paolo C Colombo; Melana Yuzefpolskaya; Shinichi Fukuhara; Lauren K Truby; Veli K Topkara; Arthur R Garan; Donna M Mancini; Yoshifumi Naka
Journal:  J Thorac Cardiovasc Surg       Date:  2015-09-28       Impact factor: 5.209

9.  Clinical outcomes after implantation of a centrifugal flow left ventricular assist device and concurrent cardiac valve procedures.

Authors:  Carmelo Milano; Francis D Pagani; Mark S Slaughter; Duc Thinh Pham; David R Hathaway; Mary V Jacoski; Kevin B Najarian; Keith D Aaronson
Journal:  Circulation       Date:  2014-09-09       Impact factor: 29.690

10.  Concomitant tricuspid valve repair or replacement during left ventricular assist device implant demonstrates comparable outcomes in the long term.

Authors:  Salil V Deo; Tal Hasin; Salah E Altarabsheh; Stephen H McKellar; Ishan K Shah; Lucian Durham; John M Stulak; Richard C Daly; Soon J Park; Lyle D Joyce
Journal:  J Card Surg       Date:  2012-11       Impact factor: 1.620

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.