| Literature DB >> 36001137 |
Charles Joseph O'Connor1, Donnacha Hogan2, Lee Chien Yap2, Louise Lyons2, Derek Barry Hennessey2.
Abstract
INTRODUCTION ANDEntities:
Keywords: Begostone; Nephrolithotomy; Percutaneous; Settings; Trilogy
Mesh:
Substances:
Year: 2022 PMID: 36001137 PMCID: PMC9512712 DOI: 10.1007/s00345-022-04127-8
Source DB: PubMed Journal: World J Urol ISSN: 0724-4983 Impact factor: 3.661
Fig. 1Set up of experiments. A IRP measurement: A = 100 cm IV stand with irrigating fluid, B = 60 cm IV stand, C = cystometer, D = Trilogy handpiece, E = nephroscope, F = porcine kidney, G = pressure transducer. B Stone clearance measurement: H = filter, I = filter outflow tubing, J = inner sponge, K = small plastic container, L = water bath, M = Trilogy handpiece
Fig. 2Effect of varying frequency and impact of total stone clearance time for the different stone phantoms. COM calcium oxalate monohydrate, CaP Calcium phosphate, Hz hertz (frequency). Statistical significance between lithotripsy settings assessed using ANOVA (Tukey's multiple comparison). **p < 0.001
Comparison of physical characteristics of artificial kidney stones to real kidney stones
| Begostone powder to water ratio | CL (m/s) | % Difference CL (m/s) | P (Kg/m3) | % Difference P (Kg/m3) | Kidney stone type | CL (m/s) [ | P (Kg/m3) [ |
|---|---|---|---|---|---|---|---|
| 15:3 | 4201 ± 27 | 7.4 | 2056 ± 6 | 0.9 | COM | 4535 ± 58 | 2038 ± 34 |
| 15:4 | 3815 ± 0 | 3 | 1825 ± 34 | 15.4 | Cap | 3932 ± 134 | 2157 ± 16 |
| 15:5 | 3374 ± 48 | 2.8 | 1717 ± 61 | 10 | Uric acid | 3471 ± 62 | 1546 ± 12 |
| 15:6 | 3325 ± 90 | 15.8 | 1753 ± 29 | 9.5 | Struvite | 2798 ± 82 | 1587 ± 68 |