Thuy Phan1, Rebecca Stephenson1, Tina Cai1, Nikol Andacic1, Genevieve McKew1,2.
Abstract
The QuickGene-810 Nucleic Acid Isolation System is a semi-automated extraction platform which may be used for RNA extraction. New methods were required to support the rapid increase in respiratory virus testing during the SARS-CoV-2 pandemic. The aim of this study was to assess SARS-CoV-2 RNA extraction using the QuickGene-810 kit compared to the EZ1 Advanced Extraction Platform for use on the AusDiagnostics SARS-CoV-2, Influenza and RSV 8-well RT-PCR assay. Qualitative results from all clinical samples were concordant between the QuickGene-810 and the EZ1 extraction methods, demonstrating that the QuickGene-810 kit is suitable for use in pathogen diagnostics. However, there was an average difference of approximately two cycles between the cycle threshold (Ct) values for both SARS-CoV-2 targets, suggesting that the EZ1 kit yields a higher concentration of nucleic acid extract, possibly related to its use of carrier RNA and/or smaller elution volume, which infers the possibility of false negative results for samples with very low viral loads.Entities:
Keywords: QuickGene-810; RNA extraction; SARS-CoV-2
Year: 2022 PMID: 36003356 PMCID: PMC9394536 DOI: 10.1099/acmi.0.000353
Source DB: PubMed Journal: Access Microbiol ISSN: 2516-8290
Comparison of the categorical detection of SARS-CoV-2 RNA from clinical samples extracted with the QuickGene-810 (Kurabo) vs. the EZ1 Advanced (Qiagen) and run on the SARS-CoV-2, Influenza and RSV 8-well assay (AusDiagnostics)
|
Extraction method |
Qiagen EZ1 Advanced | ||
|---|---|---|---|
|
Detected |
Not Detected | ||
|
Kurabo QuickGene-810 |
Detected |
27 |
0 |
|
Not detected |
0 |
83 | |
Comparison of the limit of detection (Ct values, as provided by the MT-Analyzer software; AusDiagnostics) of SARS-CoV-2 RNA from samples extracted with the QuickGene-810 (Kurabo) vs. the EZ1 Advanced (Qiagen) from samples with high, medium and low viral concentration, in dilution series run on the SARS-CoV-2, Influenza and RSV 8-well assay (AusDiagnostics)
|
Target |
Sample concentration range |
Extraction platform |
Ct of neat sample |
Ct of 1 : 10 dilution |
Ct of 1 : 100 dilution |
Ct of 1 : 1000 dilution |
Ct of 1 : 10 000 dilution |
Ct of 1 : 100 000 dilution |
|---|---|---|---|---|---|---|---|---|
|
|
High |
QuickGene-810 |
20 |
23 |
29.27 |
31.08 |
– |
– |
|
|
|
EZ1 |
17 |
21 |
25.71 |
29.09 |
32.45 |
– |
|
– |
Medium |
QuickGene-810 |
32.05 |
– |
– |
– |
– |
– |
|
|
|
EZ1 |
27.93 |
32.18 |
33.21 |
– |
– |
– |
|
|
Low |
QuickGene-810 |
– |
– |
– |
– |
– |
– |
|
|
|
EZ1 |
31.64 |
– |
– |
– |
– |
– |
|
|
High |
QuickGene-810 |
18 |
21 |
27.06 |
27.97 |
33.11 |
– |
|
|
|
EZ1 |
15 |
19 |
23 |
26.39 |
29.9 |
32.19 |
|
|
Medium |
QuickGene-810 |
29.75 |
31.71 | ||||
|
|
|
EZ1 |
25.82 |
29.19 |
32.88 |
– |
– |
– |
|
|
Low |
QuickGene-810 |
31.95 |
– |
– |
– |
– |
– |
|
|
|
EZ1 |
29.73 |
31.54 |
– |
– |
– |
– |
Comparison of the limit of detection (concentration, in relative concentration units, as provided by the MT-Analyzer software; AusDiagnostics) of SARS-CoV-2 RNA from samples extracted with the QuickGene-810 (Kurabo) vs. the EZ1 Advanced (Qiagen) from samples with high, medium and low viral concentration, in dilution series run on the SARS-CoV-2, Influenza and RSV 8-well assay (AusDiagnostics)
|
Target |
Sample concentration range |
Extraction platform |
Concentration of neat sample |
Concentration of 1 : 10 dilution |
Concentration of 1 : 100 dilution |
Concentration of 1 : 1000 dilution |
Concentration of 1 : 10 000 dilution |
Concentration of 1 : 100 000 dilution |
|---|---|---|---|---|---|---|---|---|
|
|
High |
QuickGene-810 |
33 376 |
3805 |
84 |
29 |
– |
– |
|
|
|
EZ1 |
172 806 |
18 349 |
537 |
69 |
8 |
– |
|
|
Medium |
QuickGene-810 |
15 |
– |
– |
– |
– |
– |
|
|
|
EZ1 |
166 |
11 |
5 |
– |
– |
– |
|
|
Low |
QuickGene-810 |
– |
– |
– |
– |
– |
– |
|
|
|
EZ1 |
16 |
– |
– |
– |
– |
– |
|
|
High |
QuickGene-810 |
112 840 |
15 563 |
347 |
211 |
7 |
– |
|
|
|
EZ1 |
529 511 |
64 416 |
2928 |
392 |
39 |
11 |
|
|
Medium |
QuickGene-810 |
65 |
14 |
– |
– |
– |
– |
|
|
|
EZ1 |
642 |
72 |
7 |
– |
– |
– |
|
|
Low |
QuickGene-810 |
13 |
– |
– |
– |
– |
– |
|
|
|
EZ1 |
53 |
14 |
– |
– |
– |
– |
Comparison of the limit of detection (qualitative) of SARS-CoV-2 RNA from samples extracted with the QuickGene-810 (Kurabo) vs. the EZ1 Advanced (Qiagen) from samples with high, medium and low viral concentration, in dilution series run on the SARS-CoV-2, Influenza and RSV 8-well assay (AusDiagnostics)
|
Target |
Sample concentration range |
Extraction platform |
Result of neat sample |
Result of 1 : 10 dilution |
Result of 1 : 100 dilution |
Result of 1 : 1000 dilution |
Result of 1 : 10000 dilution |
Result of 1 : 100000 dilution |
|---|---|---|---|---|---|---|---|---|
|
|
High |
QuickGene-810 |
+ |
+ |
+ |
+ |
Check* |
− |
|
|
|
EZ1 |
+ |
+ |
+ |
+ |
− |
− |
|
|
Medium |
QuickGene-810 |
+ |
Check* |
− |
− |
− |
− |
|
|
|
EZ1 |
+ |
+ |
+ |
− |
− |
− |
|
|
Low |
QuickGene-810 |
− |
− |
− |
− |
− |
− |
|
|
|
EZ1 |
+ |
− |
− |
− |
− |
− |
|
|
High |
QuickGene-810 |
+ |
+ |
+ |
+ |
+ |
− |
|
|
|
EZ1 |
+ |
+ |
+ |
+ |
+ |
+ |
|
|
Medium |
QuickGene-810 |
+ |
+ |
− |
− |
− |
− |
|
|
|
EZ1 |
+ |
+ |
+ |
− |
− |
− |
|
|
Low |
QuickGene-810 |
+ |
− |
− |
− |
− |
− |
|
|
|
EZ1 |
+ |
+ |
− |
− |
− |
− |
+, Sample result was positive by the assay’s software; −, negative result.
*As per the AusDiagnostics High-Plex IFU, a check is defined as a case when the cycling acceleration is less than the pre-set parameters and therefore no concentration estimate is provided. For this verification, a ‘check’ will be considered as positive as they are known positive samples.