Literature DB >> 33856292

Development, deployment and in-field demonstration of mobile coronavirus SARS-CoV-2 Nucleic acid amplification test.

Teagan F Paton1, Ian Marr2, Zoe O'Keefe1, Timothy J J Inglis3,1.   

Abstract

Introduction. The evolving SARS-CoV-2 coronavirus pandemic presents a series of challenges to clinical diagnostic services. Many proprietary PCR platforms deployed outside centralised laboratories have limited capacity to upscale when public health demands increase. We set out to develop and validate an open-platform mobile laboratory for remote area COVID-19 diagnosis, with a subsequent field trial.Gap Statement. In regional Western Australia, molecular diagnostic support is limited to near point-of-care devices. We therefore aimed to demonstrate open-platform capability in a rapidly deployable format within the context of the COVID-19 pandemic.Methodology. We compared, selected and validated components of a SARS-CoV-2 RT-PCR assay in order to establish a portable molecular diagnostics laboratory. The optimal combination of PCR assay equipment, reagents and consumables required for operation to national standards in regional laboratories was identified. This comprised RNA extraction and purification (QuickGene-480, Kurabo, Japan), a duplex RT-PCR assay (Logix Smart COVID-19, Co-Diagnostics, USA), a Myra liquid handling robot (Biomolecular Systems, Australia) and a magnetic induction thermal cycler (MIC, Biomolecular Systems).Results The 95 and 99% limits of detection were 1.01 copies μl-1 (5.05 copies per reaction) and 2.80 copies μl-1 (14.00 copies per reaction) respectively. The Co-Diagnostics assay amplified both SARS-CoV-1 and -2 RNA but showed no other cross reactivity. Qualitative results aligned with the reference laboratory SARS-CoV-2 assay (sensitivity 100% [95 % CI 96.48-100%], specificity 100% [95% CI 96.52-100%]). In field trials, the laboratory was operational within an hour of arrival on-site, can process up to 36 samples simultaneously, produces results in two and a half hours from specimen reception, and performed well during six consecutive runs during a 1 week deployment.Conclusion. Our mobile laboratory enables an adaptive response to increased test demand, and unlike many proprietary point-of-care PCR systems, rapid substitution with an alternative assay if gene targets change or reagent supply chains fail. We envisage operation of this RT-PCR assay as a standby capability to meet varying regional test demands under public health emergency operations guidance.

Entities:  

Keywords:  COVID-19; RT-PCR; SARS-CoV-2; coronavirus; diagnostic test performance; mobile laboratory; molecular diagnostics; nucleic acid test

Mesh:

Substances:

Year:  2021        PMID: 33856292     DOI: 10.1099/jmm.0.001346

Source DB:  PubMed          Journal:  J Med Microbiol        ISSN: 0022-2615            Impact factor:   2.472


  3 in total

1.  Estimating COVID Risk During a Period of Pandemic Decline.

Authors:  Timothy J J Inglis; Benjamin McFadden; Anthony Macali
Journal:  Front Public Health       Date:  2021-12-17

2.  Lab-in-a-van: Rapid SARS-CoV-2 testing response with a mobile laboratory.

Authors:  Susan A Ballard; Maryza Graham; Debra David; Tuyet Hoang; Angela Donald; Michelle Sait; Nicole Isles; Amelia Matlock; Sarah Yallop; Mark Bek; Benjamin P Howden; Timothy P Stinear
Journal:  EBioMedicine       Date:  2022-04-08       Impact factor: 11.205

3.  A comparison of SARS-CoV-2 RNA extraction with the QuickGene-810 Nucleic Acid Isolation System compared to the EZ1 Advanced DSP Virus Kit.

Authors:  Thuy Phan; Rebecca Stephenson; Tina Cai; Nikol Andacic; Genevieve McKew
Journal:  Access Microbiol       Date:  2022-05-27
  3 in total