Multi-energy inter-pixel coincidence counters for charge sharing correction and compensation in photon counting detectors.

Katsuyuki Taguchi1.   

Abstract

PURPOSE: Smaller pixel sizes of x-ray photon counting detectors (PCDs) are advantageous for count rate capabilities but disadvantageous for charge sharing. With charge sharing, the energy of an x-ray photon may be split and one photon may produce two or more counts at adjacent pixels, both at lower energies than the incident energy. This "double-counting" increases noise variance and degrades the spectral response. Overall, it has a significantly negative impact on the performance of PCD-based computed tomography (CT). Charge sharing is induced by the detection physics and occurs regardless of count rates; thus, it is impossible to avoid. We propose in this paper a method that has a potential to address both noise and bias added by charge sharing.
METHODS: We propose applying a multi-energy inter-pixel coincidence counter (MEICC) technique, which uses energy-dependent coincidence counters, keeps the book of charge sharing events during data acquisition, and provides the exact number of charge sharing occurrences, which can be used to either correct or compensate for them after the acquisition is completed. MEICC does not interfere with the primary counting process; therefore, PCDs with MEICC will remain as fast as those without MEICC. MEICC can be implemented using current electronics technology because its inter-pixel coincidence counters used to handle digital data are rather simple. We evaluated Cramér-Rao lower bound (CRLB) of PCDs with and without MEICC using a Monte Carlo simulation.
RESULTS: When the number of energy windows was four or larger and eight neighboring pixels were used, the CRLBs of 225-µm PCD with MEICC normalized by those of the current PCD with the same number of windows were 0.361-0.383 for water density images of two basis functions, which was only 5.7-16.4% worse than those of a PCD without charge sharing (which were at 0.329-0.358). In contrast, the normalized CRLBs of the PCD with one coincidence counter were 0.466-0.499, which were 37.3-45.6% worse than the PCD without charge sharing. The use of eight neighboring pixels provided ~10% better CRLB values than four neighboring pixels for MEICC. With four energy windows, decreasing the number of coincidence counters from 16 to 9 only slightly increased the CRLB from 0.255 to 0.269 (which corresponded to as little as a 5.5% change). The normalized CRLBs of MEICC for K-edge imaging (gold) were 0.295-0.426, while those of the one coincidence counter were 0.926-0.959 and the ideal PCDs were 0.126-0.146.
CONCLUSIONS: The proposed MEICC provides spectral information that can be used to address charge sharing problems in PCDs and is expected to satisfy the requirements for clinical x-ray CT. MEICC is very effective, especially for K-edge imaging, which requires accurate spectral information.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  charge sharing; computed tomography; photon counting; spectral response

Mesh:

Year:  2020        PMID: 31984498     DOI: 10.1002/mp.14047

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Anomalous edge response of cadmium telluride-based photon counting detectors jointly caused by high-flux radiation and inter-pixel communication.

Authors:  Xu Ji; Kevin Treb; Ke Li
Journal:  Phys Med Biol       Date:  2021-04-14       Impact factor: 3.609

2.  Comparison of quantitative imaging characteristics between a new, larger-FOV 1000 fps high-speed angiographic (HSA) photon-counting detector (Aries) with a smaller HSA detector (Actaeon).

Authors:  X Wu; A Shields; S V Setlur Nagesh; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-04-04

3.  Spectral Photon Counting CT: Imaging Algorithms and Performance Assessment.

Authors:  Adam S Wang; Norbert J Pelc
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-07-07

4.  Assessment of Multienergy Interpixel Coincidence Counters (MEICC) for Charge Sharing Correction or Compensation for Photon Counting Detectors With Boxcar Signals.

Authors:  Katsuyuki Taguchi
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-06-17
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.