Comparison of quantitative imaging characteristics between a new, larger-FOV 1000 fps high-speed angiographic (HSA) photon-counting detector (Aries) with a smaller HSA detector (Actaeon).

X Wu1, A Shields1, S V Setlur Nagesh1, D R Bednarek1, S Rudin1.   

Abstract

High Speed Angiography (HSA) requires imaging detectors with both high-temporal and high-spatial resolution. Both the Aries and Acteon detectors by Direct Conversion (Stockholm, Sweden) are CdTe direct photon-counting detectors (PCD) that have acquisition frame rates of up to 1000-fps and a 100-micrometer pixel pitch; however, the new Aries detector offers a larger field of view (512 × 768 pixels) compared to the smaller Actaeon detector (256 × 256 pixels). An expanded field of view is required for imaging of larger vasculature, thus the Aries offers this advantage. Evaluations were performed of both detectors under Anti-Coincidence Circuitry (ACC-ON) mode, which corrects for charge sharing between pixels. Initial evaluations of instrumentation noise and detector energy-threshold calibration using Am-241 gamma spectroscopy were performed for the new Aries detector. Linearity was also evaluated for the Aries for each of the 12 individual modules that compose the detector field to check for homogeneity in response to exposure throughout the detector. Finally, Normalized Noise Power Spectrum (NNPS), Modulation Transfer Function (MTF) and Detective Quantum Efficiency (DQE) were then compared between the Aries and Actaeon detectors at two different exposures and detector energy thresholds. The detectors are linear up to approximately 1000 μR and have no instrumentation noise above a threshold of 15 keV. As expected, the MTF's and DQE's are similar between the Aries and Actaeon detectors, and there are thus no tradeoff's in image quality to achieve the larger FOV.

Entities:  

Keywords:  DQE; High Speed Angiography; MTF; NPS; Photon-Counting Detectors; standard metrics

Year:  2021        PMID: 35982768      PMCID: PMC9385188          DOI: 10.1117/12.2611538

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  6 in total

1.  Intercomparison of methods for image quality characterization. I. Modulation transfer function.

Authors:  Ehsan Samei; Nicole T Ranger; James T Dobbins; Ying Chen
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

Review 2.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

3.  Evaluation of a new photon-counting imaging detector (PCD) with various acquisition modes.

Authors:  A Shankar; J Krebs; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03-09

4.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device.

Authors:  E Samei; M J Flynn; D A Reimann
Journal:  Med Phys       Date:  1998-01       Impact factor: 4.071

5.  Spectroscopy with a CdTe-based photon-counting imaging detector (PCD) having charge sharing correction capability.

Authors:  A Shankar; J Krebs; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03-09

6.  Multi-energy inter-pixel coincidence counters for charge sharing correction and compensation in photon counting detectors.

Authors:  Katsuyuki Taguchi
Journal:  Med Phys       Date:  2020-03-24       Impact factor: 4.071

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.