| Literature DB >> 35974853 |
Austin L Shiver1, Colton Webber2, Taylor Sliker3, Patrick Rushford4, Aaron Shaw5.
Abstract
Introduction Electrosurgery for dissection and hemostasis remains one of the foundational tools for the field of surgery as a whole. Monopolar cautery remains the most utilized modality for achieving the aforementioned goals. Given the prolonged history and pre-modern development of "Bovie" cautery, there remains a paucity of data regarding appropriate settings and intensity for various tissue types, procedures, or locales. As a result, utilized settings depend on precedent and personal preference. We aimed to determine the amount of secondary soft tissue injury by volume and depth beyond the electrocautery pen tip in the skin and subcutaneous tissue as well as skeletal muscle. Methods Porcine samples were used for experimental testing using two testing types: 1) skin and subcutaneous tissue and 2) Skeletal muscle. Sample sizes were standardized at 1 cm3 cubes. For skin samples, tissue injury was created with either a scalpel or electrocautery pen on cut setting, and tested at intensities from 10 to 150 in increments of 10. Skeletal muscle samples were similarly tested using the electrocautery pen only in either a cut or coagulation setting. Samples were tested at incremental intensities from 10 to 120 for both settings. Electrocautery was tested for a period of five seconds with a continuous current. All samples were placed in formalin and underwent histologic staining with hematoxylin and eosin staining to be assessed for the extent of tissue injury in terms of depth, radius, and volume. The measurements were recorded in millimeters. Results For skin incision, there was a positive and significant correlation with respect to the radius (R=.73, p=0.006). When considering intensity with an interval of 10-70 there was a positive and significant correlation with respect to the radius, depth, and volume. The cold knife incision had no notable soft tissue injury beyond the depth of the incision. Regarding skeletal muscle, again, a significant and positive correlation between increasing monopolar settings was noted for both the coagulation and cut functions (R=.84, p=.0005; R=0.84, p=0.0006). A positive correlation was found between increasing cut intensity and volume of soft tissue injury (R=0.73, p=.008); this was not reflected in the coagulation setting. When limited to an intensity range of 10-60, a significant relationship was noted for depth, radius and volume (R=.95, p= <0.001; R=0.98, p= <.001; R=.92, p=.001). Conclusion In all samples, apart from the cold knife skin incision, additional soft tissue injury beyond the tip of the electrocautery pen was noted. Given our findings, recommendations include using the lowest setting required for the purposes of the given surgical case as well as minimizing electrocautery use for skin incisions given its association with a larger volume of tissue injury in comparison with a scalpel. Additionally, electrocautery should be used with care in, and around neurovascular structures as soft tissue injury did occur several millimeters beyond the tip of the electrocautery pen. Further study is needed to see if these patterns are similar in living animals as well as human tissue and whether they bear any clinical impact on surgical wound healing or other surgical complications.Entities:
Keywords: basic orthopedic sciences; bovie; electrocautery; orthopaedic surgery; wound healing
Year: 2022 PMID: 35974853 PMCID: PMC9375435 DOI: 10.7759/cureus.26841
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Skin secondary injury depth, radius, and volume for "cut" function
R value - Pearson's Product-Moment Correlation Coefficient; T stat - Paired T-test
| Intensity Setting (watts) | Depth of Tissue Damage (mm) | Radius of Tissue Damage (mm) | Volume of Tissue Damage (mm3) |
| 10 | 5.00 | 2.45 | 31.43 |
| 20 | 4.20 | 2.80 | 34.48 |
| 30 | 6.50 | 2.30 | 36.01 |
| 40 | 5.70 | 2.90 | 50.19 |
| 50 | 7.90 | 2.80 | 64.86 |
| 60 | 6.80 | 3.80 | 102.83 |
| 70 | 8.30 | 5.20 | 235.02 |
| 80 | 6.55 | 1.75 | 21.01 |
| 90 | 3.75 | 5.65 | 125.36 |
| 100 | 3.00 | 6.60 | 136.85 |
| 110 | 4.00 | 6.25 | 163.62 |
| 120 | 3.50 | 6.90 | 174.50 |
| 130 | 3.00 | 7.05 | 156.14 |
| 140 | 3.70 | 8.40 | 273.39 |
| 150 | 6.00 | 5.65 | 200.57 |
| 300 | 3.00 | 7.30 | 167.41 |
| R Value | -0.4881 | 0.7354 | 0.6011 |
| T-Stat | -1.768 | 3.431 | 2.378 |
| P-value | 0.1074 | 0.0064 | 0.0387 |
Skeletal muscle secondary injury depth, radius, and volume for both "coagulation" and "cut" functions
R value - Pearson's Product-Moment Correlation Coefficient; T stat - Paired T-test
| Intensity Setting (watts) | Coagulation Depth (mm) | Cut depth (mm) | Coagulation Radius (mm) | Cut Radius (mm) | Coagulation Volume (mm3) | Cut Volume (mm3) |
| 10 | 0.50 | 0.55 | 2.40 | 1.50 | 3.01 | 1.29 |
| 20 | 2.10 | 0.75 | 3.00 | 3.60 | 19.79 | 10.17 |
| 30 | 2.10 | 0.85 | 5.15 | 3.10 | 58.32 | 8.55 |
| 40 | 2.60 | 1.50 | 5.20 | 3.55 | 73.62 | 19.79 |
| 50 | 3.00 | 2.10 | 7.00 | 2.35 | 153.93 | 12.14 |
| 60 | 3.80 | 0.90 | 8.45 | 3.25 | 284.13 | 9.95 |
| 70 | 2.40 | 2.25 | 8.00 | 3.67 | 160.84 | 31.82 |
| 80 | 2.00 | 3.00 | 5.95 | 7.00 | 74.14 | 153.93 |
| 90 | 4.40 | 7.00 | 7.80 | 1.52 | 280.33 | 17.04 |
| 100 | 8.80 | 7.60 | 2.50 | 3.50 | 57.59 | 97.49 |
| 110 | 7.90 | 7.00 | 3.00 | 3.70 | 74.45 | 100.35 |
| 120 | 7.60 | 4.50 | 3.20 | 4.55 | 81.49 | 97.55 |
| R value | 0.848 | 0.8407 | -0.0019 | 0.363 | 0.2763 | 0.7214 |
| T-Stat | 5.060 | 4.911 | 0.006 | 1.232 | 0.9091 | 3.294 |
| P-value | 0.0004 | 0.0006 | 0.9952 | 0.246 | 0.3846 | 0.008 |
Skeletal muscle secondary injury depth, radius and volume for "coagulation" and "cut" functions for intensity 10-60
R value - Pearson's Product-Moment Correlation Coefficient; T stat - Paired T-test
| Intensity Setting (watts) | Coagulation Depth (mm) | Cut Depth (mm) | Coagulation Radius (mm) | Cut Radius (mm) | Coagulation Volume (mm3) | Cut Volume (mm3) |
| 10 | 0.50 | 0.55 | 2.40 | 1.50 | 3.01 | 1.29 |
| 20 | 2.10 | 0.75 | 3.00 | 3.60 | 19.79 | 10.17 |
| 30 | 2.10 | 0.85 | 5.15 | 3.10 | 58.32 | 8.55 |
| 40 | 2.60 | 1.50 | 5.20 | 3.55 | 73.62 | 19.79 |
| 50 | 3.00 | 2.10 | 7.00 | 2.35 | 153.93 | 12.14 |
| 60 | 3.80 | 0.90 | 8.45 | 3.25 | 284.13 | 9.95 |
| R value | 0.9506 | 0.5935 | 0.9824 | 0.3565 | 0.9285 | 0.5416 |
| T-stat | 9.686 | 2.332 | 16.63 | 1.206 | 7.912 | 2.037 |
| P-value | 2.126 | 0.0418 | 0.0001 | 0.2552 | 1.295 | 0.0689 |
Figure 1Skin/Subcutaneous tissue at intensity 10 on cut function
Representative histologic image of skin/subcutaneous tissue with zone of secondary injury outlined
Figure 2Skin/Subcutaneous tissue at intensity 70 on cut function
Representative histologic image of skin/subcutaneous tissue with zone of secondary injury outlined
Figure 3Skeletal muscle tissue at intensity 10 on coagulation function
Representative histologic image of skeletal muscle with zone of secondary injury outlined
Figure 5Linear correlation for secondary injury for coagulation and cut functions at intensity 10-60
Figure 7Linear correlation for secondary injury for coagulation and cut functions at intensity 10-60